Accuracy and precision of a method to study kinematics of the temporomandibular joint: combination of motion data and CT imaging.

Evre Baltali1, Kristin D Zhao, Matthew F Koff, Eugene E Keller, Kai-Nan An.   

Abstract

The purpose of the study was to test the precision and accuracy of a method used to track selected landmarks during motion of the temporomandibular joint (TMJ). A precision phantom device was constructed and relative motions between two rigid bodies on the phantom device were measured using optoelectronic (OE) and electromagnetic (EM) motion tracking devices. The motion recordings were also combined with a 3D CT image for each type of motion tracking system (EM+CT and OE+CT) to mimic methods used in previous studies. In the OE and EM data collections, specific landmarks on the rigid bodies were determined using digitization. In the EM+CT and OE+CT data sets, the landmark locations were obtained from the CT images. 3D linear distances and 3D curvilinear path distances were calculated for the points. The accuracy and precision for all 4 methods were evaluated (EM, OE, EM+CT and OE+CT). In addition, results were compared with and without the CT imaging (EM vs. EM+CT, OE vs. OE+CT). All systems overestimated the actual 3D curvilinear path lengths. All systems also underestimated the actual rotation values. The accuracy of all methods was within 0.5mm for 3D curvilinear path calculations, 0.05mm for 3D linear distance calculations and 0.2 degrees for rotation calculations. In addition, Bland-Altman plots for each configuration of the systems suggest that measurements obtained from either system are repeatable and comparable.

Entities:  

Mesh:

Year:  2008        PMID: 18617178      PMCID: PMC2556512          DOI: 10.1016/j.jbiomech.2008.05.023

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  6 in total

1.  Quantitative analysis of factors affecting intraoperative precision and stability of optoelectronic and electromagnetic tracking systems.

Authors:  A Wagner; K Schicho; W Birkfellner; M Figl; R Seemann; F König; Franz Kainberger; R Ewers
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

Review 2.  Modeling of temporomandibular joint function using MRI and jaw-tracking technologies--mechanics.

Authors:  Luigi M Gallo
Journal:  Cells Tissues Organs       Date:  2005       Impact factor: 2.481

3.  A least-squares algorithm for the equiform transformation from spatial marker co-ordinates.

Authors:  F E Veldpaus; H J Woltring; L J Dortmans
Journal:  J Biomech       Date:  1988       Impact factor: 2.712

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  A procedure for determining rigid body transformation parameters.

Authors:  J H Challis
Journal:  J Biomech       Date:  1995-06       Impact factor: 2.712

6.  A method for quantifying condylar motion in patients with osteoarthritis using an electromagnetic tracking device and computed tomography imaging.

Authors:  Evre Baltali; Kristin D Zhao; Matthew F Koff; Ercan Durmuş; Kai-Nan An; Eugene E Keller
Journal:  J Oral Maxillofac Surg       Date:  2008-05       Impact factor: 1.895

  6 in total
  5 in total

1.  Three-dimensional observation of the mouse embryo by micro-computed tomography: composition of the trigeminal ganglion.

Authors:  Hidekazu Aoyagi; Kohzo Tsuchikawa; Shin-ichi Iwasaki
Journal:  Odontology       Date:  2010-02-16       Impact factor: 2.634

2.  In vivo measurement of the 3D kinematics of the temporomandibular joint using miniaturized electromagnetic trackers: technical report.

Authors:  J-P Baeyens; H Gilomen; B Erdmann; R Clijsen; J Cabri; D Vissers
Journal:  Med Biol Eng Comput       Date:  2012-12-14       Impact factor: 2.602

3.  Effect of Off-Axis Fluoroscopy Imaging on Two-Dimensional Kinematics in the Lumbar Spine: A Dynamic In Vitro Validation Study.

Authors:  Kristin D Zhao; Ephraim I Ben-Abraham; Dixon J Magnuson; Jon J Camp; Lawrence J Berglund; Kai-Nan An; Gert Bronfort; Ralph E Gay
Journal:  J Biomech Eng       Date:  2016-05       Impact factor: 2.097

4.  Characterizing fluoroscopy based kinematic accuracy as a function of pulse width and velocity.

Authors:  Arin M Ellingson; Joseph D Mozingo; Dixon J Magnuson; Mark W Pagnano; Kristin D Zhao
Journal:  J Biomech       Date:  2016-10-11       Impact factor: 2.712

5.  An Evaluation of Jaw Tracking Movements in Patients with Total Joint Replacements versus a Control Group.

Authors:  Farhana Rahman; Felice Femiano; Patrick J Louis; Chung How Kau
Journal:  Medicina (Kaunas)       Date:  2022-05-30       Impact factor: 2.948

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.