Cone beam computed tomography, a low-dose imaging technique in the postoperative assessment of cochlear implantation.

Joana Ruivo1, Koen Mermuys, Klaus Bacher, Rudolf Kuhweide, Erwin Offeciers, Jan W Casselman.   

Abstract

BACKGROUND: Cone beam computed tomography (CBCT) has become an extremely useful technique for dentomaxillofacial imaging because it provides clear images of highly contrasted structures. Previous studies evaluating the applicability of this technique in otologic imaging were very encouraging but were only performed in vitro on temporal bone specimens. The intracochlear positioning of the individual electrodes after cochlear implantation by means of CBCT has not yet been shown in vivo.
OBJECTIVE: We describe a protocol for in vivo postoperative imaging of cochlear implants by CBCT. Moreover, the effective dose was measured and compared with the effective dose used on 4- and 16-slice multislice computed tomography (MSCT) by using a RANDO-phantom. MAIN OUTCOME MEASURE: Developing a protocol for in vivo postoperative imaging of cochlear implants by CBCT.
RESULTS: CBCT provides high-resolution and almost artifact-free multiplanar reconstruction images allowing assessment of the precise intracochlear position of the electrode and visualization of each of the individual contacts. The calculated effective dose of the used CBCT and MSCT acquisitions is 80 musv for the CBCT, 3,600 musv for the 16-slice computed tomography, and 4,800 musv for the 4-slice computed tomography.
CONCLUSION: These preliminary results suggests that, for in vivo postoperative evaluation of cochlear implants, CBCT can provide at least the same information as conventional radiography, digital radiograph, and MSCT but in a more comfortable and a much more safer way.

Mesh:

Year:  2009        PMID: 19174709     DOI: 10.1097/MAO.0b013e31819679f9

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  25 in total

1.  Temporal bone imaging using digital volume tomography and computed tomography: a comparative cadaveric radiological study.

Authors:  Afshin Teymoortash; Stella Hamzei; Tobias Murthum; Behfar Eivazi; Ingo Kureck; Jochen A Werner
Journal:  Surg Radiol Anat       Date:  2010-08-13       Impact factor: 1.246

Review 2.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

3.  High-resolution secondary reconstructions with the use of flat panel CT in the clinical assessment of patients with cochlear implants.

Authors:  M S Pearl; A Roy; C J Limb
Journal:  AJNR Am J Neuroradiol       Date:  2013-12-26       Impact factor: 3.825

4.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

5.  Artifacts of the electrode in cochlea implantation and limits in analysis of deep insertion in cone beam tomography (CBT).

Authors:  C Güldner; S Wiegand; R Weiss; S Bien; A Sesterhenn; A Teymoortash; I Diogo
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-07-30       Impact factor: 2.503

6.  Cone-beam computed tomography: a new low dose, high resolution imaging technique of the wrist, presentation of three cases with technique.

Authors:  Jens De Cock; Koen Mermuys; Jean Goubau; Simon Van Petegem; Brecht Houthoofd; Jan W Casselman
Journal:  Skeletal Radiol       Date:  2011-05-21       Impact factor: 2.199

7.  Use of dentomaxillofacial cone beam computed tomography in dentistry.

Authors:  Kıvanç Kamburoğlu
Journal:  World J Radiol       Date:  2015-06-28

8.  Radiologic and functional evaluation of electrode dislocation from the scala tympani to the scala vestibuli in patients with cochlear implants.

Authors:  N Fischer; L Pinggera; V Weichbold; D Dejaco; J Schmutzhard; G Widmann
Journal:  AJNR Am J Neuroradiol       Date:  2014-11-27       Impact factor: 3.825

9.  Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects.

Authors:  Ambrose J Huang; Connie Y Chang; Bijoy J Thomas; Peter J MacMahon; William E Palmer
Journal:  Skeletal Radiol       Date:  2015-02-05       Impact factor: 2.199

Review 10.  Radiation dose in non-dental cone beam CT applications: a systematic review.

Authors:  Cosimo Nardi; Sergio Salerno; Roberto Molteni; Mariaelena Occhipinti; Giulia Grazzini; Niccolò Norberti; Cesare Cordopatri; Stefano Colagrande
Journal:  Radiol Med       Date:  2018-06-05       Impact factor: 3.469