Literature DB >> 18765891

Improving the quantitative accuracy of optical-emission computed tomography by incorporating an attenuation correction: application to HIF1 imaging.

E Kim1, J Bowsher, A S Thomas, H Sakhalkar, M Dewhirst, M Oldham.   

Abstract

Optical computed tomography (optical-CT) and optical-emission computed tomography (optical-ECT) are new techniques for imaging the 3D structure and function (including gene expression) of whole unsectioned tissue samples. This work presents a method of improving the quantitative accuracy of optical-ECT by correcting for the 'self'-attenuation of photons emitted within the sample. The correction is analogous to a method commonly applied in single-photon-emission computed tomography reconstruction. The performance of the correction method was investigated by application to a transparent cylindrical gelatin phantom, containing a known distribution of attenuation (a central ink-doped gelatine core) and a known distribution of fluorescing fibres. Attenuation corrected and uncorrected optical-ECT images were reconstructed on the phantom to enable an evaluation of the effectiveness of the correction. Significant attenuation artefacts were observed in the uncorrected images where the central fibre appeared approximately 24% less intense due to greater attenuation from the surrounding ink-doped gelatin. This artefact was almost completely removed in the attenuation-corrected image, where the central fibre was within approximately 4% of the others. The successful phantom test enabled application of attenuation correction to optical-ECT images of an unsectioned human breast xenograft tumour grown subcutaneously on the hind leg of a nude mouse. This tumour cell line had been genetically labelled (pre-implantation) with fluorescent reporter genes such that all viable tumour cells expressed constitutive red fluorescent protein and hypoxia-inducible factor 1 transcription-produced green fluorescent protein. In addition to the fluorescent reporter labelling of gene expression, the tumour microvasculature was labelled by a light-absorbing vasculature contrast agent delivered in vivo by tail-vein injection. Optical-CT transmission images yielded high-resolution 3D images of the absorbing contrast agent, and revealed highly inhomogeneous vasculature perfusion within the tumour. Optical-ECT emission images yielded high-resolution 3D images of the fluorescent protein distribution in the tumour. Attenuation-uncorrected optical-ECT images showed clear loss of signal in regions of high attenuation, including regions of high perfusion, where attenuation is increased by increased vascular ink stain. Application of attenuation correction showed significant changes in an apparent expression of fluorescent proteins, confirming the importance of the attenuation correction. In conclusion, this work presents the first development and application of an attenuation correction for optical-ECT imaging. The results suggest that successful attenuation correction for optical-ECT is feasible and is essential for quantitatively accurate optical-ECT imaging.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18765891      PMCID: PMC3223488          DOI: 10.1088/0031-9155/53/19/007

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  24 in total

1.  A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results.

Authors:  S J Doran; K K Koerkamp; M A Bero; P Jenneson; E J Morton; W B Gilboy
Journal:  Phys Med Biol       Date:  2001-12       Impact factor: 3.609

2.  High resolution gel-dosimetry by optical-CT and MR scanning.

Authors:  M Oldham; J H Siewerdsen; A Shetty; D A Jaffray
Journal:  Med Phys       Date:  2001-07       Impact factor: 4.071

Review 3.  X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.

Authors:  Paul E Kinahan; Bruce H Hasegawa; Thomas Beyer
Journal:  Semin Nucl Med       Date:  2003-07       Impact factor: 4.446

4.  Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of an optical scanner.

Authors:  J C Gore; M Ranade; M J Maryañski; R J Schulz
Journal:  Phys Med Biol       Date:  1996-12       Impact factor: 3.609

5.  Correction of nonuniform attenuation in cardiac SPECT imaging.

Authors:  B M Tsui; G T Gullberg; E R Edgerton; J G Ballard; J R Perry; W H McCartney; J Berg
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

6.  Observation of incipient tumor angiogenesis that is independent of hypoxia and hypoxia inducible factor-1 activation.

Authors:  Yiting Cao; Chuan-Yuan Li; Benjamin J Moeller; Daohai Yu; Yulin Zhao; Matthew R Dreher; Siqing Shan; Mark W Dewhirst
Journal:  Cancer Res       Date:  2005-07-01       Impact factor: 12.701

7.  Ordered subsets algorithms for transmission tomography.

Authors:  H Erdogan; J A Fessler
Journal:  Phys Med Biol       Date:  1999-11       Impact factor: 3.609

8.  Functional imaging in bulk tissue specimens using optical emission tomography: fluorescence preservation during optical clearing.

Authors:  H S Sakhalkar; M Dewhirst; T Oliver; Y Cao; M Oldham
Journal:  Phys Med Biol       Date:  2007-03-20       Impact factor: 3.609

9.  Optical CT reconstruction of 3D dose distributions using the ferrous-benzoic-xylenol (FBX) gel dosimeter.

Authors:  R G Kelly; K J Jordan; J J Battista
Journal:  Med Phys       Date:  1998-09       Impact factor: 4.071

10.  Optical projection tomography as a tool for 3D microscopy and gene expression studies.

Authors:  James Sharpe; Ulf Ahlgren; Paul Perry; Bill Hill; Allyson Ross; Jacob Hecksher-Sørensen; Richard Baldock; Duncan Davidson
Journal:  Science       Date:  2002-04-19       Impact factor: 47.728

View more
  2 in total

1.  A comprehensive method for optical-emission computed tomography.

Authors:  Andrew Thomas; James Bowsher; Justin Roper; Tim Oliver; Mark Dewhirst; Mark Oldham
Journal:  Phys Med Biol       Date:  2010-06-24       Impact factor: 3.609

2.  Angular restriction fluorescence optical projection tomography to localize micrometastases in lymph nodes.

Authors:  Veronica C Torres; Chengyue Li; Yusheng He; Lagnojita Sinha; Georgia Papavasiliou; Husain A Sattar; Jovan G Brankov; Kenneth M Tichauer
Journal:  J Biomed Opt       Date:  2019-11       Impact factor: 3.170

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.