Literature DB >> 21633798

A new system for 3D ultrasound-guided placement of cerebral ventricle catheters.

I Reinertsen1, A S Jakola, P Friderichsen, F Lindseth, O Solheim, T Selbekk, G Unsgård.   

Abstract

PURPOSE: We present a new system for 3D ultrasound-guided placement of cerebral ventricle catheters. The system has been developed with the aim to provide accurate ultrasound-based guidance with only minimal changes to the current surgical technique and workflow.
METHODS: The system consists of a pre-calibrated navigation adapter for the catheter and a reference frame attached to a standard surgical retractor in addition to an ultrasound-based navigation system with a probe that fits on top of a standard burr hole.
RESULTS: The accuracy of the pre-calibrated system has been evaluated, and our measurements indicate that the accuracy of the pre-calibrated system is better than 3 mm. We also present a clinical case.
CONCLUSIONS: The navigation accuracy is considered sufficient for clinical use, and initial clinical tests are promising. Further testing will be necessary to fully evaluate the performance of the system in a clinical setting.

Entities:  

Mesh:

Year:  2011        PMID: 21633798     DOI: 10.1007/s11548-011-0622-0

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  9 in total

1.  Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters.

Authors:  Alan T Villavicencio; Jean-Christophe Leveque; Matthew J McGirt; John S Hopkins; Herbert E Fuchs; Timothy M George
Journal:  Surg Neurol       Date:  2003-05

2.  A new tool for frameless stereotactic placement of ventricular catheters.

Authors:  Lennart H Stieglitz; Mario Giordano; Madjid Samii; Wolf O Luedemann
Journal:  Neurosurgery       Date:  2010-09       Impact factor: 4.654

3.  Shunt malfunction.

Authors:  John R W Kestle
Journal:  J Neurosurg       Date:  2010-04-16       Impact factor: 5.115

4.  Accuracy of ultrasound-guided puncture of the ventricular system.

Authors:  M Strowitzki; Y Komenda; R Eymann; W I Steudel
Journal:  Childs Nerv Syst       Date:  2007-07-03       Impact factor: 1.475

5.  Accuracy of the freehand pass technique for ventriculostomy catheter placement: retrospective assessment using computed tomography scans.

Authors:  David R Huyette; Benjamin J Turnbow; Christian Kaufman; Dale F Vaslow; Benjamin B Whiting; Michael Y Oh
Journal:  J Neurosurg       Date:  2008-01       Impact factor: 5.115

6.  Safety and accuracy of bedside external ventricular drain placement.

Authors:  Udaya K Kakarla; Louis J Kim; Steven W Chang; Nicholas Theodore; Robert F Spetzler
Journal:  Neurosurgery       Date:  2008-07       Impact factor: 4.654

7.  Application of electromagnetic technology to neuronavigation: a revolution in image-guided neurosurgery.

Authors:  Caroline Hayhurst; Patricia Byrne; Paul R Eldridge; Conor L Mallucci
Journal:  J Neurosurg       Date:  2009-12       Impact factor: 5.115

8.  Accurate placement of cerebrospinal fluid shunt ventricular catheters with real-time ultrasound guidance in older children without patent fontanelles.

Authors:  William E Whitehead; Andrew Jea; Shobhan Vachhrajani; Abhaya V Kulkarni; James M Drake
Journal:  J Neurosurg       Date:  2007-11       Impact factor: 5.115

9.  External ventricular drain insertion accuracy: is there a need for change in practice?

Authors:  Ahmed K Toma; Sophie Camp; Laurence D Watkins; Joan Grieve; Neil D Kitchen
Journal:  Neurosurgery       Date:  2009-12       Impact factor: 4.654

  9 in total
  2 in total

1.  Deformable registration of preoperative MR, pre-resection ultrasound, and post-resection ultrasound images of neurosurgery.

Authors:  Hassan Rivaz; D Louis Collins
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-11-06       Impact factor: 2.924

2.  Validation of model-guided placement of external ventricular drains.

Authors:  I Reinertsen; A S Jakola; T Selbekk; O Solheim
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-01-11       Impact factor: 2.924

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.