Literature DB >> 18997635

Foreign body or hypersensitivity granuloma of the inner ear after cochlear implantation: one possible cause of a soft failure?

Joseph B Nadol1, Donald K Eddington, Barbara J Burgess.   

Abstract

HYPOTHESIS: A tissue response in the form of foreign body or a hypersensitivity reaction to cochlear implantation is common and may be one possible cause of a soft failure of cochlear implantation.
BACKGROUND: After a successful cochlear implantation, delayed failure may occur. The causes of a "soft" failure, that is, one in which device malfunction cannot be proven, are unknown.
METHODS: The histopathology of the temporal bones of a patient who, in life, had experienced a soft failure after cochlear implantation was described. In addition, the temporal bones of 8 other subjects who, in life, had undergone cochlear implantation were studied for evidence of a foreign body or hypersensitivity reaction.
RESULTS: In the case report, a necrotizing granulomatous giant cell reaction surrounded the cochlear implant electrode track through the mastoid and the middle ear and into the cochlea in both ears. There was osteolysis of the cribrose area, otic capsule, and bone between the facial nerve and the cochlea and destruction of the organ of Corti and spiral ganglion. In the additional 8 cases studied, a similar, although less pronounced, foreign body or hypersensitivity reaction was seen in 6 (75%) of the cases.
CONCLUSION: A foreign body or hypersensitivity reaction in the form of giant cells and lymphocytic cell infiltration is common after cochlear implantation and may be one possible cause of soft failure.

Entities:  

Mesh:

Year:  2008        PMID: 18997635      PMCID: PMC4180287          DOI: 10.1097/MAO.0b013e31818c33cf

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  23 in total

1.  Re: "Silicone allergy: a new cause for cochlear implant extrusion and its management".

Authors:  Paal C Klykken; James M Curtis
Journal:  Otol Neurotol       Date:  2007-12       Impact factor: 2.311

2.  Unexpected exit of a cochlear implant electrode through the wall of the basal turn of the cochlea - a report on two patients.

Authors:  Eu Chin Ho; David Proops; Peter Andrews; John Graham
Journal:  Cochlear Implants Int       Date:  2007-09

3.  Long-term results of silicone wrist arthroplasty in patients with rheumatoid arthritis.

Authors:  Urs Kistler; Arnold-Peter C Weiss; Beat R Simmen; Daniel B Herren
Journal:  J Hand Surg Am       Date:  2005-11       Impact factor: 2.230

4.  Endolymphatic subarachnoid shunt failure caused by Silastic allergy.

Authors:  J L Pulec
Journal:  Ear Nose Throat J       Date:  1998-08       Impact factor: 1.697

Review 5.  Silicone toxicology.

Authors:  H Busch
Journal:  Semin Arthritis Rheum       Date:  1994-08       Impact factor: 5.532

Review 6.  Silicone allergy: A new cause for cochlear implant extrusion and its management.

Authors:  Larisa D Kunda; Katrina R Stidham; Michelle M Inserra; Peter S Roland; Daniel Franklin; Joseph B Roberson
Journal:  Otol Neurotol       Date:  2006-12       Impact factor: 2.311

7.  Revision cochlear implant surgery in adult patients with suspected device malfunction.

Authors:  Craig A Buchman; Carol A Higgins; Robert Cullen; Harold C Pillsbury
Journal:  Otol Neurotol       Date:  2004-07       Impact factor: 2.311

8.  Hypersensitivity reaction associated with sterile ventriculoperitoneal shunt malfunction.

Authors:  R B Snow; N Kossovsky
Journal:  Surg Neurol       Date:  1989-03

9.  A case of abdominal CSF pseudocyst associated with silicone allergy.

Authors:  Masanori Hashimoto; Akira Yokota; Eiichirou Urasaki; Shuhji Tsujigami; Masayuki Shimono
Journal:  Childs Nerv Syst       Date:  2004-03-04       Impact factor: 1.475

10.  Histologic evaluation of the tissue seal and biologic response around cochlear implant electrodes in the human.

Authors:  Joseph B Nadol; Donald K Eddington
Journal:  Otol Neurotol       Date:  2004-05       Impact factor: 2.311

View more
  20 in total

1.  Factors associated with incomplete insertion of electrodes in cochlear implant surgery: a histopathologic study.

Authors:  Joonhan Lee; Joseph B Nadol; Donald K Eddington
Journal:  Audiol Neurootol       Date:  2010-06-19       Impact factor: 1.854

2.  The pattern and degree of capsular fibrous sheaths surrounding cochlear electrode arrays.

Authors:  Reuven Ishai; Barbara S Herrmann; Joseph B Nadol; Alicia M Quesnel
Journal:  Hear Res       Date:  2017-02-17       Impact factor: 3.208

3.  Intracochlear inflammatory response to cochlear implant electrodes in humans.

Authors:  Mohammad Seyyedi; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2014-10       Impact factor: 2.311

4.  Correlation between word recognition score and intracochlear new bone and fibrous tissue after cochlear implantation in the human.

Authors:  Takefumi Kamakura; Joseph B Nadol
Journal:  Hear Res       Date:  2016-06-29       Impact factor: 3.208

5.  Post Hybrid Cochlear Implant Hearing Loss and Endolymphatic Hydrops.

Authors:  Akira Ishiyama; Joni Doherty; Gail Ishiyama; Alicia M Quesnel; Ivan Lopez; Fred H Linthicum
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311

6.  Foreign Body Response to Silicone in Cochlear Implant Electrodes in the Human.

Authors:  Jennifer T O'Malley; Barbara J Burgess; Donald Galler; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2017-08       Impact factor: 2.311

7.  Cochlear Implant Electrode Migration due to Cholesterol Granuloma: Cues from a Case.

Authors:  Andrea Di Laora; Isabelle Mosnier; Andrea Ciorba; Stefano Pelucchi; Olivier Sterkers; Daniele Bernardeschi
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

8.  Histopathologic Changes of the Inner ear in Rhesus Monkeys After Intratympanic Gentamicin Injection and Vestibular Prosthesis Electrode Array Implantation.

Authors:  Daniel Q Sun; Mohamed Lehar; Chenkai Dai; Lani Swarthout; Amanda M Lauer; John P Carey; Diana E Mitchell; Kathleen E Cullen; Charles C Della Santina
Journal:  J Assoc Res Otolaryngol       Date:  2015-03-20

9.  Delayed changes in auditory status in cochlear implant users with preserved acoustic hearing.

Authors:  Rachel A Scheperle; Viral D Tejani; Julia K Omtvedt; Carolyn J Brown; Paul J Abbas; Marlan R Hansen; Bruce J Gantz; Jacob J Oleson; Marie V Ozanne
Journal:  Hear Res       Date:  2017-04-12       Impact factor: 3.208

10.  Foreign Body Reaction After Cochlear Implantation: A Case Report.

Authors:  Yuan Xin; Ya-Sheng Yuan; Fang-Lu Chi; Jing Wang; Juan-Mei Yang
Journal:  Chin Med J (Engl)       Date:  2015-08-05       Impact factor: 2.628

View more