Performance of neonatal ventilators: the effects of changes in resistance and compliance.

G Simbruner, G A Gregory.   

Abstract

The authors tested the ability of 6 neonatal ventilators (Baby bird, Bourns BP-200, Veriflo, Bourns LS, ServoVent, SI-75) to deliver a preset tidal volume with minimal pressures when the compliance and resistance of test lung were altered. The authors also studied the effects of gas leaks from the system and the effects of PEEP. When the ventilator cycling rates exceeded 28-43/min, gas trapping occurred despite compliances and resistance appropriate for a healthy term infant. With a normal compliance and an elevated resistance, gas trapping occurred above rates of 4-6/min. Under these same conditions (normal compliance, elevated resistance), only 24-46% of the preset tidal volume was delivered. The different ventilators caused the pressure within the test lung to exceed 5, 10, and 15 cm H2O for differing amounts of time (sec)/min.

Entities:  

Mesh:

Year:  1981        PMID: 6940689

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  6 in total

1.  Adult ICU ventilators to provide neonatal ventilation: a lung simulator study.

Authors:  Andrew D Marchese; Daniel Chipman; Pedro de la Oliva; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2008-10-18       Impact factor: 17.440

2.  Assessment of neonatal ventilator performances.

Authors:  P Jouvet; P Hubert; P H Jarreau; F Lofaso; M Cloup; A Harf
Journal:  Intensive Care Med       Date:  1995-09       Impact factor: 17.440

3.  [Indirect determination of the intrapulmonary pressure course in the lung model in high frequency ventilation].

Authors:  W Hultzsch; G Lipowsky
Journal:  Klin Wochenschr       Date:  1989-09-15

4.  Neonatal and pediatric intensive care--a new section in Intensive Care Medicine.

Authors:  G Simbruner; J Pfenninger
Journal:  Intensive Care Med       Date:  1993       Impact factor: 17.440

5.  High bias gas flows increase lung injury in the ventilated preterm lamb.

Authors:  Katinka P Bach; Carl A Kuschel; Stuart B Hooper; Jean Bertram; Sue McKnight; Shirley E Peachey; Valerie A Zahra; Sharon J Flecknoe; Mark H Oliver; Megan J Wallace; Frank H Bloomfield
Journal:  PLoS One       Date:  2012-10-08       Impact factor: 3.240

6.  Simulations for mechanical ventilation in children: review and future prospects.

Authors:  Olivier Flechelles; Annie Ho; Patrice Hernert; Guillaume Emeriaud; Nesrine Zaglam; Farida Cheriet; Philippe A Jouvet
Journal:  Crit Care Res Pract       Date:  2013-03-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.