| Literature DB >> 20491996 |
B D Hedley1, M Keeney, I Chin-Yee, W Brown.
Abstract
The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was performed in our 800-bed tertiary care hospital and specifically centered upon the correlation of WBC, NRBC, and platelet (PLT) enumeration when compared to a predicate analyzer, the Coulter® LH 780, and flow cytometry (FCM) reference methods. Of particular interest were those samples with morphologically confirmed interference and extreme leukocytosis (evaluated with respect to red blood cell parameter correction). The sample set (n=272) consisted of morphologically normal and hematologically abnormal patients. Correlation of the WBC, PLT, and NRBC showed r(2) values of 0.994, 0.985, and 0.910 for the DxH 800 vs. FCM, respectively. The presence of interfering particles did not affect the accuracy of the DxH 800 with respect to WBC counts. The DxH 800 showed accurate PLT and NRBC counts in the clinically significant low range when compared to FCM. Compared to the LH 780, flagging rates were significantly reduced (NRBC flag), or equivalent (WBC, PLT flag) on the DxH 800. The DxH 800 demonstrated higher sensitivity and specificity for PLTs and NRBCs and achieved a lower NRBC false negative rate compared to the LH 780. The UniCel® DxH 800 represents a significant improvement to previous impedance analyzers in accurately detecting the presence of NRBCs at counts >1/100 WBC. Furthermore, it provides accurate PLT and WBC counts in the presence of interference and improved NRBC flagging efficiency when compared to the LH 780. Correction of red blood cell parameters is appropriate and accurate in cases of extreme leukocytosis.Entities:
Mesh:
Year: 2011 PMID: 20491996 PMCID: PMC3044820 DOI: 10.1111/j.1751-553X.2010.01239.x
Source DB: PubMed Journal: Int J Lab Hematol ISSN: 1751-5521 Impact factor: 2.877
The range of linearity of the new DxH 800 cellular analysis system and the LH 780
| Low limit | Upper limit | |||
|---|---|---|---|---|
| Parameter | DxH 800 | LH 780 | DxH 800 | LH 780 |
| WBC (×109/l) | 0.000 | 0.000 | 400.000 | 400.00 |
| RBC (×1012/l) | 0.000 | 0.000 | 8.500 | 8.000 |
| HGB (g/l) | 0.00 | 0.00 | 255.0 | 250.0 |
| MCV (fl) | 50.00 | 50.00 | 150.00 | 200.00 |
| PLT (109/l) | 0.0 | 0.0 | 3000.0 | 3000.00 |
| NRBC (/100 WBC) | 0.00 | – | 600.00 | – |
| Retic (%) | 0.000 | – | 30.000 | – |
WBC, white blood cell; PLT, platelet; NRBC, nucleated red blood cell.
Figure 1Leukocyte enumeration. (a) Leukocyte counts for all samples collected (n = 264) on the DxH 800 and LH 780 vs. flow cytometry (FCM). r2 Values for DxH 800 vs. LH 780 and DxH 800 vs. FCM were 0.998 and 0.994, respectively. (b) Leukocyte counts in the presence of cellular interference on the DxH 800 compared to the LH 780 and FCM. r2 Values for DxH 800 vs. LH 780 and DxH 800 vs. FCM were 0.996 and 0.990, respectively. (c) Leukocyte counts in the presence of nucleated red blood cells (manually confirmed) on the DxH 800 compared to LH 780 and FCM. r2 Values for DxH 800 vs. FCM and LH 780 vs. FCM were 0.998 and 0.992, respectively.
Accuracy P values, bias (mean difference), and % mean difference for the DxH 800 cellular analysis system vs. the LH 780 and flow cytometry (FCM)
| DxH 800 | DxH 800 | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Parameter | Subset | Bias | % Diff | Bias | % Diff | ||||
| WBC | None | 0.762 | 0.858 | 0.51 | 2 | 0.693 | 0.754 | 0.59 | 6 |
| WBC | Cellular Interference | 0.782 | 0.799 | 0.42 | 4 | 0.654 | 0.642 | 0.57 | 8 |
| WBC | NRBC | 0.879 | 0.931 | 0.69 | 3 | 0.736 | 0.762 | 0.73 | 6 |
| PLT | None | 0.716 | 0.625 | −8 | −7 | 0.510 | 0.657 | −16 | −9 |
| PLT | <50 × 109/l | 0.833 | 0.24 | 24 | 0.831 | −0.07 | −7 | ||
| PLT | Giant Platelets | 0.784 | 0.773 | −27 | −4 | 0.276 | 0.640 | −12 | −17 |
| PLT | Microcytosis | 0.812 | 0.605 | −12 | −4 | 0.640 | 0.987 | 1 | −5 |
WBC, white blood cell; PLT, platelet; NRBC, nucleated red blood cell.
