Literature DB >> 18460749

Dose reduction for kilovotage cone-beam computed tomography in radiation therapy.

Jing Wang1, Tianfang Li, Zhengrong Liang, Lei Xing.   

Abstract

Kilovotage cone-beam computed tomography (kV-CBCT) has shown potentials to improve the accuracy of a patient setup in radiotherapy. However, daily and repeated use of CBCT will deliver high extra radiation doses to patients. One way to reduce the patient dose is to lower mAs when acquiring projection data. This, however, degrades the quality of low mAs CBCT images dramatically due to excessive noises. In this work, we aim to improve the CBCT image quality from low mAs scans. Based on the measured noise properties of the sinogram, a penalized weighted least-squares (PWLS) objective function was constructed, and the ideal sinogram was then estimated by minimizing the PWLS objection function. To preserve edge information in the projection data, an anisotropic penalty term was designed using the intensity difference between neighboring pixels. The effectiveness of the presented algorithm was demonstrated by two experimental phantom studies. Noise in the reconstructed CBCT image acquired with a low mAs protocol was greatly suppressed after the proposed sinogram domain image processing, without noticeable sacrifice of the spatial resolution.

Entities:  

Mesh:

Year:  2008        PMID: 18460749     DOI: 10.1088/0031-9155/53/11/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  28 in total

1.  Three-dimensional anisotropic adaptive filtering of projection data for noise reduction in cone beam CT.

Authors:  Andreas Maier; Lars Wigstrom; Hannes G Hofmann; Joachim Hornegger; Lei Zhu; Norbert Strobel; Rebecca Fahrig
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

2.  Low-dose computed tomography image restoration using previous normal-dose scan.

Authors:  Jianhua Ma; Jing Huang; Qianjin Feng; Hua Zhang; Hongbing Lu; Zhengrong Liang; Wufan Chen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT.

Authors:  Dimitrios Lazos; Jeffrey F Williamson
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

4.  Iterative image reconstruction for CBCT using edge-preserving prior.

Authors:  Jing Wang; Tianfang Li; Lei Xing
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

5.  Noise suppression in scatter correction for cone-beam CT.

Authors:  Lei Zhu; Jing Wang; Lei Xing
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

6.  Towards the clinical implementation of iterative low-dose cone-beam CT reconstruction in image-guided radiation therapy: cone/ring artifact correction and multiple GPU implementation.

Authors:  Hao Yan; Xiaoyu Wang; Feng Shi; Ti Bai; Michael Folkerts; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

7.  Structure-adaptive CBCT reconstruction using weighted total variation and Hessian penalties.

Authors:  Qi Shi; Nanbo Sun; Tao Sun; Jing Wang; Shan Tan
Journal:  Biomed Opt Express       Date:  2016-08-09       Impact factor: 3.732

8.  Statistical CT noise reduction with multiscale decomposition and penalized weighted least squares in the projection domain.

Authors:  Shaojie Tang; Xiangyang Tang
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

9.  Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

Authors:  A Sisniega; W Zbijewski; J W Stayman; J Xu; K Taguchi; E Fredenberg; Mats Lundqvist; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-11-27       Impact factor: 3.609

10.  Adaptive-weighted total variation minimization for sparse data toward low-dose x-ray computed tomography image reconstruction.

Authors:  Yan Liu; Jianhua Ma; Yi Fan; Zhengrong Liang
Journal:  Phys Med Biol       Date:  2012-11-15       Impact factor: 3.609

View more