Trapping ring electrode cell: a FTICR mass spectrometer cell for improved signal-to-noise and resolving power.

Chad R Weisbrod1, Nathan K Kaiser, Gunnar E Skulason, James E Bruce.   

Abstract

A novel FTICR cell called the trapping ring electrode cell (TREC) has been conceived, simulated, developed, and tested. The performance of the TREC is compared to a closed cylindrical cell at different excited cyclotron radii. The TREC permits the ability to maintain coherent ion motion at larger initial excited cyclotron radii by decreasing the change in radial electric field with respect to z-axis position in the cell. This is accomplished through postexcitation modulation of the trapping potentials applied to segmented trap plates. Resolving power approaching the theoretical limit was achieved using the novel TREC technology; over 420,000 resolving power was observed on melittin [M + 4H] (4+) species when employed under modest magnetic field strength (3T) and a data acquisition duration of 13 s. A 10-fold gain in signal-to-noise ratio is demonstrated over the closed cylindrical cell optimized with common potentials on all ring electrodes. The observed frequency drift during signal acquisition over long time periods was also significantly reduced, resulting in improved resolving power.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18681460      PMCID: PMC2915943          DOI: 10.1021/ac800535e

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  27 in total

1.  Petroleomics: the next grand challenge for chemical analysis.

Authors:  Alan G Marshall; Ryan P Rodgers
Journal:  Acc Chem Res       Date:  2004-01       Impact factor: 22.384

2.  Global analysis of the Deinococcus radiodurans proteome by using accurate mass tags.

Authors:  Mary S Lipton; Ljiljana Pasa-Tolic'; Gordon A Anderson; David J Anderson; Deanna L Auberry; John R Battista; Michael J Daly; Jim Fredrickson; Kim K Hixson; Heather Kostandarithes; Christophe Masselon; Lye Meng Markillie; Ronald J Moore; Margaret F Romine; Yufeng Shen; Eric Stritmatter; Nikola Tolic'; Harold R Udseth; Amudhan Venkateswaran; Kwong-Kwok Wong; Rui Zhao; Richard D Smith
Journal:  Proc Natl Acad Sci U S A       Date:  2002-08-12       Impact factor: 11.205

3.  Theoretical and experimental prospects for protein identification based solely on accurate mass measurement.

Authors:  Fei He; Mark R Emmett; Kristina Håkansson; Christopher L Hendrickson; Alan G Marshall
Journal:  J Proteome Res       Date:  2004 Jan-Feb       Impact factor: 4.466

4.  Characterization of a new open cylindrical ion cyclotron resonance cell with unusual geometry.

Authors:  B Kanawati; K P Wanczek
Journal:  Rev Sci Instrum       Date:  2007-07       Impact factor: 1.523

5.  Reduction of axial kinetic energy induced perturbations on observed cyclotron frequency.

Authors:  Nathan K Kaiser; Chad R Weisbrod; Brian N Webb; James E Bruce
Journal:  J Am Soc Mass Spectrom       Date:  2008-01-06       Impact factor: 3.109

6.  Experimental Evidence for Space-Charge Effects between Ions of the Same Mass-to-Charge in Fourier-Transform Ion Cyclotron Resonance Mass Spectrometry.

Authors:  Richard L Wong; I Jonathan Amster
Journal:  Int J Mass Spectrom       Date:  2007-09-01       Impact factor: 1.986

Review 7.  Fourier transform ion cyclotron resonance mass spectrometry: a primer.

Authors:  A G Marshall; C L Hendrickson; G S Jackson
Journal:  Mass Spectrom Rev       Date:  1998 Jan-Feb       Impact factor: 10.946

8.  Fourier-transform mass spectrometry of large molecules by electrospray ionization.

Authors:  K D Henry; E R Williams; B H Wang; F W McLafferty; J Shabanowitz; D F Hunt
Journal:  Proc Natl Acad Sci U S A       Date:  1989-12       Impact factor: 11.205

9.  A "screened" electrostatic ion trap for enhanced mass resolution, mass accuracy, reproducibility, and upper mass limit in Fourier transform ion cyclotron resonance mass spectrometry.

