Carotid stenosis assessed with a 4-dimensional semiautomated Doppler system.

Flemming Forsberg1, Alan D Stein, Daniel A Merton, Kathryn J Lipcan, Donald Herzog, Laurence Parker, Laurence Needleman.   

Abstract

OBJECTIVE: The purpose of this study was to compare peak systolic velocities (PSVs) and the degree of stenosis obtained with a real-time 3-dimensional (ie, 4-dimensional) Doppler ultrasound scanner (Encore PV; VueSonix Sensors Inc, Wayne, PA) to conventional Doppler ultrasound imaging of the carotid arteries (common [CCA], internal [ICA], and external [ECA]). A secondary goal was to assess Encore volume flow measurements.
METHODS: Seventy patients referred for clinical carotid ultrasound participated in this pilot study. Peak systolic velocities of the CCA, ECA, and ICA were obtained bilaterally. The degree of stenosis in the ICA was calculated based on the ICA PSV and ICA/CCA PSV ratio. The Encore detects all 3-dimensional blood flow velocity vectors within 10-s longitudinal volumes of the ICA, ECA, and CCA. On the Encore, a reader determined the centerline of the vessels. The PSV and volume flow were then automatically calculated. The flow measurement error was obtained by comparing the CCA flow to the ICA and ECA flow. Data were compared using linear regression, intraclass correlation coefficients (ICCs), and Bland-Altman analysis.
RESULTS: Due to technical difficulties, only 59 patients (323 vessel segments) were available for analysis. There was good agreement between methods for assessing the degree of stenosis based on the ICA PSV (ICC = 0.83; P < .0001) and, to a lesser degree, on the ICA/CCA PSV ratio (ICC = 0.65; P < .0001). Peak systolic velocity measurements obtained with conventional ultrasound and the Encore correlated in all vessels (r >or= 0.32; P < .002), and Bland-Altman analysis showed reasonable variations. The Encore mean volume flow error +/- SD was -4.1% +/- 66.4% and was not biased (P = .57).
CONCLUSIONS: A new semiautomated 4-dimensional Doppler device is comparable to conventional Doppler ultrasound for assessment of carotid stenosis.

Entities:  

Mesh:

Year:  2008        PMID: 18716143      PMCID: PMC2679685          DOI: 10.7863/jum.2008.27.9.1337

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  21 in total

Review 1.  Diagnosing carotid stenosis by Doppler sonography: state of the art.

Authors:  Diana Gaitini; Michalle Soudack
Journal:  J Ultrasound Med       Date:  2005-08       Impact factor: 2.153

2.  Validating volume flow measurements from a novel semiautomated four-dimensional Doppler ultrasound scanner.

Authors:  Flemming Forsberg; Alan D Stein; Ji-Bin Liu; Xuedong Deng; William Ackerman; Donald Herzog; Kenneth Abend; Laurence Needleman
Journal:  Acad Radiol       Date:  2006-10       Impact factor: 3.173

3.  Measurement of volumetric flow.

Authors:  Oliver D Kripfgans; Jonathan M Rubin; Anne L Hall; Michael B Gordon; J Brian Fowlkes
Journal:  J Ultrasound Med       Date:  2006-10       Impact factor: 2.153

Review 4.  Intraclass correlations: uses in assessing rater reliability.

Authors:  P E Shrout; J L Fleiss
Journal:  Psychol Bull       Date:  1979-03       Impact factor: 17.737

5.  Peak velocity overestimation and linear-array spectral Doppler.

Authors:  B M Eicke; F W Kremkau; H Hinson; C H Tegeler
Journal:  J Neuroimaging       Date:  1995-04       Impact factor: 2.486

6.  Volume flow estimation using time domain correlation and ultrasonic flowmetry.

Authors:  F Forsberg; J B Liu; K M Russell; S L Guthrie; B B Goldberg
Journal:  Ultrasound Med Biol       Date:  1995       Impact factor: 2.998

7.  Doppler ultrasound measurement of blood flow volume in the extracranial internal carotid artery for evaluation of brain perfusion after aneurysmal subarachnoid hemorrhage.

Authors:  Karl-Michael Schebesch; Chris Woertgen; Jürgen Schlaier; Alexander Brawanski; Ralf D Rothoerl
Journal:  Neurol Res       Date:  2007-03       Impact factor: 2.448

8.  A real-time 3-dimensional digital Doppler method for measurement of flow rate and volume through mitral valve in children: a validation study compared with magnetic resonance imaging.

Authors:  Shuping Ge; Liping Bu; Honghai Zhang; Erik Schelbert; Michelle Disterhoft; Xiangning Li; Xiaokui Li; David Sahn; Alan Stolpen; Milan Sonka
Journal:  J Am Soc Echocardiogr       Date:  2005-01       Impact factor: 5.251

Review 9.  Atherosclerotic carotid stenosis and occlusion.

Authors:  Matthias Sitzer
Journal:  Front Neurol Neurosci       Date:  2006

10.  Volume flow rate evaluation in patients with obstructive arteriosclerotic disease.

Authors:  Claudia Doberenz; Walter Paulus; Carl D Reimers; B Martin Eicke
Journal:  Cerebrovasc Dis       Date:  2004-09-03       Impact factor: 2.762

View more
  3 in total

Review 1.  Recent technological advancements in cardiac ultrasound imaging.

Authors:  Jaydev K Dave; Maureen E Mc Donald; Praveen Mehrotra; Andrew R Kohut; John R Eisenbrey; Flemming Forsberg
Journal:  Ultrasonics       Date:  2017-11-23       Impact factor: 2.890

2.  Accuracy of volumetric flow rate measurements: an in vitro study using modern ultrasound scanners.

Authors:  Kenneth Hoyt; Felix A Hester; Randall L Bell; Mark E Lockhart; Michelle L Robbin
Journal:  J Ultrasound Med       Date:  2009-11       Impact factor: 2.153

3.  Mean volume flow estimation in pulsatile flow conditions.

Authors:  Michael S Richards; Oliver D Kripfgans; Jonathan M Rubin; Anne L Hall; J Brian Fowlkes
Journal:  Ultrasound Med Biol       Date:  2009-10-12       Impact factor: 2.998

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.