Mean glandular dose estimation using MCNPX for a digital breast tomosynthesis system with tungsten/aluminum and tungsten/aluminum+silver x-ray anode-filter combinations.

Andy K W Ma1, Dimitra G Darambara, Alexander Stewart, Spencer Gunn, Edward Bullard.   

Abstract

Breast cancer screening with x-ray mammography, using one or two projection images of the breast, is an indispensible tool in the early detection of breast cancer in women. Digital breast tomosynthesis (DBT) is a 3D imaging technique that promises higher sensitivity and specificity in breast cancer screening at a similar radiation dose to conventional two-view screening mammography. In DBT a 3D volume is reconstructed with anisotropic voxels from a limited number of x-ray projection images acquired over a limited angle. Although the benefit of early cancer detection through screening mammography outweighs the potential risks associated with radiation, the radiation dosage to women in terms of mean glandular dose (MGD) is carefully monitored. This work studies the MGD arising from a prototype DBT system under various parameters. Two anode/filter combinations (W/Al and W/Al+Ag) were investigated; the tube potential ranges from 20 to 50 kVp; and the breast size varied between 4 and 10 cm chest wall-to-nipple distance and between 3 and 7 cm compressed breast thickness. The dosimetric effect of breast positioning with respect to the imaging detector was also reviewed. It was found that the position of the breast can affect the MGD by as much as 5% to 13% depending on the breast size.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19175087     DOI: 10.1118/1.3002310

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Digital Breast Tomosynthesis: State of the Art.

Authors:  Srinivasan Vedantham; Andrew Karellas; Gopal R Vijayaraghavan; Daniel B Kopans
Journal:  Radiology       Date:  2015-12       Impact factor: 11.105

2.  Characterization of the homogeneous tissue mixture approximation in breast imaging dosimetry.

Authors:  Ioannis Sechopoulos; Kristina Bliznakova; Xulei Qin; Baowei Fei; Steve Si Jia Feng
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

3.  Digital breast tomosynthesis versus digital mammography: a clinical performance study.

Authors:  Gisella Gennaro; Alicia Toledano; Cosimo di Maggio; Enrica Baldan; Elisabetta Bezzon; Manuela La Grassa; Luigi Pescarini; Ilaria Polico; Alessandro Proietti; Aida Toffoli; Pier Carlo Muzzio
Journal:  Eur Radiol       Date:  2009-12-22       Impact factor: 5.315

Review 4.  A review of breast tomosynthesis. Part I. The image acquisition process.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

5.  Objective models of compressed breast shapes undergoing mammography.

Authors:  Steve Si Jia Feng; Bhavika Patel; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  Clinical digital breast tomosynthesis system: dosimetric characterization.

Authors:  Steve Si Jia Feng; Ioannis Sechopoulos
Journal:  Radiology       Date:  2012-02-13       Impact factor: 11.105

Review 7.  Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography.

Authors:  T M Svahn; N Houssami; I Sechopoulos; S Mattsson
Journal:  Breast       Date:  2014-12-29       Impact factor: 4.380

8.  Dosimetry in x-ray-based breast imaging.

Authors:  David R Dance; Ioannis Sechopoulos
Journal:  Phys Med Biol       Date:  2016-09-12       Impact factor: 3.609

9.  Assessment of Mean Glandular Dose in Mammography System with Different Anode-Filter Combinations Using MCNP Code.

Authors:  Lida Gholamkar; Ali Asghar Mowlavi; Mahdi Sadeghi; Mitra Athari
Journal:  Iran J Radiol       Date:  2016-08-01       Impact factor: 0.212

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.