Accuracy of continuous central venous oxygen saturation monitoring in patients undergoing cardiac surgery.

Werner Baulig1, Alexander Dullenkopf, Andreas Kobler, Barbara Baulig, Hans Rudolf Roth, Edith R Schmid.   

Abstract

OBJECTIVE: Continuous assessment of central venous oxygen saturation (S(cevox)O(2)) with the CeVOX device (Pulsion Medical Systems, Munich, Germany) was evaluated against central venous oxygen saturation (S(cv)O(2)) determined by co-oximetry.
METHODS: In 20 cardiac surgical patients, a CeVOX fiberoptic probe was introduced into a standard central venous catheter placed in the right internal jugular vein and advanced 2-3 cm beyond the catheter tip. After in vivo calibration of the probe, S(cevox)O(2), S(cv)O(2), mixed venous oxygen saturation (S(mv)O(2)) haemoglobin (Hb), body temperature, heart rate, central venous and mean arterial pressure, and cardiac index were assessed simultaneously at 30 min intervals during surgery and at 60 min intervals during recovery in the intensive care unit. Agreement between S(cevox)O(2), and S(cv)O(2) was determined by Bland-Altman analysis. Simple regression analysis was used to assess the correlation of S(cevox)O(2), and S(cv)O(2) to Hb, body temperature and haemodynamic parameters.
RESULTS: Values of S(cevox)O(2) and S(cv)O(2) (84 data pairs during surgery and 106 in the intensive care unit) ranged between 45-89% and 43-90%, respectively. Mean bias and limits of agreement of S(cevox)O(2) and S(cv)O(2) were -0.9 (-7.9/+6.1)% during surgery and -1.2 (-10.5/+8.1)% in the intensive care unit. In 37.9% of all measured data pairs, the difference between S(cevox)O(2) and S(cv)O(2) was beyond clinically acceptable limits (> or =1 s.d.). Mean bias was significantly influenced by cardiac index. Sensitivity and specificity of S(cevox)O(2) to detect substantial (> or =1 s.d.) changes in S(cv)O(2) were 89 and 82%, respectively.
CONCLUSIONS: In adult patients during and after cardiac surgery, the current version of the CeVOX device might not be the tool to replace S(cv)O(2) determined by co-oxymetry, although sensitivity and specificity of S(cevox)O(2 )to predict substantial changes in S(cv)O(2) were acceptable.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18443743     DOI: 10.1007/s10877-008-9123-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  22 in total

Review 1.  Central venous oxygen saturation monitoring in the critically ill patient.

Authors:  E P Rivers; D S Ander; D Powell
Journal:  Curr Opin Crit Care       Date:  2001-06       Impact factor: 3.687

Review 2.  Ischemia--a coagulation problem?

Authors:  B D Spiess
Journal:  J Cardiovasc Pharmacol       Date:  1996       Impact factor: 3.105

3.  Continuous measurement of oxygen saturation in sick newborn infants.

Authors:  A R Wilkinson; R H Phibbs; G A Gregory
Journal:  J Pediatr       Date:  1978-12       Impact factor: 4.406

4.  Central venous and mixed venous oxygen saturation in critically ill patients.

Authors:  C Ladakis; P Myrianthefs; A Karabinis; G Karatzas; T Dosios; G Fildissis; J Gogas; G Baltopoulos
Journal:  Respiration       Date:  2001       Impact factor: 3.580

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Importance of the sampling site for measurement of mixed venous oxygen saturation in shock.

Authors:  J D Edwards; R M Mayall
Journal:  Crit Care Med       Date:  1998-08       Impact factor: 7.598

7.  Comparison of central-venous to mixed-venous oxygen saturation during changes in oxygen supply/demand.

Authors:  K Reinhart; T Rudolph; D L Bredle; L Hannemann; S M Cain
Journal:  Chest       Date:  1989-06       Impact factor: 9.410

8.  A laboratory comparison of three pulmonary artery oximetry catheters.

Authors:  P E Scuderi; D A MacGregor; D L Bowton; R L James
Journal:  Anesthesiology       Date:  1994-07       Impact factor: 7.892

9.  Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill.

Authors:  Konrad Reinhart; Hans-Jörg Kuhn; Christiane Hartog; Donald L Bredle
Journal:  Intensive Care Med       Date:  2004-06-09       Impact factor: 17.440

10.  A comparison of two pulmonary artery mixed venous oxygen saturation catheters during the changing conditions of cardiac surgery.

Authors:  B R Hecker; D L Brown; D Wilson
Journal:  J Cardiothorac Anesth       Date:  1989-06
View more
  4 in total

1.  Prospective evaluation of regional oxygen saturation to estimate central venous saturation in sepsis.

Authors:  Christian Koch; Rainer Röhrig; Tobias Monz; Andreas Hecker; Florian Uhle; Emanuel Schneck; Markus A Weigand; Christoph Lichtenstern
Journal:  J Clin Monit Comput       Date:  2015-03-11       Impact factor: 2.502

2.  Do ScvO2 variations induced by passive leg raising predict fluid responsiveness? A prospective study.

Authors:  Raphaël Giraud; Bojana Vujovic; Benjamin Assouline; Ivo Neto Silva; Karim Bendjelid
Journal:  Physiol Rep       Date:  2021-09

3.  Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values.

Authors:  Alexander Herner; Bernhard Haller; Ulrich Mayr; Sebastian Rasch; Lea Offman; Roland Schmid; Wolfgang Huber
Journal:  PLoS One       Date:  2018-04-17       Impact factor: 3.240

Review 4.  Optimization of the target strategy of perioperative infusion therapy based on monitoring data of central hemodynamics in order to prevent complications.

Authors:  Dmytro Dmytriiev; Oleksandr Nazarchuk; Mykola Melnychenko; Bohdan Levchenko
Journal:  Front Med (Lausanne)       Date:  2022-10-03