Literature DB >> 18958444

[Complications with all-inside devices used in reconstructive meniscal surgery].

P Wilmes1, O Lorbach, P Brogard, R Seil.   

Abstract

All-inside devices have become increasingly popular in reconstructive meniscal surgery since their introduction at the beginning of the 1990s. Although the latest clinical investigations show better results for conventional suture techniques, meniscal devices are an important alternative because of the low risk of neurovascular injury and the easy handling of the instruments. Over the years, many reports on specific complications related to all-inside devices have been published. Especially chondral injuries, implant loosening, device migration and capsular or neural irritations have been described. Furthermore, some authors reported on foreign body reactions and cystic granulomas after the use of meniscal fixation devices. However, there is no evidence for a higher infection rate or for specific infections after the use of intra-articular techniques. Clinical reports on complications along with biomechanical studies on meniscal repair devices have led to the enhancement of all-inside techniques through substantial modifications of established products as well as to the development of new implants. After reviewing the latest literature, the complication rate seems to be decreasing. In many ways, all-inside devices are an interesting alternative to conventional suture techniques. A precise knowledge of their potential complications and the pitfalls during surgery however is crucial to make a risk evaluation in the choice of the right technique for meniscal reconstruction.

Entities:  

Mesh:

Year:  2008        PMID: 18958444     DOI: 10.1007/s00132-008-1307-4

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  61 in total

1.  Broken arrow: a complication of meniscal repair.

Authors:  S J Calder; P T Myers
Journal:  Arthroscopy       Date:  1999-09       Impact factor: 4.772

2.  Meniscal repair devices.

Authors:  F A Barber; M A Herbert
Journal:  Arthroscopy       Date:  2000-09       Impact factor: 4.772

3.  Chondral injury following meniscal repair with a biodegradable implant.

Authors:  K Anderson; R G Marx; J Hannafin; R F Warren
Journal:  Arthroscopy       Date:  2000-10       Impact factor: 4.772

4.  Failure of a biodegradable meniscal arrow. A case report.

Authors:  M R Hutchinson; S A Ash
Journal:  Am J Sports Med       Date:  1999 Jan-Feb       Impact factor: 6.202

5.  Failure strength of a new meniscus arrow repair technique: biomechanical comparison with horizontal suture.

Authors:  P Albrecht-Olsen; T Lind; G Kristensen; B Falkenberg
Journal:  Arthroscopy       Date:  1997-04       Impact factor: 4.772

6.  Aseptic synovitis after meniscal repair using the biodegradable meniscus arrow.

Authors:  E K Song; K B Lee; T R Yoon
Journal:  Arthroscopy       Date:  2001-01       Impact factor: 4.772

7.  Meniscal repair using the FasT-Fix all-inside meniscal repair device.

Authors:  Andrew L Haas; Anthony A Schepsis; Joshua Hornstein; Cory M Edgar
Journal:  Arthroscopy       Date:  2005-02       Impact factor: 4.772

8.  Second-look arthroscopy with removal of bioabsorbable tacks.

Authors:  J A Lombardi
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2000-02

9.  Septic arthritis following arthroscopic meniscus repair: a cluster of three cases.

Authors:  F T Blevins; J Salgado; D C Wascher; F Koster
Journal:  Arthroscopy       Date:  1999 Jan-Feb       Impact factor: 4.772

10.  Foreign-body reaction and the course of osteolysis after polyglycolide implants for fracture fixation: experimental study in sheep.

Authors:  A Weiler; H J Helling; U Kirch; T K Zirbes; K E Rehm
Journal:  J Bone Joint Surg Br       Date:  1996-05
View more
  2 in total

1.  Outcome of All-Inside Second-Generation Meniscal Repair: Minimum Five-Year Follow-up.

Authors:  Ljiljana Bogunovic; Lisa M Kruse; Amanda K Haas; Laura J Huston; Rick W Wright
Journal:  J Bone Joint Surg Am       Date:  2014-08-06       Impact factor: 5.284

2.  You Can Repair More Tears Than You Think-Tricks for All-Inside Lateral Meniscal Repair With Nonabsorbable Sutures.

Authors:  Konrad Malinowski; Michał Ebisz; Adrian Góralczyk; Robert F LaPrade; Krzysztof Hermanowicz
Journal:  Arthrosc Tech       Date:  2020-06-18
  2 in total