In vitro evaluation of root canal preparation using oscillatory and rotary systems in flattened root canals.

Luis Cardoso Rasquin1, Fabíola Bastos de Carvalho, Regina Karla de Pontes Lima.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the biomechanical preparation of flattened root canals using the following systems: Endo-Eze AET stainless steel oscillatory instruments (Ultradent) and RaCe rotary NiTi instruments (FKG Dentaire).
MATERIAL AND METHODS: Twenty extracted human mandibular incisors were randomly assigned to two groups: Group 1 - Instrumentation with oscillatory Endo-Eze AET files (oscillatory technique); Group 2 - Instrumentation with rotary NiTi RaCe files (rotary technique). The teeth were decoronated, had their apices and coronal openings sealed with sticky wax and were embedded in crystal-clear orthophtalic polyester resin. The roots were sectioned transversally with diamond discs at 10 mm (middle third) and 5 mm (apical third) from the apex and the segments were reassembled for instrumentation. The sections were photographed before and after root canal instrumentation and evaluated with respect to whether the original root canal shape was modified by instrumentation. To evaluate the differences in the root canal shape before and after biomechanical preparation, scores were given regarding the instruments touch on the intracanal walls.
RESULTS: In middle third of the root canals instrumented with the rotary system, there was a change in the original canal anatomy (p<0.05), with formation of a protuberance in the mesiodistal direction. This protuberance did not occur when the oscillatory instrumentation was used. The oscillatory system had better results in the middle and apical thirds as evaluated by Dunn's multiple-comparison test (p>0.05).
CONCLUSION: Under the tested conditions, Endo-Eze oscillatory system yielded the instrumentation of all flattened root canal walls, maintaining the canal original shape throughout the biomechanical preparation, and was more effective than RaCe rotary system.

Entities:  

Year:  2007        PMID: 19089103      PMCID: PMC4327215          DOI: 10.1590/s1678-77572007000100014

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

The basic objective of biomechanical preparation consists of pulp tissue removal keeping the apical foramen as tight as possible, developing a gradual tapered shape of the root in a crown-apex direction, with the use of endodontic instruments associated with chemical irrigating substances6. The internal anatomic variations of the root canal system have a role on the success of endodontic therapy. In flattened canals, tissue remnants can persist in recesses and ramifications9. According to Wu, et al.16 (2000) the buccal and lingual extensions of flattened root canals are areas of difficult instrumentation and filling. The success of root canal therapy depends directly on how effectively all phases of the endodontic treatment (i.e., shaping, cleaning, disinfection and obturation) are performed. Several techniques have been used for biomechanical preparation. However, due to the limitations of conventional techniques in working in curved and flattened canals, some crown-down techniques have been recently investigated in order to provide a more efficient canal preparation. Complete biomechanical preparation with stainless steel instruments involves a high risk of perforating or significantly weakening the root. On the other hand, it seems questionable whether highly flexible Ni-Ti instruments allow a controlled preparation of canal extensions12. Ni-Ti rotary instrumentation has been shown to have a limited area of action. Due to their super-elasticity, it is known that NiTi instruments cannot be pressed against the root canal walls, leaving the polar areas of flattened root canals uninstrumented. It is likely that pathogenic microorganisms remain lodged in these areas. Flattened root canals do not allow a great widening of canal diameter without the risk of creating lateral perforations. A change of instrumentation techniques should also be considered1. The oscillatory movement or the movement of balanced forces has presented superior results to the filing movements. This movement can be manually executed or can be done using special handpieces that currently have a prominent place in endodontic practice causing a lower incidence of canal transportation and better centralization of the prepared root canal3,7,11,13. The Endo-Eze AET oscillatory system (Anatomic Endodontic Technology – Ultradent, South Jordan, UT, USA) has been recently released to the market with the objective of yielding cleaning and shaping of root canals using the root canal anatomy as a guide and hence preserving the dental structure. This system is comprised of a handpiece that has right-left oscillatory movements at 30º amplitude, acting in conjunction with stainless steel files14. The purpose of this study was to evaluate the biomechanical preparation of flattened root canals using two different systems: Endo-Eze AET oscillatory system (Ultradent) and RaCe rotary system (FKG Dentaire, La-Chaux-de-Fonds, Switzerland).

