Literature DB >> 19029615

Comparison of the load-sharing characteristics between pedicle-based dynamic and rigid rod devices.

Yoon-Ho Ahn1, Wen-Ming Chen, Kwon-Yong Lee, Kyung-Woo Park, Sung-Jae Lee.   

Abstract

Recently, numerous types of posterior dynamic stabilization (PDS) devices have been introduced as an alternative to the fusion devices for the surgical treatment of degenerative lumbar spine. It is hypothesized that the use of 'compliant' materials such as Nitinol (Ni-Ti alloy, elastic modulus = 75 GPa) or polyether-etherketone (PEEK, elastic modulus = 3.2 GPa) in PDS can restore stability of the lumbar spine without adverse stress-shielding effects that have often been found with 'rigid' fusion devices made of 'rigid' Ti alloys (elastic modulus = 114 GPa). Previous studies have shown that suitably designed PDS devices made of more compliant material may be able to help retain kinematic behavior of the normal spine with optimal load sharing between the anterior and posterior spinal elements. However, only a few studies on their biomechanical efficacies are available. In this study, we conducted a finite-element (FE) study to investigate changes in load-sharing characteristics of PDS devices. The implanted models were constructed after modifying the previously validated intact model of L3-4 spine. Posterior lumbar fusion with three different types of pedicle screw systems was simulated: a conventional rigid fixation system (Ti6Al4V, Phi = 6.0 mm) and two kinds of PDS devices (one with Nitinol rod with a three-coiled turn manner, Phi = 4.0 mm; the other with PEEK rod with a uniform cylindrical shape, Phi = 6.0 mm). To simulate the load on the lumbar spine in a neutral posture, an axial compressive load (400 N) was applied. Subsequently, the changes in load-sharing characteristics and stresses were investigated. When the compressive load was applied on the implanted models (Nitinol rod, PEEK rod, Ti-alloy rod), the predicted axial compressive loads transmitted through the devices were 141.8 N, 109.8 N and 266.8 N, respectively. Axial forces across the PDS devices (Nitinol rod, PEEK rod) and rigid system (Ti-alloy rod) with facet joints were predicted to take over 41%, 33% and 71% of the applied compression load, respectively. Our results confirmed the hypothesis on the PDS devices by showing the substantial reduction in stress-shielding characteristics. Higher axial load was noted across the anterior structure with the PDS devices, which could slow the degeneration process of bony structures and lower the possibility of implant failure.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19029615     DOI: 10.1088/1748-6041/3/4/044101

Source DB:  PubMed          Journal:  Biomed Mater        ISSN: 1748-6041            Impact factor:   3.715


  26 in total

1.  Optimal stiffness of a pedicle-screw-based motion preservation implant for the lumbar spine.

Authors:  Antonius Rohlmann; Thomas Zander; Georg Bergmann; Hadi N Boustani
Journal:  Eur Spine J       Date:  2011-10-20       Impact factor: 3.134

2.  Clinical and Radiological Comparison of Semirigid (WavefleX) and Rigid System for the Lumbar Spine.

Authors:  Do-Keun Kim; Hyunkeun Lim; Dae Cheol Rim; Chang Hyun Oh
Journal:  Korean J Spine       Date:  2016-06-30

Review 3.  Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review.

Authors:  Chan Li; Lei Liu; Jian-Yong Shi; Kai-Zhong Yan; Wei-Zhong Shen; Zhen-Rong Yang
Journal:  Neurosurg Rev       Date:  2016-07-08       Impact factor: 3.042

4.  Characterization of the behavior of a novel low-stiffness posterior spinal implant under anterior shear loading on a degenerative spinal model.

Authors:  Angela D Melnyk; Jason D Chak; Vaneet Singh; Adrienne Kelly; Peter A Cripton; Charles G Fisher; Marcel F Dvorak; Thomas R Oxland
Journal:  Eur Spine J       Date:  2015-01-06       Impact factor: 3.134

5.  Biomechanical analysis of lumbar interbody fusion supplemented with various posterior stabilization systems.

Authors:  Wei Fan; Li-Xin Guo; Ming Zhang
Journal:  Eur Spine J       Date:  2021-05-04       Impact factor: 3.134

6.  Pedicle screw anchorage of carbon fiber-reinforced PEEK screws under cyclic loading.

Authors:  Richard A Lindtner; Rene Schmid; Thomas Nydegger; Marko Konschake; Werner Schmoelz
Journal:  Eur Spine J       Date:  2018-03-01       Impact factor: 3.134

7.  Stress analysis of the implants in transforaminal lumbar interbody fusion under static and vibration loadings: a comparison between pedicle screw fixation system with rigid and flexible rods.

Authors:  Wei Fan; Li-Xin Guo; Dan Zhao
Journal:  J Mater Sci Mater Med       Date:  2019-10-18       Impact factor: 3.896

8.  Multiple-rod constructs in adult spinal deformity surgery for pelvic-fixated long instrumentations: an integral matched cohort analysis.

Authors:  Fernando Guevara-Villazón; Louis Boissiere; Kazunori Hayashi; Daniel Larrieu; Soufiane Ghailane; Jean-Marc Vital; Olivier Gille; Vincent Pointillart; Ibrahim Obeid; Anouar Bourghli
Journal:  Eur Spine J       Date:  2020-01-28       Impact factor: 3.134

9.  Effect of Device Rigidity and Physiological Loading on Spinal Kinematics after Dynamic Stabilization : An In-Vitro Biomechanical Study.

Authors:  Kwonsoo Chun; Inchul Yang; Namhoon Kim; Dosang Cho
Journal:  J Korean Neurosurg Soc       Date:  2015-11-30

10.  Change of lumbar motion after multi-level posterior dynamic stabilization with bioflex system : 1 year follow up.

Authors:  Hunho Park; Ho-Yeol Zhang; Bo Young Cho; Jeong Yoon Park
Journal:  J Korean Neurosurg Soc       Date:  2009-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.