Dose accuracy and injection force of different insulin glargine pens.

Arnd Friedrichs1, Janine Bohnet, Volker Korger, Steffen Adler, Manfred Schubert-Zsilavecz, Mona Abdel-Tawab.   

Abstract

BACKGROUND: Dose accuracy and injection force, representing key parameters of insulin pens, were determined for three pens delivering insulin glargine-based copies, Pen Royale (WR) and DispoPen (WD) for Glaritus® (Wockhardt) and GanLee Pen (GL) for Basalin® (Gan & Lee), compared with pens of the originator, ClikSTAR® (CS) and SoloSTAR® (SS) for Lantus® (Sanofi).
METHODS: Using the weighing procedure recommended by DIN EN ISO 11608-1:2000, dose accuracy was evaluated based on nonrandomized delivery of low (5 U), mid (30 U), and high (60 U) dosage levels. Injection force was measured by dispensing the maximum dose of insulin (60 U for the GL, WR, and WD; 80 U for the SS and CS) at dose speeds of 6 and 10 U/s.
RESULTS: All tested pens delivered comparable average doses within the DIN EN ISO 11608-1:2000 limits at all dosage levels. The GL revealed a higher coefficient of variation (CV) at 5 U, and the WR and WD had higher CVs at all dosage levels compared with the CS and SS. Injection force was higher for the WR, WD, and GL compared with the CS and SS at both dose speeds. In contrast to the CS and SS with an end-of-content feature, doses exceeding the remaining insulin could be dialed with the WR, GL, and WD and, apparently, dispensed with the WD.
CONCLUSIONS: All pens fulfilled the dose accuracy requirements defined by DIN EN ISO 11608-1:2000 standards at all three dosage levels, with the WR, WD, and GL showing higher dosage variability and injection force compared with the SS and CS. Thus, the devices that deliver insulin glargine copies show different performance characteristics compared with the originator.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24124963      PMCID: PMC3876380          DOI: 10.1177/193229681300700526

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

1.  Study on the dosing accuracy of commonly used disposable insulin pens.

Authors:  Meike Krzywon; Thomas van der Burg; Uwe Fuhr; Manfred Schubert-Zsilavecz; Mona Abdel-Tawab
Journal:  Diabetes Technol Ther       Date:  2012-06-13       Impact factor: 6.118

2.  Dosing accuracy of two insulin pre-filled pens.

Authors:  Toshinari Asakura; Hiroaki Seino; Miho Kageyama; Noriaki Yohkoh
Journal:  Curr Med Res Opin       Date:  2008-04-08       Impact factor: 2.580

3.  Hand syndromes associated with diabetes: impairments and obesity predict disability.

Authors:  Christine L Redmond; Gregory I Bain; Laura L Laslett; Julian D McNeil
Journal:  J Rheumatol       Date:  2009-11-02       Impact factor: 4.666

4.  The effects of the diabetes related soft tissue hand lesions and the reduced hand strength on functional disability of hand in type 2 diabetic patients.

Authors:  Serpil Savaş; Banu Kale Köroğlu; Hasan Rifat Koyuncuoğlu; Ertuğrul Uzar; Hakan Celik; Numan Mehmet Tamer
Journal:  Diabetes Res Clin Pract       Date:  2006-12-01       Impact factor: 5.602

5.  Preference for insulin delivery systems among current insulin users and nonusers.

Authors:  Kent H Summers; Sheryl L Szeinbach; Sheila M Lenox
Journal:  Clin Ther       Date:  2004-09       Impact factor: 3.393

6.  Medication adherence and the associated health-economic impact among patients with type 2 diabetes mellitus converting to insulin pen therapy: an analysis of third-party managed care claims data.

Authors:  Won Chan Lee; Sanjeev Balu; David Cobden; Ashish V Joshi; Chris L Pashos
Journal:  Clin Ther       Date:  2006-10       Impact factor: 3.393

Review 7.  Reusable and disposable insulin pens for the treatment of diabetes: understanding the global differences in user preference and an evaluation of inpatient insulin pen use.

Authors:  Riccardo Perfetti
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

8.  Development of the SoloSTAR insulin pen device: design verification and validation.

Authors:  Andreas Bode
Journal:  Expert Opin Drug Deliv       Date:  2009-01       Impact factor: 6.648

Review 9.  Insulin delivery with FlexPen: dose accuracy, patient preference and adherence.

Authors:  A Pfützner; T Asakura; B Sommavilla; Wc Lee
Journal:  Expert Opin Drug Deliv       Date:  2008-08       Impact factor: 6.648

Review 10.  Limitation of finger joint mobility in diabetes mellitus.

Authors:  A L Rosenbloom
Journal:  J Diabet Complications       Date:  1989 Apr-Jun
View more
  8 in total

Review 1.  Boluses in Insulin Therapy.

Authors:  Ralph Ziegler; Guido Freckmann; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-07-10

2.  Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen.

Authors:  David Klonoff; Irina Nayberg; Ivana Rabbone; Catherine Domenger; Udo Stauder; Hamid Oualali; Thomas Danne
Journal:  J Diabetes Sci Technol       Date:  2015-01-28

Review 3.  Half-Unit Insulin Pens: Disease Management in Patients With Diabetes Who Are Sensitive to Insulin.

Authors:  David C Klonoff; Irina Nayberg; Udo Stauder; Hamid Oualali; Catherine Domenger
Journal:  J Diabetes Sci Technol       Date:  2016-11-28

4.  Clinical Use and Evaluation of Insulin Pens.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2015-08-31

5.  Dialing torque preferences of people with diabetes when using insulin pens: a pilot study.

Authors:  Arnd Friedrichs; Mario Schmitz; Stefan Kamlot; Steffen Adler
Journal:  Diabetes Ther       Date:  2015-01-27       Impact factor: 2.945

6.  Development of devices for self-injection: using tribological analysis to optimize injection force.

Authors:  Jakob Lange; Leos Urbanek; Stefan Burren
Journal:  Med Devices (Auckl)       Date:  2016-05-11

Review 7.  A review of reusable insulin pens and features of TouStar-a new reusable pen with a dedicated cartridge.

Authors:  Robert Veasey; Carolin A Ruf; Dmitri Bogatirsky; Jukka Westerbacka; Arnd Friedrichs; Mona Abdel-Tawab; Steffen Adler; Senthilnathan Mohanasundaram
Journal:  Diabetol Metab Syndr       Date:  2021-12-19       Impact factor: 3.320

8.  Open-Label, Randomized, Two-Way, Crossover Study Assessing the Bioequivalence of the Liquid Formulation versus the Freeze-Dried Formulation of Recombinant Human FSH and Recombinant Human LH in a Fixed 2:1 Combination (Pergoveris®) in Pituitary-Suppressed Healthy Women.

Authors:  Wilhelmina Bagchus; Özkan Yalkinoglu; Peter Wolna
Journal:  Front Endocrinol (Lausanne)       Date:  2018-01-11       Impact factor: 5.555

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.