The Visian myopic implantable collamer lens does not significantly affect axial length measurement with the IOLMaster.

Donald R Sanders1, David A Bernitsky, Paul J Harton, Robert R Rivera.   

Abstract

PURPOSE: To compare pre- to postoperative axial length measurements with the IOLMaster (Zeiss Humphrey Systems) to determine whether a correction factor for later intraocular lens (IOL) power calculation is necessary with an Implantable Collamer Lens (ICL; STAAR Surgical) in place.
METHODS: The commercially approved Visian ICL was implanted in 60 eyes from 3 separate sites (20 eyes each). Pre- and postoperative axial length measurements were obtained using partial coherence interferometry.
RESULTS: The pre- and postoperative axial lengths were highly correlated with an R2 value of 0.9963. For all ranges of axial lengths tested, the use of postoperative axial length instead of preoperative axial length resulted in a <0.1-mm error in axial length measurements.
CONCLUSIONS: Based on the SRK formula, a 0.1-mm axial length difference would result in a 0.25-diopter (D) change in calculated IOL power; possibly a 0.20-D change in manifest refraction. These differences are so small as to make any correction to IOL power calculation formula based on the presence of an ICL unnecessary.

Entities:  

Mesh:

Year:  2008        PMID: 19044239     DOI: 10.3928/1081597X-20081101-17

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  6 in total

1.  Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry.

Authors:  Çağatay Çağlar; Sücattin İlker Kocamış; Emre Demir; Mustafa Durmuş
Journal:  Int Ophthalmol       Date:  2016-07-08       Impact factor: 2.031

2.  Evaluation of impact of posterior phakic IOL implantation on biometry and effectiveness of concomitant use of anterior segment OCT on IOL power calculation for cataract surgery.

Authors:  Masayuki Ouchi
Journal:  J Cataract Refract Surg       Date:  2021-09-01       Impact factor: 3.528

3.  Comparison of preoperative and postoperative ocular biometry in eyes with phakic intraocular lens implantations.

Authors:  Joo Youn Shin; Jae Bum Lee; Kyoung Yul Seo; Eung Kweon Kim; Tae-im Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

4.  Long-term refractive outcomes of posterior chamber phakic (spheric and toric implantable collamer lens) intraocular lens implantation.

Authors:  Arturo Gomez-Bastar; Martha Jaimes; Enrique O Graue-Hernández; Tito Ramirez-Luquin; Arturo Ramirez-Miranda; Alejandro Navas
Journal:  Int Ophthalmol       Date:  2013-10-11       Impact factor: 2.031

5.  Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry.

Authors:  Tomoaki Higashiyama; Hazuki Mori; Fumi Nakajima; Masahito Ohji
Journal:  PLoS One       Date:  2018-04-24       Impact factor: 3.240

6.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909