Performance Evaluation of Fully Automated 5-Part Differential Hematology Analyzer ELite 580 (ERBA).

Shanaz Khodaiji1, Dia Mansukhani1.   

Abstract

The ELite 580 (Erba Lachema) Hematology Analyzer is a newly developed automated hematology analyzer that uses impedance and flow cytometry technologies. The analyzer provides 25 reportable parameters including a 5-part differential and additionally 4 research parameters with a throughput of 80 samples per hour. To evaluate the performance of the fully automated 5-part differential hematology analyzer ELite 580 as per the international standards and correlate the results with a reference analyser, LH 780 from Beckman Coulter. We evaluated the analyzer's performance by determining background noise, carryover, precision, linearity and correlation of complete blood cell count parameters between the ELite 580 from Erba Lachema and the LH 780 from Beckman Coulter at a tertiary care hospital, according to the Clinical and Laboratory Standards Institute guidelines. The ELite 580 showed minimal background noise and carryover, high precision, accuracy and linearity over a wide analytical range for white blood cells, hemoglobin, red blood cells, hematocrit, and platelet parameters. Correlation between the ELite 580 and the LH 780 was good for all parameters (R2 > 0.90) except for mean corpuscular hemoglobin concentration (MCHC) and the basophil count. ELite 580 Hematology Analyzer with its precise and accurate results and ease of operation is suitable for clinical use in medium to large sized hematology laboratories. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s12288-021-01423-y. © Indian Society of Hematology and Blood Transfusion 2021.

Entities:  

Keywords:  Background check; Calibration; Carryover; Linearity; Precision

Year:  2021        PMID: 35125717      PMCID: PMC8804059          DOI: 10.1007/s12288-021-01423-y

Source DB:  PubMed          Journal:  Indian J Hematol Blood Transfus        ISSN: 0971-4502            Impact factor:   0.900


  8 in total

Review 1.  Possibilities and limitations of automated platelet counting procedures in the thrombocytopenic range.

Authors:  D Kunz
Journal:  Semin Thromb Hemost       Date:  2001-06       Impact factor: 4.180

2.  Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion.

Authors:  H C Segal; C Briggs; S Kunka; A Casbard; P Harrison; S J Machin; M F Murphy
Journal:  Br J Haematol       Date:  2005-02       Impact factor: 6.998

Review 3.  Continuing developments with the automated platelet count.

Authors:  C Briggs; P Harrison; S J Machin
Journal:  Int J Lab Hematol       Date:  2007-04       Impact factor: 2.877

4.  Performance evaluation of Abbott CELL-DYN Ruby for routine use.

Authors:  T Lehto; P Hedberg
Journal:  Int J Lab Hematol       Date:  2008-10       Impact factor: 2.877

5.  Performance evaluation of the Sysmex XS-1000i automated haematology analyser.

Authors:  T Ghys; R Malfait; J VAN den Bossche
Journal:  Int J Lab Hematol       Date:  2009-06-18       Impact factor: 2.877

6.  Performance evaluation of the Celltac F haematology analyser.

Authors:  I Longair; C Briggs; S J Machin
Journal:  Int J Lab Hematol       Date:  2011-01-26       Impact factor: 2.877

7.  Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. International Council for Standardization in Haematology: prepared by the ICSH Expert Panel on Cytometry.

Authors: 
Journal:  Clin Lab Haematol       Date:  1994-06

8.  Performance evaluation of Samsung LABGEO(HC10) Hematology Analyzer.

Authors:  Il Joong Park; Sunhyun Ahn; Young In Kim; Seon Joo Kang; Sung Ran Cho
Journal:  Arch Pathol Lab Med       Date:  2014-08       Impact factor: 5.534

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.