Objective models of compressed breast shapes undergoing mammography.

Steve Si Jia Feng1, Bhavika Patel, Ioannis Sechopoulos.   

Abstract

PURPOSE: To develop models of compressed breasts undergoing mammography based on objective analysis, that are capable of accurately representing breast shapes in acquired clinical images and generating new, clinically realistic shapes.
METHODS: An automated edge detection algorithm was used to catalogue the breast shapes of clinically acquired cranio-caudal (CC) and medio-lateral oblique (MLO) view mammograms from a large database of digital mammography images. Principal component analysis (PCA) was performed on these shapes to reduce the information contained within the shapes to a small number of linearly independent variables. The breast shape models, one of each view, were developed from the identified principal components, and their ability to reproduce the shape of breasts from an independent set of mammograms not used in the PCA, was assessed both visually and quantitatively by calculating the average distance error (ADE).
RESULTS: The PCA breast shape models of the CC and MLO mammographic views based on six principal components, in which 99.2% and 98.0%, respectively, of the total variance of the dataset is contained, were found to be able to reproduce breast shapes with strong fidelity (CC view mean ADE = 0.90 mm, MLO view mean ADE = 1.43 mm) and to generate new clinically realistic shapes. The PCA models based on fewer principal components were also successful, but to a lesser degree, as the two-component model exhibited a mean ADE = 2.99 mm for the CC view, and a mean ADE = 4.63 mm for the MLO view. The four-component models exhibited a mean ADE = 1.47 mm for the CC view and a mean ADE = 2.14 mm for the MLO view. Paired t-tests of the ADE values of each image between models showed that these differences were statistically significant (max p-value = 0.0247). Visual examination of modeled breast shapes confirmed these results. Histograms of the PCA parameters associated with the six principal components were fitted with Gaussian distributions. The six-component model was also used to generate CC and MLO view mammogram breast shapes, using the mean PCA parameter values of these distributions and randomly generated values based on the fitted Gaussian distributions, which resemble clinically encountered breasts. A spreadsheet with the data necessary to apply this model is provided as the supplementary material.
CONCLUSIONS: Our PCA models of breast shapes in both mammographic views successfully reproduce analyzed breast shapes and generate new clinically relevant shapes. This work can aid in research applications which incorporate breast shape modeling, such as x-ray scatter correction, dosimetry, and image registration.

Entities:  

Mesh:

Year:  2013        PMID: 23464317      PMCID: PMC3585829          DOI: 10.1118/1.4789579

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  37 in total

1.  Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise.

Authors:  I Reiser; R M Nishikawa
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

2.  Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols.

Authors:  D R Dance; K C Young; R E van Engen
Journal:  Phys Med Biol       Date:  2010-12-30       Impact factor: 3.609

3.  Dosimetric characterization of a dedicated breast computed tomography clinical prototype.

Authors:  Ioannis Sechopoulos; Steve Si Jia Feng; Carl J D'Orsi
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

4.  Calculation of the properties of digital mammograms using a computer simulation.

Authors:  R A Hunt; D R Dance; P R Bakic; A D A Maidment; M Sandborg; G Ullman; G Alm Carlsson
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

5.  Computation of the glandular radiation dose in digital tomosynthesis of the breast.

Authors:  Ioannis Sechopoulos; Sankararaman Suryanarayanan; Srinivasan Vedantham; Carl D'Orsi; Andrew Karellas
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

6.  Methodology for generating a 3D computerized breast phantom from empirical data.

Authors:  Christina M Li; W Paul Segars; Georgia D Tourassi; John M Boone; James T Dobbins
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

7.  Development of a multi-spectral, multi-geometry computational model for X-ray breast imaging.

Authors:  N Buls; I Wathion; L Mommaerts; C Breucq; J de Mey
Journal:  Radiat Prot Dosimetry       Date:  2010-02-16       Impact factor: 0.972

8.  Development and characterization of an anthropomorphic breast software phantom based upon region-growing algorithm.

Authors:  Predrag R Bakic; Cuiping Zhang; Andrew D A Maidment
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

9.  Mean glandular dose estimation using MCNPX for a digital breast tomosynthesis system with tungsten/aluminum and tungsten/aluminum+silver x-ray anode-filter combinations.

Authors:  Andy K W Ma; Dimitra G Darambara; Alexander Stewart; Spencer Gunn; Edward Bullard
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

10.  Glandular radiation dose in tomosynthesis of the breast using tungsten targets.

Authors:  Ioannis Sechopoulosa; Carl J D'Orsi
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

View more
  4 in total

1.  The compressed breast during mammography and breast tomosynthesis: in vivo shape characterization and modeling.

Authors:  Alejandro Rodríguez-Ruiz; Greeshma A Agasthya; Ioannis Sechopoulos
Journal:  Phys Med Biol       Date:  2017-08-07       Impact factor: 3.609

2.  Improvements of an objective model of compressed breasts undergoing mammography: Generation and characterization of breast shapes.

Authors:  Alejandro Rodríguez-Ruiz; Steve Si Jia Feng; Jan van Zelst; Suzan Vreemann; Jessica Rice Mann; Carl Joseph D'Orsi; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2017-04-25       Impact factor: 4.071

3.  X-ray scatter correction in breast tomosynthesis with a precomputed scatter map library.

Authors:  Steve Si Jia Feng; Carl J D'Orsi; Mary S Newell; Rebecca L Seidel; Bhavika Patel; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

4.  Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system.

Authors:  Christian Fedon; Marco Caballo; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2018-06-29       Impact factor: 4.071

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.