Reproducibility and clinical relevance of the ocular response analyzer in nonoperated eyes: corneal biomechanical and tonometric implications.

Javier Moreno-Montañés1, Miguel J Maldonado, Noelia García, Loreto Mendiluce, Pio J García-Gómez, María Seguí-Gómez.   

Abstract

PURPOSE: To assess the reproducibility of the ocular response analyzer (ORA) in nonoperated eyes and the impact of corneal biomechanical properties on intraocular pressure (IOP) measurements in normal and glaucomatous eyes.
METHODS: In the reliability study, two independent examiners obtained repeated ORA measurements in 30 eyes. In the clinical study, the examiners analyzed ORA and IOP-Goldmann values from 220 normal and 42 glaucomatous eyes. In both studies, Goldmann-correlated IOP measurement (IOP-ORAg), corneal-compensated IOP (IOP-ORAc), corneal hysteresis (CH), and corneal resistance factor (CRF) were evaluated. IOP differences of 3 mm Hg or greater between the IOP-ORAc and IOP-ORAg were considered outcome significant.
RESULTS: Intraexaminer intraclass correlation coefficients and interexaminer concordance correlation coefficients ranged from 0.78 to 0.93 and from 0.81 to 0.93, respectively, for all parameters. CH reproducibility was highest, and the IOP-ORAg readings were lowest. The median IOP was 16 mm Hg with the Goldmann tonometer, 14.5 mm Hg with IOP-ORAg (P < 0.001), and 15.7 mm Hg with IOP-ORAc (P < 0.001). Outcome-significant results were found in 77 eyes (29.38%). The IOP-ORAc, CH, and CRF were correlated with age (r = 0.22, P = 0.001; r = -0.23, P = 0.001; r = -0.14, P = 0.02, respectively), but not the IOP-ORAg or IOP-Goldmann.
CONCLUSIONS: The ORA provides reproducible corneal biomechanical and IOP measurements in nonoperated eyes. Considering the effect of ORA, corneal biomechanical metrics produces an outcome-significant IOP adjustment in at least one quarter of glaucomatous and normal eyes undergoing noncontact tonometry. Corneal viscoelasticity (CH) and resistance (CRF) appear to decrease minimally with increasing age in healthy adults.

Entities:  

Mesh:

Year:  2008        PMID: 18326720     DOI: 10.1167/iovs.07-0280

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  26 in total

1.  Evaluation of corneal biomechanical properties following penetrating keratoplasty using the ocular response analyzer.

Authors:  Joo Young Shin; Jin Seok Choi; Joo Youn Oh; Mee Kum Kim; Jin Hak Lee; Won Ryang Wee
Journal:  Korean J Ophthalmol       Date:  2010-06-05

2.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

3.  Intraocular pressure measured by dynamic contour tonometer and ocular response analyzer in normal tension glaucoma.

Authors:  Tetsuya Morita; Nobuyuki Shoji; Kazutaka Kamiya; Mana Hagishima; Fusako Fujimura; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-20       Impact factor: 3.117

4.  Corneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas.

Authors:  Gabor Nemeth; Eszter Szalai; Ziad Hassan; Agnes Lipecz; Zsuzsa Flasko; Laszlo Modis
Journal:  Int J Ophthalmol       Date:  2017-02-18       Impact factor: 1.779

5.  Corneal hysteresis in patients with dry eye.

Authors:  P G Firat; S Doganay
Journal:  Eye (Lond)       Date:  2011-09-09       Impact factor: 3.775

6.  Impact of dehydration and fasting on intraocular pressure and corneal biomechanics measured by the Ocular Response Analyzer.

Authors:  Betul Seher Uysal; Necati Duru; Umut Ozen; Mucella Arikan Yorgun; Emine Akcay; Mehtap Caglayan; Nurullah Cagil
Journal:  Int Ophthalmol       Date:  2017-03-02       Impact factor: 2.031

7.  Reproducibility of ocular response analyzer measurements and their correlation with central corneal thickness.

Authors:  Joanna Wasielica-Poslednik; Fatmire Berisha; Shakhsanam Aliyeva; Norbert Pfeiffer; Esther M Hoffmann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-10       Impact factor: 3.117

8.  [Influence of corneal biomechanical properties on myopic regression after laser in situ keratomileusis].

Authors:  D Uthoff; K Hebestedt; G I W Duncker; E Spörl
Journal:  Ophthalmologe       Date:  2013-01       Impact factor: 1.059

9.  Ocular biomechanical measurements on post-keratoplasty corneas using a Scheimpflug-based noncontact device.

Authors:  Laszlo Modis; Ziad Hassan; Eszter Szalai; Zsuzsanna Flaskó; Andras Berta; Gabor Nemeth
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

10.  Evaluation of biomechanical properties of the cornea in patients with primary hyperparathyroidism.

Authors:  Ali Bülent Çankaya; Seyfullah Kan; Muhammed Kizilgul; Aslı Tokmak; Merve Inanc; Mustafa Caliskan; Emrullah Beyazyildiz; Ugur Açar; Tuncay Delibaşı
Journal:  Int Ophthalmol       Date:  2016-07-16       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.