Design and fabrication of phantoms using stereolithography for small-animal imaging systems.

Mi-Ae Park1, Robert E Zimmerman, Andrew Taberner, Michael W Kaye, Stephen C Moore.   

Abstract

PURPOSE: We have investigated a new technology for fabricating phantoms with fine details for use in small-animal imaging. PROCEDURES: We used a high-resolution, 3-D stereolithography (SL) system to produce performance-evaluation phantoms such as cold-rod Derenzo, hot-channel Derenzo, and Defrise phantoms. SL performance was estimated by measuring the dimensions of many structures using a microscope. We also evaluated the degree of water absorption by two different SL resins, Somos 11120 and Accura 40, after curing.
RESULTS: The average bias and precision of the cold-rod structures over the size range 0.5 to 1.0 mm, were -0.4% and 1.74%, respectively. The water absorption study showed that Somos 11120 is a more suitable material for nuclear medicine applications.
CONCLUSION: We have demonstrated that SL is a robust and accurate method for fabrication of phantoms for small-animal imaging systems.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18546047      PMCID: PMC2848074          DOI: 10.1007/s11307-008-0150-7

Source DB:  PubMed          Journal:  Mol Imaging Biol        ISSN: 1536-1632            Impact factor:   3.488


  7 in total

1.  Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications.

Authors:  Jennifer R Stickel; Jinyi Qi; Simon R Cherry
Journal:  J Nucl Med       Date:  2007-01       Impact factor: 10.057

2.  Accuracy of treatment planning based on stereolithography in computer assisted surgery.

Authors:  Kurt Schicho; Michael Figl; Rudolf Seemann; Rolf Ewers; J Thomas Lambrecht; Arne Wagner; Franz Watzinger; Arnulf Baumann; Franz Kainberger; Julia Fruehwald; Clemens Klug
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

3.  Stereolithography of three-dimensional bioactive poly(ethylene glycol) constructs with encapsulated cells.

Authors:  Karina Arcaute; Brenda K Mann; Ryan B Wicker
Journal:  Ann Biomed Eng       Date:  2006-08-01       Impact factor: 3.934

4.  Coronary arteries: new design for three-dimensional arterial phantoms.

Authors:  C P Renaudin; B Barbier; R Roriz; D Revel; M Amiel
Journal:  Radiology       Date:  1994-02       Impact factor: 11.105

5.  Emission computer assisted tomography with single-photon and positron annihilation photon emitters.

Authors:  T F Budinger; S E Derenzo; G T Gullberg; W L Greenberg; R H Huesman
Journal:  J Comput Assist Tomogr       Date:  1977-01       Impact factor: 1.826

6.  Submillimeter total-body murine imaging with U-SPECT-I.

Authors:  Brendan Vastenhouw; Freek Beekman
Journal:  J Nucl Med       Date:  2007-03       Impact factor: 10.057

7.  CT-guided stereolithography as a new tool in craniofacial surgery.

Authors:  H Anderl; D Zur Nedden; W Mühlbauer; K Twerdy; E Zanon; K Wicke; R Knapp
Journal:  Br J Plast Surg       Date:  1994-01
  7 in total
  1 in total

1.  A fillable micro-hollow sphere lesion detection phantom using superposition.

Authors:  Frank P Difilippo; Sven L Gallo; Ryan S Klatte; Sagar Patel
Journal:  Phys Med Biol       Date:  2010-08-24       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.