Abstract
This study has investigated the relationship between linear attenuation coefficients (mu) and Hounsfield units (HUs) for six materials covering the range of values found clinically. Narrow-beam mu values were measured by performing radionuclide transmission scans using (99m)Tc, (123)I, (131)I, (201)Tl and (111)In. The mu values were compared to published data. The relationships between mu and HU were determined. These relationships can be used to convert computed tomography (CT) images to mu-maps for single photon emission computed tomography (SPECT) attenuation correction.Mesh:
Substances:
Year: 2008 PMID: 18662614 DOI: 10.1016/j.apradiso.2008.01.002