Evaluation of a body mass index-adapted protocol for low-dose 64-MDCT coronary angiography with prospective ECG triggering.

Fuminari Tatsugami1, Lars Husmann, Bernhard A Herzog, Nina Burkhard, Ines Valenta, Oliver Gaemperli, Philipp A Kaufmann.   

Abstract

OBJECTIVE: Because an increase in body mass index (weight in kilograms divided by height squared in meters) confers higher image noise at coronary CT angiography, we evaluated a body mass index-adapted scanning protocol for low-dose 64-MDCT coronary angiography with prospective ECG triggering. SUBJECTS AND METHODS: One hundred one consecutively registered patients underwent coronary CTA with prospective ECG triggering with a fixed contrast protocol (80 mL of iodixanol, 50-mL saline chaser, flow rate of 5 mL/s). Tube voltage (range, 100-120 kV) and current (range, 450-700 mA) were adapted to body mass index. Attenuation was measured, and contrast-to-noise ratio was calculated for the proximal right coronary artery and left main coronary artery. Image noise was determined for each patient as the SD of attenuation in the ascending aorta.
RESULTS: Body mass index ranged from 18.2 to 38.8, and mean effective radiation dose from 1.0 to 3.2 mSv. There was no correlation between body mass index and image noise (r = 0.11, p = 0.284), supporting the validity of the body mass index-adapted scanning protocol. However, body mass index was inversely correlated with vessel attenuation (right coronary artery, r = -0.45, p < 0.001; left main coronary artery, r = -0.47, p < 0.001) and contrast-to-noise ratio (right coronary artery, r = -0.39, p < 0.001; left main coronary artery, r = -0.37, p < 0.001).
CONCLUSION: Use of the proposed body mass index-adapted scanning parameters results in similar image noise regardless of body mass index. Increased bolus dilution due to larger blood volume may account for the decrease in contrast-to-noise ratio and vessel attenuation in patients with higher body mass index, but the contrast bolus was not adapted to body mass index in this study.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19234258     DOI: 10.2214/AJR.08.1390

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  28 in total

1.  The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography.

Authors:  F Tatsugami; M Matsuki; G Nakai; Y Inada; S Kanazawa; Y Takeda; H Morita; H Takada; S Yoshikawa; K Fukumura; Y Narumi
Journal:  Br J Radiol       Date:  2012-01-17       Impact factor: 3.039

2.  Individually adapted coronary 64-slice CT angiography based on precontrast attenuation values, using different kVp and tube current settings: evaluation of image quality.

Authors:  Jean-François Paul
Journal:  Int J Cardiovasc Imaging       Date:  2011-10-18       Impact factor: 2.357

3.  Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp.

Authors:  Jin Hyeok Kim; Ki Seok Choo; Tae Yong Moon; Jun Woo Lee; Ung Bae Jeon; Tae Un Kim; Jae Yeon Hwang; Myeong-Ja Yun; Dong Wook Jeong; Soo Jin Lim
Journal:  Eur Radiol       Date:  2015-10-20       Impact factor: 5.315

4.  Deep learning-based image restoration algorithm for coronary CT angiography.

Authors:  Fuminari Tatsugami; Toru Higaki; Yuko Nakamura; Zhou Yu; Jian Zhou; Yujie Lu; Chikako Fujioka; Toshiro Kitagawa; Yasuki Kihara; Makoto Iida; Kazuo Awai
Journal:  Eur Radiol       Date:  2019-04-08       Impact factor: 5.315

5.  Multimodality cardiac imaging.

Authors:  Oliver Gaemperli; Philipp A Kaufmann
Journal:  J Nucl Cardiol       Date:  2010 Jan-Feb       Impact factor: 5.952

6.  Validation of a new contrast material protocol adapted to body surface area for optimized low-dose CT coronary angiography with prospective ECG-triggering.

Authors:  Aju P Pazhenkottil; Lars Husmann; Ronny R Buechel; Bernhard A Herzog; René Nkoulou; Irene A Burger; Andrea Vetterli; Ines Valenta; Jelena R Ghadri; Patrick von Schulthess; Philipp A Kaufmann
Journal:  Int J Cardiovasc Imaging       Date:  2010-02-04       Impact factor: 2.357

7.  Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition.

Authors:  Jelena R Ghadri; Silke M Küest; Robert Goetti; Michael Fiechter; Aju P Pazhenkottil; Rene N Nkoulou; Felix P Kuhn; Carsten Pietsch; Patrick von Schulthess; Oliver Gaemperli; Christian Templin; Philipp A Kaufmann
Journal:  Int J Cardiovasc Imaging       Date:  2011-07-10       Impact factor: 2.357

8.  Image quality and radiation reduction of 320-row area detector CT coronary angiography with optimal tube voltage selection and an automatic exposure control system: comparison with body mass index-adapted protocol.

Authors:  Jiyeon Lim; Eun-Ah Park; Whal Lee; Hackjoon Shim; Jin Wook Chung
Journal:  Int J Cardiovasc Imaging       Date:  2015-01-22       Impact factor: 2.357

9.  A low-dose and an ultra-low-dose contrast agent protocol for coronary CT angiography in a clinical setting: quantitative and qualitative comparison to a standard dose protocol.

Authors:  Dominik C Benz; Christoph Gräni; Beatrice Hirt Moch; Fran Mikulicic; Jan Vontobel; Tobias A Fuchs; Julia Stehli; Olivier F Clerc; Mathias Possner; Aju P Pazhenkottil; Oliver Gaemperli; Ronny R Buechel; Philipp A Kaufmann
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

10.  Increased accuracy in computed tomography coronary angiography; a new body surface area adapted protocol.

Authors:  E E van der Wall; J E van Velzen; F R de Graaf; M M Boogers; J D Schuijf; J J Bax
Journal:  Int J Cardiovasc Imaging       Date:  2010-04-16       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.