| Literature DB >> 18986868 |
Abstract
This paper introduces a novel sidelobe and clutter suppression method in ultrasound imaging called dual apodization with cross-correlation or DAX. DAX dramatically improves the contrast-to-noise ratio (CNR) allowing for easier visualization of anechoic cysts and blood vessels. This technique uses dual apodization or weighting strategies that are effective in removing or minimizing clutter and efficient in terms of computational load and hardware/software needs. This dual apodization allows us to determine the amount of mainlobe versus clutter contribution in a signal by cross-correlating RF data acquired from 2 apodization functions. Simulation results using a 128 element 5 MHz linear array show an improvement in CNR of 139% compared with standard beamformed data with uniform apodization in a 3 mm diameter anechoic cylindrical cyst. Experimental CNR using a tissue-mimicking phantom with the same sized cyst shows an improvement of 123% in a DAX processed image.Entities:
Mesh:
Year: 2008 PMID: 18986868 PMCID: PMC2905597 DOI: 10.1109/TUFFC.919
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725