Dynamic contour tonometry in comparison to intracameral IOP measurements.

Andreas G Boehm1, Anja Weber, Lutz E Pillunat, Rainer Koch, Eberhard Spoerl.   

Abstract

PURPOSE: The purpose of the study was to examine the effect of central corneal thickness (CCT), corneal curvature, astigmatism, axial length (AL), and age on measurements with the Pascal Dynamic Contour Tonometer (DCT).
METHODS: In a prospective clinical trial 75 eyes of 75 patients undergoing phacoemulsification were examined. Before phacoemulsification, the anterior chamber was cannulated at the temporal corneal limbus. In a closed system the IOP was directly set to 15, 20, or 35 mm Hg with a manometric water column. IOP measurements taken by DCT were compared to intracameral measurements with a precision reference pressure sensor.
RESULTS: Measurements from 60 patients were suitable for statistical analysis. At IOP of 15 mm Hg, the mean difference between IOP measured by DCT and intracameral IOP was -0.02 +/- 1.32 mm Hg; at 20 mm Hg it was -0.2 +/- 1.44 mm Hg and at 35 mm Hg, -0.84 +/- 1.90 mm Hg. The concordance coefficient according to Lin was 0.9763, showing good agreement between DCT- and intracamerally measured IOP. There was a statistically significant correlation between the difference in IOP measured by DCT minus intracameral IOP and CCT (P = 0.0291, R2 = 0.00012). All other parameters had no statistically significant effect on the difference between DCT and intracameral IOP (corneal curvature, P = 0.6094, R2 = 0.00367; age, P = 0.9198, R2 = 0.000003; astigmatism, P = 0.1564, R2 = 0.08497; and axial length, P = 0.9484, R2 = 0.00008).
CONCLUSIONS: Measurements with the DCT showed good concordance with intracameral IOP. CCT exerted a statistically significant but clinically irrelevant effect on measurements with the DCT.

Entities:  

Mesh:

Year:  2008        PMID: 18316699     DOI: 10.1167/iovs.07-1366

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  33 in total

1.  [Influence of corneal thickness on intraocular pressure measurements following Descemet's stripping automated endothelial keratoplasty (DSAEK)].

Authors:  M K J Klamann; A-K B Maier; J Gonnermann; N Torun; P C Ruokonen
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

2.  Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery.

Authors:  Antonios P Aristeidou; Georgios Labiris; Andreas Katsanos; Michalis Fanariotis; Nikitas C Foudoulakis; Vassilios P Kozobolis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-06-25       Impact factor: 3.117

3.  Evaluation of ocular pulse amplitude and choroidal thickness in diabetic macular edema.

Authors:  Y Totan; T K Akyüz; E Güler; F B Güragaç
Journal:  Eye (Lond)       Date:  2015-11-13       Impact factor: 3.775

Review 4.  [Measuring intraocular pressure by different methods].

Authors:  J Lamparter; E M Hoffmann
Journal:  Ophthalmologe       Date:  2009-08       Impact factor: 1.059

5.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

6.  Non-invasive measurement of choroidal volume change and ocular rigidity through automated segmentation of high-speed OCT imaging.

Authors:  L Beaton; J Mazzaferri; F Lalonde; M Hidalgo-Aguirre; D Descovich; M R Lesk; S Costantino
Journal:  Biomed Opt Express       Date:  2015-04-13       Impact factor: 3.732

7.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

8.  Clinical evaluation of multiparameter correction equations for Goldmann applanation tonometry.

Authors:  P G Davey; A Elsheikh; D F Garway-Heath
Journal:  Eye (Lond)       Date:  2013-03-15       Impact factor: 3.775

9.  [Evaluation of correction formulas for tonometry : The Goldmann applanation tonometry in approximation to dynamic contour tonometry].

Authors:  J Wachtl; M Töteberg-Harms; S Frimmel; C Kniestedt
Journal:  Ophthalmologe       Date:  2017-08       Impact factor: 1.059

10.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.