Repeatability and interdevice reliability of two portable color selection devices in matching and measuring tooth color.

Panagiotis E Lagouvardos1, Aggeliki G Fougia, Sofia A Diamantopoulou, Gregory L Polyzois.   

Abstract

STATEMENT OF PROBLEM: There is a need to know how shade selection devices perform in matching and measuring tooth color, since these functions are usually evaluated independently and may present significant discrepancies.
PURPOSE: The purpose of this study was to evaluate 2 devices which offer both functions to test the null hypothesis that they present no differences in their repeatability and interdevice reliability relating to the 2 functions.
MATERIAL AND METHODS: Thirty-one extracted anterior human teeth were measured twice, with each of the devices (ShadeEye NCC and VITA EasyShade), by one investigator experienced and calibrated with both devices. L*a*b* values and shade matches to VITA Classical and Vitapan 3D-Master shade guide systems were determined for all teeth. Paired t tests and intraclass correlation coefficients (ICCs) were used to statistically analyze the data (alpha=.05).
RESULTS: The results showed that L*a*b* values with VITA EasyShade were significantly higher than those with ShadeEye NCC (P<.01), with a mean difference of 10.01 Delta E units. Device repeatability ICCs for measuring tooth color ranged from 0.693 to 0.929, with significant differences between devices only with respect to parameter L* (P=.042). Intradevice ICCs for matching tooth color ranged from 0.714 to 0.894, with significant differences between devices only with respect to the VITA Classical shade system (P=.042). Interdevice ICCs for measuring tooth color ranged from 0.402 to 0.486, with no significant differences between L*a*b* parameters, while interdevice matching reliability ranged from 0.189 to 0.430, with no significant differences between shade systems (P>.05).
CONCLUSIONS: All tooth color parameters were higher with the VITA EasyShade device. Measuring repeatability of the VITA EasyShade was higher than the ShadeEye NCC only for the L* parameter. Matching repeatability of the ShadeEye NCC was higher than the VITA EasyShade for the VITA Classical system. Interdevice measuring reliabilities were not different for the color parameters, but matching reliability of the VITA Classical system was higher than that of the Vitapan 3D-Master.

Entities:  

Mesh:

Year:  2009        PMID: 19105990     DOI: 10.1016/S0022-3913(08)60289-9

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  8 in total

1.  Comparison of the Easyshade Compact and Advance in vitro and in vivo.

Authors:  Andreas Zenthöfer; Tomas Cabrera; Nicoleta Corcodel; Peter Rammelsberg; Alexander J Hassel
Journal:  Clin Oral Investig       Date:  2013-10-08       Impact factor: 3.573

2.  Interdevice agreement of eight equivalent dental color measurement devices.

Authors:  M Weyhrauch; C Igiel; A M Pabst; S Wentaschek; H Scheller; K M Lehmann
Journal:  Clin Oral Investig       Date:  2015-03-24       Impact factor: 3.573

3.  The repeatability of digital shade measurement--a clinical study.

Authors:  Constanze Olms; Jürgen M Setz
Journal:  Clin Oral Investig       Date:  2012-07-25       Impact factor: 3.573

4.  A novel regression model from RGB image data to spectroradiometric correlates optimized for tooth colored shades.

Authors:  Melody N Carney; William M Johnston
Journal:  J Dent       Date:  2016-05-31       Impact factor: 4.379

5.  In Vivo and in Vitro Evaluations of Repeatability and Accuracy of VITA Easyshade® Advance 4.0 Dental Shade-Matching Device.

Authors:  Dubravka Knezović; Davor Zlatarić; Iva Ž Illeš; Maja Alajbeg
Journal:  Acta Stomatol Croat       Date:  2015-06

6.  Evaluation of accuracy and characteristics of tooth-color matching by intraoral scanners based on Munsell color system: an in vivo study.

Authors:  Mingming Huang; Hongqiang Ye; Hu Chen; Yongsheng Zhou; Yunsong Liu; Yong Wang; Yuchun Sun
Journal:  Odontology       Date:  2022-02-25       Impact factor: 2.885

7.  Study of the most frequent natural tooth colors in the Spanish population using spectrophotometry.

Authors:  Cristina Gómez-Polo; Miguel Gómez-Polo; Juan Antonio Martínez Vázquez de Parga; Alicia Celemín Viñuela
Journal:  J Adv Prosthodont       Date:  2015-12-30       Impact factor: 1.904

8.  Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study.

Authors:  Hee-Kyung Kim
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.