Antigoni Katsoulidou1, Maria Petrodaskalaki, Vana Sypsa, Eleni Papachristou, Cleo G Anastassopoulou, Panagiotis Gargalianos, Anastasia Karafoulidou, Marios Lazanas, Theodoros Kordossis, Anastasia Andoniadou, Angelos Hatzakis.
Abstract
The COBAS TaqMan HIV-1 test (Roche Diagnostics) was compared with the LCx HIV RNA quantitative assay (Abbott Laboratories), the Versant HIV-1 RNA 3.0 (bDNA) assay (Bayer) and the COBAS Amplicor HIV-1 Monitor v1.5 test (Roche Diagnostics), using plasma samples of various viral load levels from HIV-1-infected individuals. In the comparison of TaqMan with LCx, TaqMan identified as positive 77.5% of the 240 samples versus 72.1% identified by LCx assay, while their overall agreement was 94.6% and the quantitative results of samples that were positive by both methods were strongly correlated (r=0.91). Similarly, in the comparison of TaqMan with bDNA 3.0, both methods identified 76.3% of the 177 samples as positive, while their overall agreement was 95.5% and the quantitative results of samples that were positive by both methods were strongly correlated (r=0.95). Finally, in the comparison of TaqMan with Monitor v1.5, TaqMan identified 79.5% of the 156 samples as positive versus 80.1% identified by Monitor v1.5, while their overall agreement was 95.5% and the quantitative results of samples that were positive by both methods were strongly correlated (r=0.96). In conclusion, the new COBAS TaqMan HIV-1 test showed excellent agreement with other widely used commercially available tests for the quantitation of HIV-1 viral load.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16194575 DOI: 10.1016/j.jviromet.2005.07.014
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.014