Effect of object location on the density measurement and Hounsfield conversion in a NewTom 3G cone beam computed tomography unit.

M O Lagravère1, J Carey, M Ben-Zvi, G V Packota, P W Major.   

Abstract

OBJECTIVES: The purpose of this study was to determine the effect of an object's location in a cone beam CT imaging chamber (CBCT-NewTom 3G) on its apparent density and to develop a linear conversion coefficient for Hounsfield units (HU) to material density (g cm(-3)) for the NewTom 3G Scanner.
METHODS: Three cylindrical models of materials with different densities were constructed and scanned at five different locations in a NewTom 3G Volume Scanner. The average HU value for each model at each location was obtained using two different types of software. Next, five cylinders of different known densities were scanned at the exact centre of a NewTom 3G Scanner. The collected data were analysed using the same two types of software to determine a standard linear relationship between density and HU for each type of software.
RESULTS: There is no statistical significance of location of an object within the CBCT scanner on determination of its density. A linear relationship between the density of an object and the HU of a scan was rho = 0.001(HU)+1.19 with an R2 value of 0.893 (where density, rho, is measured in g cm(-3)). This equation is to be used on a range between 1.42 g cm(-3) and 0.4456 g cm(-3).
CONCLUSIONS: A linear relationship can be used to determine the density of materials (in the density range of bone) from the HU values of a CBCT scan. This relationship is not affected by the object's location within the scanner itself.

Mesh:

Year:  2008        PMID: 18757714     DOI: 10.1259/dmfr/65993482

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  21 in total

1.  Influence of graft quality and marginal bone loss on implants placed in maxillary grafted sinus: a finite element study.

Authors:  Samroeng Inglam; Siriwan Suebnukarn; Wichit Tharanon; Tratat Apatananon; Kriskrai Sitthiseripratip
Journal:  Med Biol Eng Comput       Date:  2010-04-29       Impact factor: 2.602

2.  Bone density changes around teeth during orthodontic treatment.

Authors:  Jui-Ting Hsu; Hsun-Wen Chang; Heng-Li Huang; Jian-Hong Yu; Yu-Fen Li; Ming-Gene Tu
Journal:  Clin Oral Investig       Date:  2011-08       Impact factor: 3.573

3.  Effects of orthodontic tooth movement on alveolar bone density.

Authors:  Hsing-Wen Chang; Heng-Li Huang; Jian-Hong Yu; Jui-Ting Hsu; Yu-Fen Li; Yi-Fan Wu
Journal:  Clin Oral Investig       Date:  2011-04-26       Impact factor: 3.573

Review 4.  CBCT-based bone quality assessment: are Hounsfield units applicable?

Authors:  R Pauwels; R Jacobs; S R Singer; M Mupparapu
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

5.  3D patient-specific model of the tibia from CT for orthopedic use.

Authors:  Raide A González-Carbonell; Armando Ortiz-Prado; Victor H Jacobo-Armendáriz; Yosbel A Cisneros-Hidalgo; Armando Alpízar-Aguirre
Journal:  J Orthop       Date:  2015-01-31

Review 6.  Bone mineral density in cone beam computed tomography: Only a few shades of gray.

Authors:  Marcio José da Silva Campos; Thainara Salgueiro de Souza; Sergio Luiz Mota Júnior; Marcelo Reis Fraga; Robert Willer Farinazzo Vitral
Journal:  World J Radiol       Date:  2014-08-28

7.  Soft tissue changes in the orofacial region after rapid maxillary expansion : A cone beam computed tomography study.

Authors:  Gülşilay Sayar Torun
Journal:  J Orofac Orthop       Date:  2017-05       Impact factor: 1.938

8.  Assessment of CT numbers in limited and medium field-of-view scans taken using Accuitomo 170 and Veraviewepocs 3De cone-beam computed tomography scanners.

Authors:  Matheus L Oliveira; Guilherme M Tosoni; David H Lindsey; Kristopher Mendoza; Sotirios Tetradis; Sanjay M Mallya
Journal:  Imaging Sci Dent       Date:  2014-11-25

9.  Evaluation of dehiscences using cone beam computed tomography.

Authors:  Nicholas Ising; Ki Beom Kim; Eustaquio Araujo; Peter Buschang
Journal:  Angle Orthod       Date:  2011-07-27       Impact factor: 2.079

10.  Variability of dental cone beam CT grey values for density estimations.

Authors:  R Pauwels; O Nackaerts; N Bellaiche; H Stamatakis; K Tsiklakis; A Walker; H Bosmans; R Bogaerts; R Jacobs; K Horner
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.