| Literature DB >> 18923696 |
W Stevens1, G Sherman, R Downing, L M Parsons, C-Y Ou, S Crowley, G M Gershy-Damet, K Fransen, M Bulterys, L Lu, J Homsy, T Finkbeiner, J N Nkengasong.
Abstract
A two day meeting hosted by the World Health Organization (WHO) and the U.S. Centers for Disease Control and Prevention (CDC) was held in May 2006 in Entebbe, Uganda to review the laboratory performance of virologic molecular methods, particularly the Roche Amplicor DNA PCR version 1.5 assay, in the diagnosis of HIV-1 infection in infants. The meeting was attended by approximately 60 participants from 17 countries. Data on the performance and limitations of the HIV-1 DNA PCR assay from 9 African countries with high-burdens of HIV/AIDS were shared with respect to different settings and HIV- subtypes. A consensus statement on the use of the assay for early infant diagnosis was developed and areas of needed operational research were identified. In addition, consensus was reached on the usefulness of dried blood spot (DBS) specimens in childhood as a means for ensuring greater accessibility to serologic and virologic HIV testing for the paediatric population.Entities:
Keywords: HIV DNA PCR; HIV infection; early infant diagnosis; laboratory assays
Year: 2008 PMID: 18923696 PMCID: PMC2556199 DOI: 10.2174/1874613600802010017
Source DB: PubMed Journal: Open AIDS J ISSN: 1874-6136
Commercial Assays Can Potentially be Used for Diagnosis of HIV Infection in Infants and Young Children
| Source/Name of Assay | Sample Used | Assay/Target | Volume (in ml) | Cost Per Test ($) |
|---|---|---|---|---|
| Roche | Whole Blood | Qual/DNA | 0.5 | 35- 90 |
| Amplicor HIV-1 | ||||
| DNA v1.5 | ||||
| Roche | Plasma or serum | Quan/RNA | 0.2-0.5 | 20-50 |
| Amplicor HIV-1 Monitor v 1.5 manual | ||||
| RocheCobas Amplicor HIV-1 Monitor v 1.5 | Plasma or serum | Quan/RNA | 0.2-0.75 | 20-50 |
| Roche | Plasma or serum | Quan/RNA | 0.5-1 | 50 |
| TagMan 48 | ||||
| Bayer | Plasma or serum | Quan/RNA | 1 | 85-90 |
| Versant | ||||
| HIV-1 RNA | ||||
| Biomerieux NucleiSens EasyQ | Plasma or serum | Quan/RNA | 1 | 80 |
| Abbott | Plasma or serum | Quan/RNA | 0.2-1 | 20-70 |
| Real Time HIV | ||||
| CavidiExaVir Load | Plasma or serum | Quan/ | 1 | 13-15 |
| RT | ||||
| Perkin Elmer | Plasma or serum | Quan/ | 50 µl | 10 |
| Up24 antigen | p24 antigen |
*Qual- Qualitative and **Quan- Quantitative.