Abstract

OBJECTIVE: In a previous study, we assessed the conformability limitations of self-expandable stents to a curved vascular model. The LEO stent (Balt Extrusion, Montmorency, France), one of the current self-expandable models available for intracranial aneurysm stenting, displayed 2 adverse mechanics: flattening of the stent midsection and inward crimping of the proximal and distal ends. We present a follow-up study in which we evaluate the conformability to curved vessels of a second-generation stent, LEO PLUS.
METHODS: A 3.5-x 25-mm LEO PLUS stent was deployed inside a 3-mm x 10-cm polytetrafluoroethylene tube (vascular model) with a simulated 5-mm aneurysm neck at its midsection. The polytetrafluoroethylene tube was then placed in a polystyrene block (styrofoam; Dow Chemical Co., Midland, MI) and bent at different angles ranging from 0 to 150 degrees. For each angle, a rotational radiogram was performed using a C-arm angiographic system with a 30-x 43-cm CsI/amorphous silicon flat detector operated with 23-second rotations, 0.80-degree increments, 166 projections, and a 2480 x 1920 matrix (2K matrix).
RESULTS: The LEO PLUS stent showed symmetric deployment at all tested degrees of curvature, without flattening or kinking. The stent retained its round cylindrical shape at all curvatures without inward crimping of its proximal and distal ends.
CONCLUSION: The previously documented adverse mechanics of the LEO stent were not observed with the new LEO PLUS stent. This suggests better conformability to curved or tortuous vasculature owing to design improvements.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19240562     DOI: 10.1227/01.NEU.0000336304.26756.21

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  10 in total

1.  Contrast-enhanced angiographic cone-beam CT of cerebrovascular stents: experimental optimization and clinical application.

Authors:  N V Patel; M J Gounis; A K Wakhloo; N Noordhoek; J Blijd; D Babic; D Takhtani; S-K Lee; A Norbash
Journal:  AJNR Am J Neuroradiol       Date:  2010-10-21       Impact factor: 3.825

2.  An in vitro study of silk stent morphology.

Authors:  Thaweesak Aurboonyawat; Raphaël Blanc; Paul Schmidt; Michel Piotin; Laurent Spelle; Amir Nakib; Jacques Moret
Journal:  Neuroradiology       Date:  2010-10-28       Impact factor: 2.804

3.  LEO Baby Stent Use following Balloon-Assisted Coiling: Single- and Dual-Stent Technique--Immediate and Midterm Results of 29 Consecutive Patients.

Authors:  P Machi; V Costalat; K Lobotesis; C Ruiz; Y B Cheikh; O Eker; G Gascou; F Danière; C Riquelme; A Bonafé
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

4.  The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study.

Authors:  Chang-Young Lee; Seong-Ho Park; Chang-Hyun Kim; Goetz Benndorf
Journal:  Neuroradiology       Date:  2014-05-28       Impact factor: 2.804

5.  FRED Flow Diverter: A Study on Safety and Efficacy in a Consecutive Group of 50 Patients.

Authors:  H Luecking; T Engelhorn; S Lang; P Goelitz; S Kloska; K Roessler; A Doerfler
Journal:  AJNR Am J Neuroradiol       Date:  2017-01-19       Impact factor: 3.825

6.  Stent-assisted coiling of cerebral aneurysms with the Neuroform Atlas stent.

Authors:  Ferdi Cay; Ahmet Peker; Anıl Arat
Journal:  Interv Neuroradiol       Date:  2018-01-19       Impact factor: 1.610

7.  Angiographic and artefact-free computed tomography imaging of experimental aneurysms embolised with hydrogel filaments.

Authors:  Mark R McCoy; Gregory M Cruise; Monika Killer
Journal:  Eur Radiol       Date:  2009-09-16       Impact factor: 5.315

8.  Coil embolization of intracranial saccular aneurysms using the Low-profile Visualized Intraluminal Support (LVIS™) device.

Authors:  Young Dae Cho; Chul-Ho Sohn; Hyun-Seung Kang; Jeong Eun Kim; Won-Sang Cho; Gyojun Hwang; O-Ki Kwon; Mi-Sun Ko; Nam-Mi Park; Moon Hee Han
Journal:  Neuroradiology       Date:  2014-04-17       Impact factor: 2.804

9.  Effect of structural remodeling (retraction and recoil) of the pipeline embolization device on aneurysm occlusion rate.

Authors:  L-D Jou; B D Mitchell; H M Shaltoni; M E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-10       Impact factor: 3.825

10.  Analysis of branch artery orifice angulation: Feasibility of the shelf technique for the treatment of wide-neck bifurcation aneurysms.

Authors:  Mehmet Onay; Ali Burak Binboga; Cetin Murat Altay
Journal:  Interv Neuroradiol       Date:  2020-11-22       Impact factor: 1.764

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.