Complete polyethylene wear-through and secondary breakage of the expansion cup in a ceramic-polyethylene total hip.

Massimo Mariconda1, Alessandro Silvestro, Gelsomina Mansueto, Domenico Marinò.   

Abstract

INTRODUCTION: Catastrophic polyethylene failure is an uncommon complication of ceramic-on-polyethylene total hip arthroplasty (THA) because of the favourable tribological characteristics of these implants. We present a case of a 50-year-old woman who formerly underwent bilateral ceramic-on-polyethylene THA with expansion acetabular cups and eleven years later presented with unilateral THA dislocation, secondary to catastrophic polyethylene failure and metal shell fracture. CASE
PRESENTATION: The patient came to our hospital for worsening pain in her left hip and an acute incapacity to bear weight on her left lower limb. Twelve and eleven years earlier she had undergone bilateral ceramic-on-polyethylene THA with acetabular expansion components of identical size on both sides. Radiographically, the left femoral head appeared superiorly dislocated and severe polyethylene wear was detected. The inclination angles of the left and right cups were 60° and 44°, respectively. The patient underwent left acetabular revision, and complete polyethylene wear-through with fracture of a cranial lobe of the expansion metal shell was noted at surgery. One large osteolytic lesion in the roof of the acetabulum and diffuse periarticular metallosis were also present. These findings required the use of a Burch–Schneider reinforcement cage. Two years later the patient is functioning well and has full autonomy in her activities of daily living.
CONCLUSION: The correct inclination of the acetabular component is necessary to prevent accelerated polyethylene wear in THA, even though favourable articular bearing surfaces have been used (e.g., ceramic-on-polyethylene coupling). Should the cup appear well fixed and fairly oriented on follow-up radiographies, the early detection of severe polyethylene wear may permit a revision of only the femoral head and acetabular liner.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19242700     DOI: 10.1007/s00402-009-0843-7

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  7 in total

1.  [Measurement of acetabular cup inclination in anteroposterior pelvic radiogram : An indicator of quality after primary total hip arthroplasty?]

Authors:  T Loitsch; T Freitag; F Leucht; H Reichel; R Bieger
Journal:  Orthopade       Date:  2018-12       Impact factor: 1.087

2.  Sciatic nerve compression by an extrapelvic cyst secondary to wear debris after a cementless total hip arthroplasty: A case report and literature review.

Authors:  Murat Mert; Yusuf Oztürkmen; Ethem Ayhan Unkar; Sinan Erdoğan; Onat Uzümcügil
Journal:  Int J Surg Case Rep       Date:  2013-07-29

3.  Metallosis: CT findings in a total hip arthroplasty.

Authors:  Jared R Kirkham; Jonelle M Petscavage; Michael L Richardson
Journal:  Radiol Case Rep       Date:  2015-11-06

4.  Catastrophic wear in uncemented acetabular cups after Total Hip Arthroplasty. A case series.

Authors:  Salomon Jasqui Remba; Ricardo Esquivel Gomez; Luis Gutiérrez Ambriz; Daniel Portman Santos; Isaac Baley Amiga
Journal:  Int J Surg Case Rep       Date:  2021-03-30

5.  Influence of Accelerated Aging on the Wear Behavior of Cross-Linked Polyethylene Liners-A Hip Simulator Study.

Authors:  Rebecca H Dammer; Carmen Zietz; Jörn Reinders; Michael Teske; Rainer Bader
Journal:  J Clin Med       Date:  2022-08-01       Impact factor: 4.964

6.  Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty.

Authors:  Givenchy Manzano; Rayna A C Levin; Michael B Mayor; Ran Schwarzkopf
Journal:  J Orthop Case Rep       Date:  2014 Apr-Jun

7.  Do Not Postpone Revision of Worn Conventional Liners in Ceramic-on-Polyethylene Total Hip Arthroplasty: A New Dramatic Failure.

Authors:  Thorsten Gehrke; Mustafa Citak; Hussein Abdelaziz
Journal:  Arthroplast Today       Date:  2021-07-19
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.