Literature DB >> 25662506

Smart stylet: the development and use of a bedside external ventricular drain image-guidance system.

Vaibhav Patil1, Rajiv Gupta, Raúl San José Estépar, Ronilda Lacson, Arnold Cheung, Judith M Wong, A John Popp, Alexandra Golby, Christopher Ogilvy, Kirby G Vosburgh.   

Abstract

BACKGROUND: Placement accuracy of ventriculostomy catheters is reported in a wide and variable range. Development of an efficient image-guidance system may improve physician performance and patient safety.
OBJECTIVE: We evaluate the prototype of Smart Stylet, a new electromagnetic image-guidance system for use during bedside ventriculostomy.
METHODS: Accuracy of the Smart Stylet system was assessed. System operators were evaluated for their ability to successfully target the ipsilateral frontal horn in a phantom model.
RESULTS: Target registration error across 15 intracranial targets ranged from 1.3 to 4.6 mm (mean 3.1 mm). Using Smart Stylet guidance, a test operator successfully passed a ventriculostomy catheter to a shifted ipsilateral frontal horn 20/20 (100%) times from the frontal approach in a skull phantom. Without Smart Stylet guidance, the operator was successful 4/10 (40%) times from the right frontal approach and 6/10 (60%) times from the left frontal approach. In a separate experiment, resident operators were successful 2/4 (50%) times when targeting the shifted ipsilateral frontal horn with Smart Stylet guidance and 0/4 (0%) times without image guidance using a skull phantom.
CONCLUSIONS: Smart Stylet may improve the ability to successfully target the ventricles during frontal ventriculostomy.
© 2015 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2015        PMID: 25662506      PMCID: PMC4423620          DOI: 10.1159/000368906

Source DB:  PubMed          Journal:  Stereotact Funct Neurosurg        ISSN: 1011-6125            Impact factor:   1.875


  24 in total

1.  Change in ventricular size and effect of ventricular catheter placement in pediatric patients with shunted hydrocephalus.

Authors:  S Tuli; B O'Hayon; J Drake; M Clarke; J Kestle
Journal:  Neurosurgery       Date:  1999-12       Impact factor: 4.654

2.  Application accuracy of an electromagnetic field-based image-guided navigation system.

Authors:  Joshua M Rosenow; W Keith Sootsman
Journal:  Stereotact Funct Neurosurg       Date:  2006-12-12       Impact factor: 1.875

3.  Comparison of magnetic tracking and optical tracking by simultaneous use of two independent frameless stereotactic systems.

Authors:  Christopher R Mascott
Journal:  Neurosurgery       Date:  2005-10       Impact factor: 4.654

4.  Electromagnetic tracking for image-guided abdominal procedures: overall system and technical issues.

Authors:  Kevin Cleary; Hui Zhang; Neil Glossop; Elliot Levy; Bradford Wood; Filip Banovac
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2005

5.  Is good good enough?

Authors:  David W Roberts
Journal:  Neurocrit Care       Date:  2009-02-12       Impact factor: 3.210

6.  Application of electromagnetic technology to neuronavigation: a revolution in image-guided neurosurgery.

Authors:  Caroline Hayhurst; Patricia Byrne; Paul R Eldridge; Conor L Mallucci
Journal:  J Neurosurg       Date:  2009-12       Impact factor: 5.115

7.  Least-squares fitting of two 3-d point sets.

Authors:  K S Arun; T S Huang; S D Blostein
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1987-05       Impact factor: 6.226

8.  Ventriculostomy-associated infections: incidence and risk factors.

Authors:  Yaseen Arabi; Ziad A Memish; Hanan H Balkhy; Christine Francis; Ahmad Ferayan; Abdullah Al Shimemeri; Maha A Almuneef
Journal:  Am J Infect Control       Date:  2005-04       Impact factor: 2.918

9.  Computer-assisted image-guided surgery using the Regulus Navigator.

Authors:  J S Rousu; P E Kohls; B Kall; P J Kelly
Journal:  Stud Health Technol Inform       Date:  1998

10.  Malplacement of ventricular catheters by neurosurgeons: a single institution experience.

Authors:  Andrea Saladino; J Bradley White; Eelco F M Wijdicks; Giuseppe Lanzino
Journal:  Neurocrit Care       Date:  2008-10-16       Impact factor: 3.210

View more
  3 in total

1.  Recent advances in surgical planning & navigation for tumor biopsy and resection.

Authors:  Defeng Wang; Diya Ma; Matthew Lun Wong; Yì Xiáng J Wáng
Journal:  Quant Imaging Med Surg       Date:  2015-10

2.  Optimizing accuracy of freehand cannulation of the ipsilateral ventricle for intracranial pressure monitoring in patients with brain trauma.

Authors:  Zhongyi Sun; Lin Wu; Zhixiong Liu; Weiming Zhong; Zhifeng Kou; Jinfang Liu
Journal:  Quant Imaging Med Surg       Date:  2020-11

3.  Bullseye EVD: preclinical evaluation of an intra-procedural system to confirm external ventricular drainage catheter positioning.

Authors:  Adam Hopfgartner; David Burns; Suganth Suppiah; Allan R Martin; Michael Hardisty; Cari M Whyne
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-05-28       Impact factor: 2.924

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.