[Biometry of the anterior segment using optical coherence tomography : Evaluation of different devices and analysis programs].

A Viestenz1, S Vogt, A Langenbucher, S Walter, W Behrens-Baumann.   

Abstract

BACKGROUND: Digital visualisation of the anterior eye segment is becoming more and more important. Two commercially available optical coherence tomographs (OCTs) - Visante OCT (Carl-Zeiss Meditec Jena) and Slit Lamp-OCT (SL-OCT. Heidelberg Engineering) - were evaluated. Additionally, a manual and a semiautomatic analysis mode for the anterior segment biometry were compared using the SL-OCT. PATIENTS AND METHODS: Fifty eyes were examined by three ophthalmologists with complete eye status in a standardised fashion. Three anterior segment scans (horizontal and vertical) were performed using the Visante OCT and the SL-OCT. The manual integrated analysis mode of the Visante OCT was used. The best centred SL-OCT scan was analysed with a manual (EyelabGlobal/4Optics) and a semiautomatic procedure (HEYEX). Central corneal thickness (CCT), anterior chamber depth (ACD), pupillary diameter (PD) and chamber angle-angle distance (CAAD) were compared.
RESULTS: Comparison of Visante OCT and SL-OCT: The horizontal CAAD was 11.65+/-0.47 mm for the Visante-OCT and 12.0+/-0.57 mm for the SL-OCT (p=0,002), vertical scans were comparable in 10 eyes only. The CCT was 568+/-61 microm with the Visante-OCT and 581+/-48 microm with the SL-OCT (n=35, p=0.03) in horizontal scans and 565+/-62 microm with the Visante-OCT and 568+/-51 microm with the SL-OCT in vertical scans (n=27, not significant). Deviation of the two measurement methods varied between -54.7 and 80.8 microm in horizontal scans and between -84.9 and 91.1 microm in vertical scans from the mean difference of the two methods. Comparison of the analysis programs: Horizontal scans: Manual analysis correlated with semiautomatic analysis in CCT as 581+/-51 microm vs. 572+/-53 microm (r=0.903), ACD 2.89+/-0.74 mm vs. 2.08+/-0.72 mm (r=0.98), PD 5.22+/-2.12 mm vs. 5.14+/-1.91 mm (r=0.917) and CAAD 11.59+/-1.02 mm vs. 11.79+/-0.6 mm (r=0.47). The manual and a semiautomatic analysis modes for CCT and ACD differed significantly (p<0.026). Vertical scan: Manual versus semiautomatic analysis modes correlated in CCT as 578+/-65 microm vs. 573+/-63 microm (r=0.593), ACD with 3.04+/-0.83 mm vs. 3.03+/-0.75 mm (r=0.92), PD with 5.28+/-1.99 mm vs. 5.45+/-2.00 mm (r=0.899) and CAAD with 11.75+/-0.66 mm vs. 11.82+/-0.6 mm (r=0.537).
CONCLUSIONS: A near complete investigation of the cornea-anterior-chamber-iris segment is provided by the horizontal scan compared with the vertical scan. The noncontact OCT method is an easily handled tool for visualisation and biometry of the anterior eye segment. The investigated OCTs show a limited comparability. A complete analysis was possible in all eyes with the EyelabGlobal system in contrast to the HEYEX analysis software. Both analysis programs provided comparable measurements of the anterior eye segment. The semiautomatic biometrical mode may reduce the analysis time in qualitative excellent and well-centred scans to 10%.

Entities:  

Mesh:

Year:  2009        PMID: 18830603     DOI: 10.1007/s00347-008-1836-6

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  16 in total

1.  Corneal optical coherence tomography before and immediately after excimer laser photorefractive keratectomy.

Authors:  C Wirbelauer; C Scholz; H Hoerauf; R Engelhardt; R Birngruber; H Laqua
Journal:  Am J Ophthalmol       Date:  2000-12       Impact factor: 5.258

2.  [Examination of the cornea using optical coherence tomography].

Authors:  C Wirbelauer; C Scholz; H Hoerauf; G O Bastian; R Engelhardt; R Birngruber; H Laqua
Journal:  Ophthalmologe       Date:  2001-02       Impact factor: 1.059

3.  Noncontact corneal pachymetry with slit lamp-adapted optical coherence tomography.

Authors:  Christopher Wirbelauer; Christian Scholz; Hans Hoerauf; Duy Thoai Pham; Horst Laqua; Reginald Birngruber
Journal:  Am J Ophthalmol       Date:  2002-04       Impact factor: 5.258

4.  Anterior chamber width measurement by high-speed optical coherence tomography.

Authors:  Jason A Goldsmith; Yan Li; Maria Regina Chalita; Volker Westphal; Chetan A Patil; Andrew M Rollins; Joseph A Izatt; David Huang
Journal:  Ophthalmology       Date:  2005-02       Impact factor: 12.079

5.  Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles.

Authors:  Sunita Radhakrishnan; Jason Goldsmith; David Huang; Volker Westphal; David K Dueker; Andrew M Rollins; Joseph A Izatt; Scott D Smith
Journal:  Arch Ophthalmol       Date:  2005-08

6.  Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry.

Authors:  Hanna Y Kim; Donald L Budenz; Pak S Lee; William J Feuer; Keith Barton
Journal:  Am J Ophthalmol       Date:  2007-12-11       Impact factor: 5.258

7.  [Comparison of the corneal thickness as measured by Visante anterior segment OCT versus ultrasound technique].

Authors:  K Wildner; M Müller; J Dawczynski; J Strobel
Journal:  Klin Monbl Augenheilkd       Date:  2007-11       Impact factor: 0.700

8.  [Imaging of the anterior eye chamber with optical coherence tomography].

Authors:  C Wirbelauer; R Gochmann; D T Pham
Journal:  Klin Monbl Augenheilkd       Date:  2005-11       Impact factor: 0.700

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

10.  Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry.

Authors:  Paul S Zhao; Tien Yin Wong; Wan-Ling Wong; Seang-Mei Saw; Tin Aung
Journal:  Am J Ophthalmol       Date:  2007-06       Impact factor: 5.258

View more
  2 in total

1.  Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

Authors:  Jehn-Yu Huang; Melike Pekmezci; Stephanie Yaplee; Shan Lin
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-30       Impact factor: 3.117

2.  [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices].

Authors:  L Kiraly; G Duncker
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.