Literature DB >> 18386846

Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements-An in vitro study.

Ian Leslie1, Sophie Williams, Chris Brown, Graham Isaac, Zhongmin Jin, Eileen Ingham, John Fisher.   

Abstract

Hip resurfacing arthroplasty has become a popular alternative to conventional hip surgery. Surface replacements with bearing sizes of 55 mm (n = 5) and 39 mm (n = 5) were tested in a hip simulator for 15 million cycles (Mc). Wear debris was isolated from the serum lubricant and characterized by field emmitting gun scanning electorn microscopy, and ion levels were measured via inductively coupled plasma mass spectroscopy at intervals throughout the test. The 39 mm bearings showed significantly greater bedding in volumetric wear (2.58 mm(3)) compared with the 55 mm bearings (1.15 mm(3)). There was no significant difference between the steady state wear rates (1-15 Mc) between the two sizes (0.10 and 0.09 mm(3)/Mc, respectively); however, this parity only became clear after 7 Mc. The wear debris isolated was oval in morphology with a mean particle size of 28 nm and a range of 9-108 nm. The Co levels measured at 0.13 Mc were significantly greater than at 3.6 Mc for both bearing sizes (10926 ppb and 176 ppb, respectively). After 0.5 Mc, the Co levels from the 39 mm bearings were significantly higher than the 55 mm (11,007 vs. 1475 ppb). The wear results support previous findings showing that increasing the femoral head size decreased volumetric bedding in wear. The ion levels measured suggest both bearing sizes have similar initial wear rates; however, the 55 mm bearings reach steady state wear more rapidly. (c) 2008 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18386846     DOI: 10.1002/jbm.b.31087

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  21 in total

1.  The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles.

Authors:  Fabrizio Billi; Paul Benya; Aaron Kavanaugh; John Adams; Harry McKellop; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  High cup angle and microseparation increase the wear of hip surface replacements.

Authors:  Ian J Leslie; Sophie Williams; Graham Isaac; Eileen Ingham; John Fisher
Journal:  Clin Orthop Relat Res       Date:  2009-04-11       Impact factor: 4.176

3.  Squeaking: Current knowledge and how to avoid it.

Authors:  Arjuna M Imbuldeniya; Simon J Pearce; William L Walter; Bernard A Zicat; William K Walter
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

4.  Edge-loading severity as a function of cup lip radius in metal-on-metal total hips--a finite element analysis.

Authors:  Jacob M Elkins; Karen M Kruger; Douglas R Pedersen; John J Callaghan; Thomas D Brown
Journal:  J Orthop Res       Date:  2011-08-02       Impact factor: 3.494

5.  The withdrawn ASR™ THA and hip resurfacing systems: how have our patients fared over 1 to 6 years?

Authors:  Kevin T Hug; Tyler S Watters; Thomas P Vail; Michael P Bolognesi
Journal:  Clin Orthop Relat Res       Date:  2013-02       Impact factor: 4.176

6.  What is the midterm survivorship and function after hip resurfacing?

Authors:  Luthfur Rahman; Sarah K Muirhead-Allwood; Muhannad Alkinj
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

7.  Acute inflammatory response to cobalt chromium orthopaedic wear debris in a rodent air-pouch model.

Authors:  Moeed Akbar; Alasdair R Fraser; Gerard J Graham; James M Brewer; M Helen Grant
Journal:  J R Soc Interface       Date:  2012-04-18       Impact factor: 4.118

8.  Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Authors:  Ian C Clarke; Jean-Yves Lazennec; Adrien Brusson; Christina Savisaar; John G Bowsher; Michelle Burgett; Thomas K Donaldson
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

9.  Conversion of a failed hip resurfacing arthroplasty to total hip arthroplasty: pearls and pitfalls.

Authors:  Jacob A Haynes; Jeffrey B Stambough; Robert L Barrack; Denis Nam
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

Review 10.  Review on squeaking hips.

Authors:  Yadin David Levy; Selin Munir; Shane Donohoo; William Lindsay Walter
Journal:  World J Orthop       Date:  2015-11-18
View more