Literature DB >> 18757715

Image quality vs radiation dose of four cone beam computed tomography scanners.

M Loubele1, R Jacobs, F Maes, K Denis, S White, W Coudyzer, I Lambrichts, D van Steenberghe, P Suetens.   

Abstract

OBJECTIVES: To evaluate image quality by examining segmentation accuracy and assess radiation dose for cone beam CT (CBCT) scanners.
METHODS: A skull phantom, scanned by a laser scanner, and a contrast phantom were used to evaluate segmentation accuracy. The contrast phantom consisted of a polymethyl methacrylate (PMMA) cylinder with cylindrical inserts of air, bone and PMMA. The phantoms were scanned on the (1) Accuitomo 3D, (2) MercuRay, (3) NewTom 3G, (4) i-CAT and (5) Sensation 16. The structures were segmented with an optimal threshold. Thicknesses of the bone of the mandible and the diameter of the cylinders in the contrast phantom were measured across lines at corresponding places in the CT image vs a ground truth. The accuracy was in the 95th percentile of the difference between corresponding measurements. The correlation between accuracy in skull and contrast phantom was calculated. The radiation dose was assessed by DPI(100,c) (dose profile integral (100,c)) at the central hole of a CT dose index (CTDI) phantom.
RESULTS: The results for the DPI(100,c) were 107 mGy mm for (1), 1569 mGy mm for (2), 446 mGy mm for (3), 249 mGy mm for (4) and 1090 mGy mm for (5). The segmentations in the contrast phantom were submillimeter accurate in all scanners. The segmentation accuracy of the mandible was 2.9 mm for (1), 4.2 mm for (2), 3.4 mm for (3), 1.0 mm for (4) and 1.2 mm for (5). The correlation between measurements in the contrast and skull phantom was below 0.37 mm.
CONCLUSIONS: The best radiation dose vs image quality was found for the i-CAT.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18757715     DOI: 10.1259/dmfr/16770531

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  41 in total

1.  Correlation of panoramic radiography and cone beam CT findings in the assessment of the relationship between impacted mandibular third molars and the mandibular canal.

Authors:  F S Neves; T C Souza; S M Almeida; F Haiter-Neto; D Q Freitas; F N Bóscolo
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  The influence of exposure parameters on jawbone model accuracy using cone beam CT and multislice CT.

Authors:  B Vandenberghe; S Luchsinger; J Hostens; E Dhoore; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

3.  Effect of electric potential and current on mandibular linear measurements in cone beam CT.

Authors:  S Panmekiate; W Apinhasmit; A Petersson
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

Review 4.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

5.  Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children.

Authors:  Jose A Hidalgo Rivas; Keith Horner; Badri Thiruvenkatachari; Jonathan Davies; Chrysoula Theodorakou
Journal:  Br J Radiol       Date:  2015-08-17       Impact factor: 3.039

6.  Comparison of cone beam CT device and field of view for the detection of simulated periapical bone lesions.

Authors:  M Hedesiu; M Baciut; G Baciut; O Nackaerts; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2012-05-03       Impact factor: 2.419

7.  Comparison of observer reliability of three-dimensional cephalometric landmark identification on subject images from Galileos and i-CAT cone beam CT.

Authors:  R A Katkar; C Kummet; D Dawson; L Moreno Uribe; V Allareddy; M Finkelstein; A Ruprecht
Journal:  Dentomaxillofac Radiol       Date:  2013-07-05       Impact factor: 2.419

8.  Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom.

Authors:  J Bamba; K Araki; A Endo; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

9.  Influence of CBCT parameters on the output of an automatic edge-detection-based endodontic segmentation.

Authors:  J Michetti; M Georgelin-Gurgel; J-P Mallet; F Diemer; K Boulanouar
Journal:  Dentomaxillofac Radiol       Date:  2015-06-02       Impact factor: 2.419

Review 10.  Clinical guidelines for dental cone-beam computed tomography.

Authors:  Takafumi Hayashi; Yoshinori Arai; Toru Chikui; Sachiko Hayashi-Sakai; Kazuya Honda; Hiroko Indo; Taisuke Kawai; Kaoru Kobayashi; Shumei Murakami; Masako Nagasawa; Munetaka Naitoh; Eiji Nakayama; Yutaka Nikkuni; Hideyoshi Nishiyama; Noriaki Shoji; Shigeaki Suenaga; Ray Tanaka
Journal:  Oral Radiol       Date:  2018-01-11       Impact factor: 1.852

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.