Comparison of a new device for the continuous intra-gastric measurement of intra-abdominal pressure (CiMon) with direct intra-peritoneal measurements in cirrhotic patients during paracentesis.

Valentin Becker1, Roland M Schmid, Andreas Umgelter.   

Abstract

OBJECTIVE: To evaluate a new commercially available system for the intra-gastric measurement of intra-abdominal pressure (IAP). DESIGN AND
SETTING: Prospective interventional trial in an intensive care unit of a university hospital. PATIENTS: Ten patients with cirrhosis and tense ascites scheduled for paracentesis and instrumented with a ballon-tipped nasogastric tube. INTERVENTION: Intermittent paracentesis with repeat paired measurements of IAP. MEASUREMENTS AND
RESULTS: Intra-abdominal pressure was measured directly through the paracentesis catheter using a scaled measurement gauge. Simultaneously, intra-gastric measurements were performed with the CiMon device (Pulsion Medical Systems, Munich, Germany), by means of a ballon-tipped nasogastric tube connected to a pressure transducer. The range of IAP was 6.5-28 mmHg for direct measurements and 3.7-16 mmHg for measurements obtained with the CiMon device. The regression equation according to Passing and Bablok for the two methods had an intercept of 0.054 (95% CI-1.475-1.669) and a slope of 0.674 (95% CI 0.538-0.813). The cusum test showed a significant deviation from linearity (P < 0.05). In the Bland-Altman test bias and limits of agreement were - 4.9 (+/-6.8) mmHg or, expressed as percentage of the average, -44% (+/-49%).
CONCLUSION: Intra-gastric measurements of IAP using the CiMon device were not reliable enough to be clinically useful in these ascitic patients.

Entities:  

Mesh:

Year:  2009        PMID: 19242674     DOI: 10.1007/s00134-009-1451-2

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  19 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

Review 2.  Abdominal compartment syndrome.

Authors:  Michael Sugrue
Journal:  Curr Opin Crit Care       Date:  2005-08       Impact factor: 3.687

3.  THE EFFECT OF INCREASED INTRA-ABDOMINAL PRESSURE ON RENAL FUNCTION IN MAN.

Authors:  S E Bradley; G P Bradley
Journal:  J Clin Invest       Date:  1947-09       Impact factor: 14.808

Review 4.  What is the best animal model for ACS?

Authors:  A Schachtrupp; J Wauters; A Wilmer
Journal:  Acta Clin Belg       Date:  2007       Impact factor: 1.264

5.  Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions.

Authors:  Manu L N G Malbrain; Michael L Cheatham; Andrew Kirkpatrick; Michael Sugrue; Michael Parr; Jan De Waele; Zsolt Balogh; Ari Leppäniemi; Claudia Olvera; Rao Ivatury; Scott D'Amours; Julia Wendon; Ken Hillman; Kenth Johansson; Karel Kolkman; Alexander Wilmer
Journal:  Intensive Care Med       Date:  2006-09-12       Impact factor: 17.440

6.  Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study.

Authors:  Manu L N G Malbrain; Davide Chiumello; Paolo Pelosi; David Bihari; Richard Innes; V Marco Ranieri; Monica Del Turco; Alexander Wilmer; Nicola Brienza; Vincenzo Malcangi; Jonathan Cohen; Andre Japiassu; Bart L De Keulenaer; Ronny Daelemans; Luc Jacquet; Pierre-François Laterre; Günther Frank; Paulo de Souza; Bruno Cesana; Luciano Gattinoni
Journal:  Crit Care Med       Date:  2005-02       Impact factor: 7.598

7.  Intra-abdominal hypertension and acute renal failure in critically ill patients.

Authors:  Lidia Dalfino; Livio Tullo; Ilaria Donadio; Vincenzo Malcangi; Nicola Brienza
Journal:  Intensive Care Med       Date:  2007-12-19       Impact factor: 17.440

8.  Incidence and clinical effects of intra-abdominal hypertension in critically ill patients.

Authors:  Maria Gabriela Vidal; Javier Ruiz Weisser; Francisco Gonzalez; Maria America Toro; Cecilia Loudet; Carina Balasini; Hector Canales; Rosa Reina; Elisa Estenssoro
Journal:  Crit Care Med       Date:  2008-06       Impact factor: 7.598

9.  Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study.

Authors:  Manu L N G Malbrain; Davide Chiumello; Paolo Pelosi; Alexander Wilmer; Nicola Brienza; Vincenzo Malcangi; David Bihari; Richard Innes; Jonathan Cohen; Pierre Singer; Andre Japiassu; Elizabeth Kurtop; Bart L De Keulenaer; Ronny Daelemans; Monica Del Turco; P Cosimini; Marco Ranieri; Luc Jacquet; Pierre-François Laterre; Luciano Gattinoni
Journal:  Intensive Care Med       Date:  2004-02-03       Impact factor: 17.440

10.  Effects of plasma expansion with albumin and paracentesis on haemodynamics and kidney function in critically ill cirrhotic patients with tense ascites and hepatorenal syndrome: a prospective uncontrolled trial.

Authors:  Andreas Umgelter; Wolfgang Reindl; Katrin S Wagner; Michael Franzen; Konrad Stock; Roland M Schmid; Wolfgang Huber
Journal:  Crit Care       Date:  2008-01-15       Impact factor: 9.097

View more
  6 in total

1.  Do we need to know body anthropomorphic data whilst measuring abdominal pressure?

Authors:  Manu L N G Malbrain; Inneke De Laet
Journal:  Intensive Care Med       Date:  2009-10-20       Impact factor: 17.440

2.  Comparison of measurement methods: an endless application of wrong statistical methods.

Authors:  B M Cesana; P Antonelli; E Gallazzi; A Marino
Journal:  Intensive Care Med       Date:  2011-03-11       Impact factor: 17.440

Review 3.  Acute kidney injury in the ICU: from injury to recovery: reports from the 5th Paris International Conference.

Authors:  Rinaldo Bellomo; Claudio Ronco; Ravindra L Mehta; Pierre Asfar; Julie Boisramé-Helms; Michael Darmon; Jean-Luc Diehl; Jacques Duranteau; Eric A J Hoste; Joannes-Boyau Olivier; Matthieu Legrand; Nicolas Lerolle; Manu L N G Malbrain; Johan Mårtensson; Heleen M Oudemans-van Straaten; Jean-Jacques Parienti; Didier Payen; Sophie Perinel; Esther Peters; Peter Pickkers; Eric Rondeau; Miet Schetz; Christophe Vinsonneau; Julia Wendon; Ling Zhang; Pierre-François Laterre
Journal:  Ann Intensive Care       Date:  2017-05-04       Impact factor: 6.925

4.  Evaluating intra-abdominal pressures in a porcine model of acute lung injury by using a wireless motility capsule.

Authors:  Stefan Rauch; Amelie Johannes; Bernd Zollhöfer; Ralf M Muellenbach
Journal:  Med Sci Monit       Date:  2012-05

5.  Comparing intra-abdominal pressures in different body positions via a urinary catheter and nasogastric tube: a pilot study.

Authors:  Nirooshan Rooban; Adrian Regli; Wendy A Davis; Bart L De Keulenaer
Journal:  Ann Intensive Care       Date:  2012-07-05       Impact factor: 6.925

Review 6.  The neglected role of abdominal compliance in organ-organ interactions.

Authors:  Manu L N G Malbrain; Yannick Peeters; Robert Wise
Journal:  Crit Care       Date:  2016-03-16       Impact factor: 9.097

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.