MATERIAL AND METHODS

Sample

The research protocol was reviewed by the Ethics in Research Committee of the State University of Feira de Santana (BA, Brazil) and the study design was approved (CAAE – 0035.0.059.000-05). Twenty human mandibular incisors were used in this study. The teeth had a buccolingual distance twice as long as the mesiodistal distance, confirmed by clinical evaluation of the shape of the flattened mesiodistal root, and had been extracted for periodontal reasons and/or extensive caries. The teeth were randomly assigned to two groups: Group 1 - Instrumentation with oscillatory Endo-Eze AET files (oscillatory technique); Group 2 - Instrumentation with rotary NiTi RaCe files (rotary technique).

In vitro Evaluation Preparation of teeth

The teeth were decoronated with a 40-mm diamond disc (KG Sorensen, São Paulo, SP, Brazil) at the cementoenamel junction. A #10 stainless steel K-file (Maillefer, Ballaigues, Switzerland) was then passively introduced into the canal until its tip was seen at the apical foramen. The instrument was retracted 1 mm to determine the working length (WL) by the visual method. Mesiodistal and buccolingual radiographs were taken to confirm the existence of only one root canal. The teeth had the apices and coronal openings sealed with sticky wax and were placed inside a PVC ring (25 mm in diameter; 25 mm in height) that was filled with crystal-clear orthophtalic polyester resin (Polyfibras - Bahia - Brazil) at a ratio of 10% monomer per kilogram of resin. The PVC ring had its base perforated to allow the placement of a thread and a screw (FischerÒ MQR 19, Rio de Janeiro, RJ, Brazil) with the purpose of facilitating the removal of the sections for undertaking photographs during the experiment2 (Figures 1a and 1b). The roots were sectioned transversally with diamond discs (Discoflex 40 mm) at 10 mm (middle third) and 5 mm (apical third) from the apex and the segments were reassembled for instrumentation.
FIGURE 1

Frontal view of the PVC ring (a) and PVC with its base perforated (b)

Instrumentation Technique

In Group 1, the teeth were instrumented using Endo-Eze AET oscillatory system (Ultradent) with Endo-Eze files. The following sequence of instruments belonging to the “Shaping Files” set was used. First, the oscillatory S1 file (yellow shank, D0 = 10 mm, taper 2.5%) was used at WL until it became loose into the canal. This file was followed by the SC instrument (red shank, D0 = 13 mm, taper 3.5%), the S2 instrument (blue shank, D0 = 13 mm, taper 4.5%) and the S3 instrument (green shank, D0 = 13mm, taper 6%). After using the oscillatory files, the apical third was instrumented using manual files belonging to “Apical Files” set of the system (#15 to #30). In Group 2, the teeth were instrumented using RaCe rotary system (FKG Dentaire) according to the crown-down technique in the following sequence: Pre-RaCe 40.08, RaCe 35.02, 30.04, 30.02, 25.04 and 25.02. Apical preparation was completed with RaCe 25.04 or 30.02 files, depending on root canal anatomy. In this group, all root canals were prepared using a low-speed handpiece (300 rpm) with high-torque motor with torque control (Easy Endo; Easy, Belo Horizonte, MG, Brazill). The root canal sections were photographed before and after biomechanical preparation. During the instrumentation, the root canals of both groups were irrigated with 5.25% sodium hypochlorite.

Evaluation of root canal deviation and shape

Using superimposition of images obtained from magnified and standardized photographs taken before and after biomechanical preparation, the middle and apical thirds were evaluated to determine whether there were changes on root canal original anatomical shape. In order to evaluate the differences between the root canal contours before and after biomechanical preparation, scores were given according to instrument touch on the intracanal walls (Table 1).
Table 1

Scores attributed to instrument touching on the root canal walls

Score 4 Touch of the instrument on the four root canal walls
Score 3 Touch of the instrument on three root canal walls
Score 2 Touch of the instrument on two root canal walls
Score 1 Touch of the instrument on one wall of the root canal
Score 0 The instrument did not touch any of the root canal walls

Statistical analysis

The score means were submitted to statistical analysis using the Kruskal-Wallis test (p<0.05) and Dunn’s multiple comparison test.

