A novel Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer has been developed for improved biomolecule analysis. A flared metal capillary and an electrodynamic ion funnel were installed in the source region of the instrument for improved ion transmission. The transfer quadrupole is divided into 19 segments, with the capacity for independent control of DC voltage biases for each segment. Restrained ion population transfer (RIPT) is used to transfer ions from the ion accumulation region to the ICR cell. The RIPT ion guide reduces mass discrimination that occurs as a result of time-of-flight effects associated with gated trapping. Increasing the number of applied DC bias voltages from 8 to 18 increases the number of ions that are effectively trapped in the ICR cell. The RIPT ion guide with a novel voltage profile applied during ion transfer provides a 3- to 4-fold increase in the number of ions that are trapped in the ICR cell compared with gated trapping for the same ion accumulation time period. A novel ICR cell was incorporated in the instrument to reduce radial electric field variation for ions with different z-axis oscillation amplitudes. With the ICR cell, called trapping ring electrode cell (TREC), we can tailor the shape of the trapping electric fields to reduce dephasing of coherent cyclotron motion of an excited ion packet. With TREC, nearly an order of magnitude increase in sensitivity is observed. The performance of the instrument with the combination of RIPT, TREC, flared inlet, and ion funnel is presented.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19200753      PMCID: PMC2763776          DOI: 10.1016/j.jasms.2008.12.022

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  27 in total

1.  Techview: biochemistry. Biomolecule mass spectrometry.

Authors:  F W McLafferty; E K Fridriksson; D M Horn; M A Lewis; R A Zubarev
Journal:  Science       Date:  1999-05-21       Impact factor: 47.728

2.  Characterization of an improved electrodynamic ion funnel interface for electrospray ionization mass spectrometry.

Authors:  S A Shaffer; A Tolmachev; D C Prior; G A Anderson; H R Udseth; R D Smith
Journal:  Anal Chem       Date:  1999-08-01       Impact factor: 6.986

3.  Role of accurate mass measurement (+/- 10 ppm) in protein identification strategies employing MS or MS/MS and database searching.

Authors:  K R Clauser; P Baker; A L Burlingame
Journal:  Anal Chem       Date:  1999-07-15       Impact factor: 6.986

4.  Baseline mass resolution of peptide isobars: a record for molecular mass resolution.

Authors:  F He; C L Hendrickson; A G Marshall
Journal:  Anal Chem       Date:  2001-02-01       Impact factor: 6.986

5.  Top-down mass spectrometry of a 29-kDa protein for characterization of any posttranslational modification to within one residue.

Authors:  Siu Kwan Sze; Ying Ge; HanBin Oh; Fred W McLafferty
Journal:  Proc Natl Acad Sci U S A       Date:  2002-02-12       Impact factor: 11.205

6.  Resolution of 11,000 compositionally distinct components in a single electrospray ionization Fourier transform ion cyclotron resonance mass spectrum of crude oil.

Authors:  Christine A Hughey; Ryan P Rodgers; Alan G Marshall
Journal:  Anal Chem       Date:  2002-08-15       Impact factor: 6.986

7.  Improved ion extraction from a linear octopole ion trap: SIMION analysis and experimental demonstration.

Authors:  Bruce E Wilcox; Christopher L Hendrickson; Alan G Marshall
Journal:  J Am Soc Mass Spectrom       Date:  2002-11       Impact factor: 3.109

8.  Parts-per-billion mass measurement accuracy achieved through the combination of multiple linear regression and automatic gain control in a Fourier transform ion cyclotron resonance mass spectrometer.

Authors:  D Keith Williams; David C Muddiman
Journal:  Anal Chem       Date:  2007-06-01       Impact factor: 6.986

9.  Cyclotron motion of two Coulombically interacting ion clouds with implications to Fourier-transform ion cyclotron resonance mass spectrometry.

Authors: 
Journal:  Phys Rev E Stat Phys Plasmas Fluids Relat Interdiscip Topics       Date:  1995-10

10.  Counting individual sulfur atoms in a protein by ultrahigh-resolution Fourier transform ion cyclotron resonance mass spectrometry: experimental resolution of isotopic fine structure in proteins.

Authors:  S D Shi; C L Hendrickson; A G Marshall
Journal:  Proc Natl Acad Sci U S A       Date:  1998-09-29       Impact factor: 11.205

View more
  6 in total

1.  A novel 9.4 tesla FTICR mass spectrometer with improved sensitivity, mass resolution, and mass range.

Authors:  Nathan K Kaiser; John P Quinn; Gregory T Blakney; Christopher L Hendrickson; Alan G Marshall
Journal:  J Am Soc Mass Spectrom       Date:  2011-05-05       Impact factor: 3.109

2.  Ion Behavior in an Electrically Compensated Ion Cyclotron Resonance Trap.

Authors:  Adam M Brustkern; Don L Rempel; Michael L Gross
Journal:  Int J Mass Spectrom       Date:  2011-03-01       Impact factor: 1.986

3.  Excite-coupled trapping ring electrode cell (eTREC): radial trapping field control, linearized excitation, and improved detection.

Authors:  Chad R Weisbrod; Nathan K Kaiser; Gunnar E Skulason; James E Bruce
Journal:  Anal Chem       Date:  2010-07-15       Impact factor: 6.986

4.  Improved ion optics for introduction of ions into a 9.4-T Fourier transform ion cyclotron resonance mass spectrometer.

Authors:  Yu Chen; Franklin E Leach; Nathan K Kaiser; Xibei Dang; Yehia M Ibrahim; Randolph V Norheim; Gordon A Anderson; Richard D Smith; Alan G Marshall
Journal:  J Mass Spectrom       Date:  2015-01       Impact factor: 1.982

Review 5.  Towards Higher Sensitivity of Mass Spectrometry: A Perspective From the Mass Analyzers.

Authors:  Chang Li; Shiying Chu; Siyuan Tan; Xinchi Yin; You Jiang; Xinhua Dai; Xiaoyun Gong; Xiang Fang; Di Tian
Journal:  Front Chem       Date:  2021-12-21       Impact factor: 5.221

Review 6.  Developments in FTICR-MS and Its Potential for Body Fluid Signatures.

Authors:  Simone Nicolardi; Bogdan Bogdanov; André M Deelder; Magnus Palmblad; Yuri E M van der Burgt
Journal:  Int J Mol Sci       Date:  2015-11-13       Impact factor: 5.923

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.