Evaluation of the DxH 800 automated RBC correction compared to manual RBC correction on high white blood cell (WBC) counts
| Red cell count (×1012/l) | Hemoglobin (g/l) | Mean cell volume (fl) | RDW (%) | |||||
|---|---|---|---|---|---|---|---|---|
| WBC (×109/l) | DxH | Manual | DxH | Manual | DxH | Manual | DxH | Manual |
| 161.2 | 2.45 | 2.42 | 84.0 | 81.0 | 99.69 | 97.5 | 14.56 | 14.7 |
| 101.8 | 3.23 | 3.16 | 96.9 | 100.0 | 100.56 | 98.3 | 17.26 | 18.1 |
| 114.3 | 3.12 | 2.90 | 104.1 | 102.0 | 109.16 | 102.4 | 15.01 | 15.5 |
| 149.1 | 4.00 | 3.84 | 122.0 | 122.0 | 92.20 | 88.3 | 16.29 | 16.6 |
| 137.1 | 4.08 | 3.79 | 123.4 | 119.0 | 98.06 | 92.0 | 14.08 | 14.1 |
| 120.1 | 3.72 | 3.47 | 112.0 | 114.0 | 99.03 | 91.4 | 13.48 | 13.6 |
| 108.1 | 2.74 | 2.73 | 77.8 | 83.0 | 86.00 | 83.8 | 15.58 | 16.2 |
| 208.6 | 3.35 | 3.33 | 110.8 | 108.0 | 98.16 | 95.5 | 15.58 | 15.1 |
Figure 2Platelet (PLT) enumeration. (a) PLT counts for all samples collected (n = 257) on the DxH 800 and LH 780 vs. flow cytometry (FCM). r2 Values for DxH 800 vs. LH 780 and DxH 800 vs. FCM were 0.996 and 0.985, respectively. (b) PLT counts in the presence of megathrombocytes (manually confirmed) on DxH 800 and LH 780 compared to the reference FCM. r2 Values for DxH 800 vs. LH 780 and DxH 800 vs. FCM were 0.996 and 0.974, respectively. (c) PLT counts in the presence of microcytosis (manually confirmed) on the DxH 800 vs. LH 780 compared to the reference FCM. r2 Values for DxH 800 vs. LH 780 and DxH 800 vs. FCM were 0.998 and 0.966, respectively.
Platelet (PLT) sensitivity and specificity values for the DxH 800 vs. flow cytometry
| PLT count(×109/l) | Sensitivity TP/(TP + FN) | Specificity TN/(TN + FP) | False negative |
|---|---|---|---|
| <10 | 92.3% | 100% | 1/13 |
| <20 | 83.3% | 97.6% | 5/30 |
| <50 | 96.6% | 95.1% | 2/58 |
This single value has a value of 10.3.
All samples were flagged for review by the DxH 800.
These two samples only showed a small difference in PLT count.
Figure 3Nucleated red blood cell (NRBC) enumeration. (a) NRBC counts for all samples collected (n = 201) on the DxH 800 vs. flow cytometry (FCM). r2 Value for DxH 800 vs. FCM was 0.910. (b) Receiver Operator Curves for DxH 800 vs. FCM. Sensitivity for DxH 800 vs. FCM was 73% and specificity for DxH 800 vs. FCM was 100%.
Nucleated red blood cell false positive and false negative rates, sensitivity, specificity, and biases on the LH 780 and DxH 800
| LH 780 | DxH 800 | |
|---|---|---|
| FP rate | 3% | 0% |
| FN rate | 10% | 6% |
| Sensitivity | 56% | 73% |
| Specificity | 96% | 100% |
| Bias | −0.29 | −0.60 |