Safety and accuracy of bedside external ventricular drain placement.

Udaya K Kakarla1, Louis J Kim, Steven W Chang, Nicholas Theodore, Robert F Spetzler.   

Abstract

OBJECTIVE: To study the safety and accuracy of ventriculostomy by neurosurgical trainees.
METHODS: Initial computed tomographic studies of 346 consecutive patients who underwent bedside ventriculostomy were reviewed retrospectively. Diagnosis, catheter tip location, midline shift, and procedural complications were tabulated. To analyze catheter placement, we used a new grading system: Grade 1, optimal placement in the ipsilateral frontal horn or third ventricle; Grade 2, functional placement in the contralateral lateral ventricle or noneloquent cortex; and Grade 3, suboptimal placement in the eloquent cortex or nontarget cerebrospinal fluid space, with or without functional drainage. Statistical analysis was performed using Fisher's exact test and a weighted kappa coefficient.
RESULTS: Diagnoses included the following: subarachnoid hemorrhage, n = 153 (44%); trauma, n = 64 (18%); intracerebral hemorrhage/intraventricular hemorrhage, n = 63 (18%); and other, n = 66 (20%). There were 266 (77%) Grade 1, 34 (10%) Grade 2, and 46 (13%) Grade 3 catheter placements. Hemorrhagic complications occurred in 17 (5%). Four patients (1.2%) were symptomatic, with two (0.6%) requiring surgery. Inter- and intraobserver agreement was almost perfect (kappa = 0.846 and 0.922, respectively) as applied to our grading system. Rates of suboptimal placement were highest in patients with midline shift (P = 0.059) and trauma (P = 0.0001). Rates of optimal placement were highest in patients with subarachnoid hemorrhage (P = 0.003) and when the catheter was placed ipsilateral to the side of midline shift (P = 0.063). Neither the resident's training experience nor the side of placement seemed to affect accuracy.
CONCLUSION: Bedside ventriculostomy is a safe and accurate procedure for intracranial pressure monitoring and cerebrospinal fluid drainage.

Entities:  

Mesh:

Year:  2008        PMID: 18728595     DOI: 10.1227/01.neu.0000335031.23521.d0

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  32 in total

Review 1.  External ventricular drainage for intraventricular hemorrhage.

Authors:  Mahua Dey; Jennifer Jaffe; Agnieszka Stadnik; Issam A Awad
Journal:  Curr Neurol Neurosci Rep       Date:  2012-02       Impact factor: 5.081

2.  Smart stylet: the development and use of a bedside external ventricular drain image-guidance system.

Authors:  Vaibhav Patil; Rajiv Gupta; Raúl San José Estépar; Ronilda Lacson; Arnold Cheung; Judith M Wong; A John Popp; Alexandra Golby; Christopher Ogilvy; Kirby G Vosburgh
Journal:  Stereotact Funct Neurosurg       Date:  2015-01-31       Impact factor: 1.875

Review 3.  The ventricular system of the brain: a comprehensive review of its history, anatomy, histology, embryology, and surgical considerations.

Authors:  M M Mortazavi; N Adeeb; C J Griessenauer; H Sheikh; S Shahidi; R I Tubbs; R S Tubbs
Journal:  Childs Nerv Syst       Date:  2013-11-16       Impact factor: 1.475

4.  Cingulum injury by external ventricular drainage procedure: diffusion tensor tractography study.

Authors:  H G Kwon; S H Jang
Journal:  Clin Neuroradiol       Date:  2013-11-13       Impact factor: 3.649

5.  Transcranial access using fluoroscopic flat panel detector CT navigation.

Authors:  D L Cooke; M Levitt; L J Kim; D K Hallam; B Ghodke
Journal:  AJNR Am J Neuroradiol       Date:  2010-03-18       Impact factor: 3.825

6.  A new system for 3D ultrasound-guided placement of cerebral ventricle catheters.

Authors:  I Reinertsen; A S Jakola; P Friderichsen; F Lindseth; O Solheim; T Selbekk; G Unsgård
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-06-02       Impact factor: 2.924

7.  Determinants of accuracy of freehand external ventricular drain placement by neurosurgical trainees.

Authors:  Masoud Pishjoo; Kasra Khatibi; Hamid Etemadrezaie; Samira Zabihyan; Babak Ganjeifar; Mohammad Safdari; Humain Baharvahdat
Journal:  Acta Neurochir (Wien)       Date:  2021-01-06       Impact factor: 2.216

8.  Accuracy of bolt external ventricular drain insertion by neurosurgeons of different experience.

Authors:  Ivan Cabrilo; Claudia L Craven; Debayan Dasgupta; Ugan Reddy; Ahmed K Toma
Journal:  Acta Neurochir (Wien)       Date:  2021-01-21       Impact factor: 2.216

9.  Bleeding and infection with external ventricular drainage: a systematic review in comparison with adjudicated adverse events in the ongoing Clot Lysis Evaluating Accelerated Resolution of Intraventricular Hemorrhage Phase III (CLEAR-III IHV) trial.

Authors:  Mahua Dey; Agnieszka Stadnik; Fady Riad; Lingjiao Zhang; Nichol McBee; Carlos Kase; J Ricardo Carhuapoma; Malathi Ram; Karen Lane; Noeleen Ostapkovich; Francois Aldrich; Charlene Aldrich; Jack Jallo; Ken Butcher; Ryan Snider; Daniel Hanley; Wendy Ziai; Issam A Awad
Journal:  Neurosurgery       Date:  2015-03       Impact factor: 4.654

10.  Validation of model-guided placement of external ventricular drains.

Authors:  I Reinertsen; A S Jakola; T Selbekk; O Solheim
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-01-11       Impact factor: 2.924