Literature DB >> 18830731

Validation of the WristOx 3100 oximeter for the diagnosis of sleep apnea/hypopnea syndrome.

Carlos Alberto Nigro1, Silvia Aimaretti, Sergio Gonzalez, Edgardo Rhodius.   

Abstract

OBJECTIVE: To evaluate the diagnostic accuracy of the Nonin WristOx 3100 and its software (nVision 5.0) in patients with suspicion of sleep apnea/hypopnea syndrome (SAHS).
METHODS: All participants (168) had the oximetry and polysomnography simultaneously. The two recordings were interpreted blindly. The software calculated: adjusted O(2) desaturation index [ADI]-mean number of O(2) desaturation per hour of total recording analyzed time of > or = 2%, 3%, 4%, 5%, and 6% (ADI2, 3, 4, 5, and 6) and AT90-accumulated time at SO(2) < 90%. The ADI2, 3, 4, 5, and 6 and the AT90 cutoff points that better discriminated between subjects with or without SAHS arose from the receiver operating characteristic curve analysis. The sensitivity (S), specificity (E), and positive and negative likelihood ratio (LR+, LR-) for the different thresholds for ADI were calculated.
RESULTS: One hundred and fifty-four patients were included (119 men, mean age 51, median apnea/hypopnea index [AHI] 14, median body mass index [BMI] 28.3 kg/m(2)). The best cutoff points of ADI were: SAHS = AHI > or = 5: ADI2 > 19.3 (S 89%, E 94%, LR+ 15.5 LR- 0.11); SAHS =AHI > or = 10: ADI3 > 10.5 (S 88%, E 94%, LR+ 15 LR- 0.12); SAHS = AHI > or = 15: ADI3 > 13.4 (S 88%, E 90%, LR+ 8.9, LR- 0.14). AT90 had the lowest diagnosis accuracy. An ADI2 < or = 12.2 excluded SAHS (AHI > or = 5 and 10; S 100%, LR- 0) and ADI3 > 4.3 (AHI > or = 5 and 10) or 32 (AHI > or = 15) confirmed SAHS (E 100%).
CONCLUSIONS: A negative oximetry defined as ADI2 < or = 12.2 excluded SAHS defined as AHI > or = 5 or 10 with a sensitivity and negative likelihood ratio of 100% and 0%, respectively. Furthermore, a positive oximetry defined as an ADI3 > 32 (SAHS = AHI > or = 15) had a specificity of 100% to confirm the pathology.

Entities:  

Mesh:

Year:  2008        PMID: 18830731     DOI: 10.1007/s11325-008-0217-3

Source DB:  PubMed          Journal:  Sleep Breath        ISSN: 1520-9512            Impact factor:   2.816


  52 in total

1.  Usefulness of the oximetry test for the diagnosis of sleep apnea syndrome in Japan.

Authors:  Takayuki Takeda; Yoshihiro Nishimura; Miyako Satouchi; Hiroshi Kamiryo; Kaori Takenaka; Daisuke Kasai; Yoshiko Urata; Kazuyuki Kobayashi; Temiko Shimada; Sho Yoshimura; Teruaki Nishiuma; Mitsuhiro Yokoyama
Journal:  Am J Med Sci       Date:  2006-06       Impact factor: 2.378

2.  Hand-scoring of MESAM 4 recordings is more accurate than automatic analysis in screening for obstructive sleep apnoea.

Authors:  M Koziej; J K Cieślicki; K Gorzelak; P Sliwiński; J Zieliński
Journal:  Eur Respir J       Date:  1994-10       Impact factor: 16.671

3.  A limited diagnostic investigation for obstructive sleep apnea syndrome. Oximetry and static charge sensitive bed.

Authors:  E Svanborg; H Larsson; B Carlsson-Nordlander; R Pirskanen
Journal:  Chest       Date:  1990-12       Impact factor: 9.410

4.  Prediction of sleep-disordered breathing by unattended overnight oximetry.

Authors:  L G Olson; A Ambrogetti; S G Gyulay
Journal:  J Sleep Res       Date:  1999-03       Impact factor: 3.981

5.  [Role of nocturnal oximetry in screening for sleep apnea syndrome in pulmonary medicine. Study of 329 patients].