Authors:  M Wang; A G Marshall
Journal:  Anal Chem       Date:  1989-06-01       Impact factor: 6.986

10.  Identification of modification sites in large biomolecules by stable isotope labeling and tandem high resolution mass spectrometry. The active site nucleophile of thiaminase I.

Authors:  N L Kelleher; R B Nicewonger; T P Begley; F W McLafferty
Journal:  J Biol Chem       Date:  1997-12-19       Impact factor: 5.157

View more
  11 in total

1.  Initial experimental characterization of a new ultra-high resolution FTICR cell with dynamic harmonization.

Authors:  Eugene N Nikolaev; Ivan A Boldin; Roland Jertz; Gökhan Baykut
Journal:  J Am Soc Mass Spectrom       Date:  2011-04-19       Impact factor: 3.109

2.  A novel 9.4 tesla FTICR mass spectrometer with improved sensitivity, mass resolution, and mass range.

Authors:  Nathan K Kaiser; John P Quinn; Gregory T Blakney; Christopher L Hendrickson; Alan G Marshall
Journal:  J Am Soc Mass Spectrom       Date:  2011-05-05       Impact factor: 3.109

3.  Peak coalescence, spontaneous loss of coherence, and quantification of the relative abundances of two species in the plasma regime: particle-in-cell modeling of Fourier transform ion cyclotron resonance mass spectrometry.

Authors:  M Takeshi Nakata; Grant W Hart; Bryan G Peterson
Journal:  J Am Soc Mass Spectrom       Date:  2010-06-11       Impact factor: 3.109

4.  Ion trap with narrow aperture detection electrodes for Fourier transform ion cyclotron resonance mass spectrometry.

Authors:  Konstantin O Nagornov; Anton N Kozhinov; Oleg Y Tsybin; Yury O Tsybin
Journal:  J Am Soc Mass Spectrom       Date:  2015-03-14       Impact factor: 3.109

5.  Twelve million resolving power on 4.7 T Fourier transform ion cyclotron resonance instrument with dynamically harmonized cell--observation of fine structure in peptide mass spectra.

Authors:  Igor A Popov; Konstantin Nagornov; Gleb N Vladimirov; Yury I Kostyukevich; Eugene N Nikolaev
Journal:  J Am Soc Mass Spectrom       Date:  2014-03-07       Impact factor: 3.109

6.  Multiparticle Simulations of Quadrupolar Ion Detection in an Ion Cyclotron Resonance Cell with Four Narrow Aperture Detection Electrodes.

Authors:  Joshua A Driver; Konstantin O Nagornov; Anton N Kozhinov; Yury O Tsybin; Andriy Kharchenko; I Jonathan Amster
Journal:  J Am Soc Mass Spectrom       Date:  2017-10-16       Impact factor: 3.109

7.  Excite-coupled trapping ring electrode cell (eTREC): radial trapping field control, linearized excitation, and improved detection.

Authors:  Chad R Weisbrod; Nathan K Kaiser; Gunnar E Skulason; James E Bruce
Journal:  Anal Chem       Date:  2010-07-15       Impact factor: 6.986

8.  A novel Fourier transform ion cyclotron resonance mass spectrometer with improved ion trapping and detection capabilities.

Authors:  Nathan K Kaiser; Gunnar E Skulason; Chad R Weisbrod; James E Bruce
Journal:  J Am Soc Mass Spectrom       Date:  2009-01-07       Impact factor: 3.109

Review 9.  Evaluation of major historical ICR cell designs using electric field simulations.

Authors:  Evgeny Nikolaev; Anton Lioznov
Journal:  Mass Spectrom Rev       Date:  2020-11-25       Impact factor: 9.011

Review 10.  Developments in FTICR-MS and Its Potential for Body Fluid Signatures.

Authors:  Simone Nicolardi; Bogdan Bogdanov; André M Deelder; Magnus Palmblad; Yuri E M van der Burgt
Journal:  Int J Mol Sci       Date:  2015-11-13       Impact factor: 5.923

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.