Volume flow estimation using time domain correlation and ultrasonic flowmetry.

F Forsberg1, J B Liu, K M Russell, S L Guthrie, B B Goldberg.   

Abstract

A comparison of two volumetric blood flow measurement techniques, CVI-Q (based on time domain correlation) and ultrasonic flowmetry, has been performed in vitro and in vivo. A pulsatile flowpump was used to simulate carotid and femoral type waveforms which were measured simultaneously using the two methods. Five dogs had their common carotid and femoral arteries exposed, and the instantaneous maximum volume flow and the mean flow were measured. Each vessel was partially occluded halfway through the experiment, simulating a 90% stenosis. In vitro, both techniques achieved absolute errors below +/- 5% for flow rates over 100 mL/min, but ultrasonic flowmetry had statistically significant larger errors for slower flow rates. In vivo correlation coefficients ranging from 0.73 to 0.95 were obtained with regression line slopes close to unity. The two techniques were in reasonable agreement, but with standard deviations of 20% to 28%. These studies indicate that noninvasive CVI-Q measurements of blood flow in the carotid and femoral arteries are linear and accurate compared to invasive ultrasonic flowmetry.

Entities:  

Mesh:

Year:  1995        PMID: 8553498     DOI: 10.1016/0301-5629(95)00024-l

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  3 in total

1.  Intraobserver reproducibility and interobserver agreement of ultrasonic flowmetry using the velocity profile color doppler method in the common carotid artery.

Authors:  Nobuyuki Taniguchi; Ryuuichi Takano; Kouichi Itoh; Syukuko Kaneko; Akira Onoguchi; Yi Wang; Tomoyuki Kuwata; Kouichiro Shigeta; Tomoko Ono
Journal:  J Med Ultrason (2001)       Date:  2003-03       Impact factor: 1.314

2.  Carotid stenosis assessed with a 4-dimensional semiautomated Doppler system.

Authors:  Flemming Forsberg; Alan D Stein; Daniel A Merton; Kathryn J Lipcan; Donald Herzog; Laurence Parker; Laurence Needleman
Journal:  J Ultrasound Med       Date:  2008-09       Impact factor: 2.153

3.  Epidural Oscillating Cardiac-Gated Intracranial Implant Modulates Cerebral Blood Flow.

Authors:  Mark G Luciano; Stephen M Dombrowski; Serge El-Khoury; Jun Yang; Suraj Thyagaraj; Sara Qvarlander; Syed Khalid; Ian Suk; Amir Manbachi; Francis Loth
Journal:  Neurosurgery       Date:  2020-11-16       Impact factor: 4.654

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.