Authors:  L Lacassagne; A Didier; M Murris-Espin; J P Charlet; P Chollet; M L Léophonte-Domairon; M Tiberge; J J Pessey; P Léophonte
Journal:  Rev Mal Respir       Date:  1997-06       Impact factor: 0.622

6.  Non-Invasive detection of respiratory effort-related arousals (REras) by a nasal cannula/pressure transducer system.

Authors:  I Ayappa; R G Norman; A C Krieger; A Rosen; R L O'malley; D M Rapoport
Journal:  Sleep       Date:  2000-09-15       Impact factor: 5.849

7.  Oxygen saturation regularity analysis in the diagnosis of obstructive sleep apnea.

Authors:  Félix del Campo; Roberto Hornero; Carlos Zamarrón; Daniel E Abasolo; Daniel Alvarez
Journal:  Artif Intell Med       Date:  2005-12-28       Impact factor: 5.326

8.  Pulseoximetry: sufficient to diagnose severe sleep apnea.

Authors:  Britt Oeverland; Olav Skatvedt; Kari J Kvaerner; Harriet Akre
Journal:  Sleep Med       Date:  2002-03       Impact factor: 3.492

9.  Model for investigating snorers with suspected sleep apnoea.

Authors:  H Rauscher; W Popp; H Zwick
Journal:  Thorax       Date:  1993-03       Impact factor: 9.139

10.  Prediction of the apnea-hypopnea index from overnight pulse oximetry.

Authors:  Ulysses J Magalang; Jacek Dmochowski; Sateesh Veeramachaneni; Azmi Draw; M Jeffery Mador; Ali El-Solh; Brydon J B Grant
Journal:  Chest       Date:  2003-11       Impact factor: 9.410

View more
  10 in total

1.  Accuracy of the clinical parameters and oximetry to initiate CPAP in patients with suspected obstructive sleep apnea.

Authors:  Carlos Alberto Nigro; Eduardo Dibur; Edgardo Rhodius
Journal:  Sleep Breath       Date:  2011-10-14       Impact factor: 2.816

2.  Effect of continuous positive airway pressure on regional cerebral blood flow during wakefulness in obstructive sleep apnea.

Authors:  Satomi Shiota; Yuichi Inoue; Hidenori Takekawa; Masaki Kotajima; Mami Nakajyo; Chie Usui; Yasuko Yoshioka; Tatsuo Koga; Kazuhisa Takahashi
Journal:  Sleep Breath       Date:  2013-09-13       Impact factor: 2.816

3.  Case-control studies of novel hemoglobin anomalies as differential diagnosis in sleep apnea syndrome.

Authors:  Berndt Zur; Michael Ludwig; Birgit Stoffel-Wagner
Journal:  Sleep Breath       Date:  2012-05-13       Impact factor: 2.816

4.  Comparison of two home sleep testing devices with different strategies for diagnosis of OSA.

Authors:  Tyler Gumb; Akosua Twumasi; Shahnaz Alimokhtari; Alan Perez; Kathleen Black; David M Rapoport; Jag Sunderram; Indu Ayappa
Journal:  Sleep Breath       Date:  2017-08-19       Impact factor: 2.816

5.  A novel, simple, and accurate pulse oximetry indicator for screening adult obstructive sleep apnea.

Authors:  Carlos Alberto Nigro; Gonzalo Castaño; Ignacio Bledel; Alfredo Colombi; María Cecilia Zicari
Journal:  Sleep Breath       Date:  2021-09-23       Impact factor: 2.655

6.  Difference between apnea-hypopnea index (AHI) and oxygen desaturation index (ODI): proportional increase associated with degree of obesity.

Authors:  G Ernst; M Bosio; A Salvado; E Dibur; C Nigro; E Borsini
Journal:  Sleep Breath       Date:  2016-03-30       Impact factor: 2.816

Review 7.  Computer-Assisted Diagnosis of the Sleep Apnea-Hypopnea Syndrome: A Review.

Authors:  Diego Alvarez-Estevez; Vicente Moret-Bonillo
Journal:  Sleep Disord       Date:  2015-07-21

8.  The effect of obstructive sleep apnea and treatment with continuous positive airway pressure on stroke rehabilitation: rationale, design and methods of the TOROS study.

Authors:  Justine A Aaronson; Coen A M van Bennekom; Winni F Hofman; Tijs van Bezeij; Joost G van den Aardweg; Erny Groet; Wytske A Kylstra; Ben A Schmand
Journal:  BMC Neurol       Date:  2014-02-25       Impact factor: 2.474

9.  Pulse oximetry for the detection of obstructive sleep apnea syndrome: can the memory capacity of oxygen saturation influence their diagnostic accuracy?

Authors:  Carlos A Nigro; Eduardo Dibur; Edgardo Rhodius
Journal:  Sleep Disord       Date:  2011-10-19

10.  Ambulatory pulse oximetry monitoring in Japanese COPD outpatients not receiving oxygen therapy.

Authors:  Seigo Minami; Suguru Yamamoto; Yoshitaka Ogata; Takeshi Nakatani; Yoshiko Takeuchi; Masanari Hamaguchi; Taro Koba; Kiyoshi Komuta
Journal:  Multidiscip Respir Med       Date:  2014-04-17
  10 in total