Literature DB >> 19166241

Quality assurance of an image guided intracranial stereotactic positioning system.

Emilie T Soisson1, Geoff Sobering, Daniel Lucas, Edward Chao, Gustavo Olivera, Wolfgang A Tomé.   

Abstract

This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.

Entities:  

Mesh:

Year:  2009        PMID: 19166241     DOI: 10.1177/153303460900800106

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  4 in total

1.  Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy.

Authors:  Emilie T Soisson; Nick Hardcastle; Wolfgang A Tomé
Journal:  J Neurooncol       Date:  2010-05-27       Impact factor: 4.130

2.  Radiosurgery technology development and use.

Authors:  Sanford L Meeks; Jason Pukala; Naren Ramakrishna; Twyla R Willoughby; Francis J Bova
Journal:  J Radiosurg SBRT       Date:  2011

3.  A technique for stereotactic radiosurgery treatment planning with helical tomotherapy.

Authors:  Emilie T Soisson; Peter W Hoban; Thomas Kammeyer; Jeffrey M Kapatoes; David C Westerly; Amar Basavatia; Wolfgang A Tomé
Journal:  Med Dosim       Date:  2011       Impact factor: 1.482

4.  Clinical commissioning of a new patient positioning system, SyncTraX FX4, for intracranial stereotactic radiotherapy.

Authors:  Satoshi Tanabe; Osamu Umetsu; Toshikazu Sasage; Satoru Utsunomiya; Ryota Kuwabara; Toshiki Kuribayashi; Hiromasa Takatou; Gen Kawaguchi; Hidefumi Aoyama
Journal:  J Appl Clin Med Phys       Date:  2018-10-01       Impact factor: 2.102

  4 in total