RESULTS

Table 2 gives a general overview of the scores attributed after analysis of the photographs taken before and after biomechanical preparation with both tested techniques. In Group 1 (oscillatory technique), one section changed from the original root canal shape in the middle third, that is the instrumentation touched only on one wall (score 1), while in the apical third two sections did not maintain the original root canal shape (one section with score 2 and the other with score 1). In Group 2 (rotary technique), six sections had score 2 (the biomechanical preparation produced a circular protuberance in the root canals leaving uninstrumented the isthmus walls) and three sections had score 1 in the middle third (Figure 2). In the apical third, five sections presented change in the original anatomy of the root canal (two sections with score 2 and three sections with score 1) (Figure 3).
Table 2

General distribution of scores

Score 4Score 3Score 2Score 1Score 0
Oscillatory techniqueMiddle Third9* 0010
Oscillatory techniqueApical Third80110
Rotary techniqueMiddle Third10630p< 0.05
Rotary techniqueApical Third50230

Number of sections

FIGURE 2

Middle third before (a) and after (b) preparation with RaCe system (Score 2)

FIGURE 3

Apical third before (a) and after (b) preparation with RaCe system (Score 2)

Number of sections In 7 out of 10 mandibular incisors instrumented with the oscillatory system, the biomechanical preparation maintained the original root canal shape up to the working length (Figure 4 and 5). On the other hand, only 1 out of 10 mandibular incisors instrumented with the rotary system maintained the original root canal shape up to the working length.
FIGURE 4

Middle third before (a) and after (b) preparation with Endo-Eze system (Score 4)

FIGURE 5

Apical third before (a) and after (b) preparation with Endo-Eze system (Score 4)

The statistical analysis showed that the middle third of the root canals instrumented with the rotary system presented a statistically significant change from the original anatomy (p<0.05), a fact that was not observed in the apical third. With the oscillatory instrumentation, the changes that occurred in the original shape of the canals were not statistically significant.

DISCUSSION

Root canal instrumentation is one of the most important steps to the success of endodontic treatment and several biomechanical instrumentation techniques have been proposed. However, despite the limitations of the conventional techniques in working in curved and flattened canals, crown-down techniques have been recently proposed with the aim of achieving a more efficient preparation. The quality of the biomechanical preparation is related to the internal anatomical characteristics of the root canal system1. Manual crown-down instrumentation techniques have been more effective in cleaning mesiodistally flattened root canals than rotary techniques. In this study, the root canals that had mesiodistal flattening and were instrumented with the RaCe rotary system presented changes in their original anatomy with the formation of a protuberance in the mesiodistal direction, leaving the walls of the isthmus uninstrumented. This showed that the rotary instrumentation had a limited working area, in according with previous studies5,10. The flexibility of the NiTi instruments does not allow a controlled instrumentation of the buccal and lingual extensions of flattened canals. These instruments frequently produce a circular protuberance in the root canals, maintaining the extensions uninstrumented5,12. Wu and Wesselink15 (2001) also demonstrated that, after biomechanical preparation of mandibular incisors with oval root canals using balanced force, 65% of the canals presented non-instrumented isthmuses. In the present study, the group instrumented with the rotary system had better results in the apical third, with only two sections having non-instrumented buccal and lingual extensions (isthmuses) versus six middle third sections This may be attributed to the fact that most flattened root canals become more round towards the apex, as previously shown in a study with extracted teeth16. Several automated systems that connect different types of files to different handpieces are being introduced to Endodontics. Amongst these are the alternating systems that use oscillatory movements, which have been described as the best type of movement for biomechanical preparation of flattened root canals8. Endo-Eze instruments have recently been introduced to the market and, according to its manufacturer, the use of stainless steel instruments attached to an oscillatory handpiece allows the preparation of oval anatomical irregularities of root canals, thus cleaning the intracanal space without weakening the root. In this study, the oscillatory system (Endo-Eze) was used in flattened root canals to allow comparisons with the data of previous studies1,10,12, which carried out similar studies under the same conditions using similar techniques, but used rotary NiTi instead. The present study evaluated the amount of uninstrumented root canal area after biomechanical preparation using the oscillatory technique. Our results were better than those of previous studies that used rotary instrumentation in flattened root canals and found that the rotary instrumentation was not capable of reaching all the root canal walls1,10,12. In this study, the number of sections that maintained the original root canal shape when using the oscillatory technique in middle third was statistically significant (9 sections), if compared to the rotary technique (section 1) (Table 2). In the apical third, both techniques maintained the original root canal shape. This result disagree with those of Camargo4 (2004), according to whom Endo-Eze instrumentation caused greater displacement of root canal center than the RaCe instrumentation in the apical third of artificial premolars.

CONCLUSION

Under the tested conditions, Endo-Eze oscillatory system yielded the instrumentation of all flattened root canal walls, maintaining the canal original shape throughout the biomechanical preparation, and was more effective than RaCe rotary system.
  11 in total

1.  A primary observation on the preparation and obturation of oval canals.

Authors:  M K Wu; P R Wesselink
Journal:  Int Endod J       Date:  2001-03       Impact factor: 5.264

2.  Prevalence and extent of long oval canals in the apical third.

Authors:  M K Wu; A R'oris; D Barkis; P R Wesselink
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2000-06

3.  Effectiveness of manual and rotary instrumentation techniques for cleaning flattened root canals.

Authors:  João Vicente Baroni Barbizam; Luis Fernando Fariniuk; Melissa Andréia Marchesan; Jesus Djalma Pecora; Manoel D Sousa-Neto
Journal:  J Endod       Date:  2002-05       Impact factor: 4.171

4.  Quality of preparation of oval distal root canals in mandibular molars using nickel-titanium instruments.

Authors:  T Rödig; M Hülsmann; M Mühge; F Schäfers
Journal:  Int Endod J       Date:  2002-11       Impact factor: 5.264

5.  Comparison of root canal preparation using RaCe and ProTaper rotary Ni-Ti instruments.

Authors:  F Paqué; U Musch; M Hülsmann
Journal:  Int Endod J       Date:  2005-01       Impact factor: 5.264

6.  A comparison of maintenance of canal curvature using balanced-force instrumentation with three different file types.

Authors:  J R Royal; J C Donnelly
Journal:  J Endod       Date:  1995-06       Impact factor: 4.171

7.  Morphometrical analysis of cleaning capacity using nickel-titanium rotary instrumentation associated with irrigating solutions in mesio-distal flattened root canals.

Authors:  Melissa Andréia Marchesan; Marcos Porto de Arruda; Yara T Correa Silva-Sousa; Paulo César Saquy; Jesus Djalma Pecora; Manoel D Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2003-03       Impact factor: 2.698

8.  A methodology for evaluation of root canal instrumentation.

Authors:  C M Bramante; A Berbert; R P Borges
Journal:  J Endod       Date:  1987-05       Impact factor: 4.171

9.  The effects of four instrumentation techniques on root canal shape.

Authors:  G Calhoun; S Montgomery
Journal:  J Endod       Date:  1988-06       Impact factor: 4.171

10.  The "balanced force" concept for instrumentation of curved canals.

Authors:  J B Roane; C L Sabala; M G Duncanson
Journal:  J Endod       Date:  1985-05       Impact factor: 4.171

View more
  2 in total

1.  Effect of passive ultrasonic agitation during final irrigation on cleaning capacity of hybrid instrumentation.

Authors:  Marcilene Coelho Vinhorte; Eduardo Hideki Suzuki; Maíra Sousa de Carvalho; André Augusto Franco Marques; Emílio Carlos Sponchiado Júnior; Lucas da Fonseca Roberti Garcia
Journal:  Restor Dent Endod       Date:  2014-03-21

2.  Cleaning capacity promoted by motor-driven or manual instrumentation using ProTaper Universal system: Histological analysis.

Authors:  Matheus Franco da Frota; Idomeo Bonetti Filho; Fábio Luiz Camargo Villela Berbert; Emilio Carlos Sponchiado; André Augusto Franco Marques; Lucas da Fonseca Roberti Garcia
Journal:  J Conserv Dent       Date:  2013-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.