BACKGROUND: While most reports of component fracture following alumina-on-alumina total hip arthroplasty have involved the acetabular liner, few have involved fracture of the alumina femoral head. In the present multicenter study, we investigated ceramic head fractures in a cohort of patients who underwent third-generation alumina-on-alumina total hip arthroplasty.
METHODS: We performed a retrospective study of 312 patients (367 hips) who underwent alumina-on-alumina total hip arthroplasty without cement at four participating centers with the use of a 28-mm BIOLOX forte femoral head and a BIOLOX forte liner from July 2001 to October 2003. Three hundred and five patients (359 hips) were evaluated at a mean of forty-five months postoperatively. Clinical follow-up with use of the Harris hip score and radiographic evaluation were performed at six weeks; at three, six, and twelve months; and every six months thereafter. Retrieved ceramic implants were examined by means of visual inspection.
RESULTS: Five hips (1.4%) in five patients were revised because of a ceramic head fracture during the follow-up period. The ceramic head fractures occurred during normal daily activities at a mean of 22.6 months postoperatively. A short neck had been used in all five hips in which a fracture occurred, compared with 121 (34.2%) of the 354 hips in which a fracture did not occur (p = 0.009). The fracture involved a circular crack along the circumference of the thinnest portion of the head component at the proximal edge of the bore. The fracture also involved multiple vertical cracks extending radially along the longitudinal axis from the circumference of the circular crack line to the lower edge of the head component.
CONCLUSIONS: In the present study, the rate of ceramic head fracture associated with one design of a short-neck modular alumina femoral head was 1.4% (five of 359). The extent to which these findings are generalizable to other designs that utilize this type of femoral head is unknown.

Entities:  

Mesh:

Year:  2008        PMID: 18245593     DOI: 10.2106/JBJS.F.01489

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  37 in total

1.  Squeaking: Current knowledge and how to avoid it.

Authors:  Arjuna M Imbuldeniya; Simon J Pearce; William L Walter; Bernard A Zicat; William K Walter
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

Review 2.  Surgical management of hip osteoarthritis.

Authors:  Rajiv Gandhi; Anthony V Perruccio; Nizar N Mahomed
Journal:  CMAJ       Date:  2013-10-21       Impact factor: 8.262

3.  Modular sleeves with ceramic heads in isolated acetabular cup revision in younger patients-laboratory and experimental analysis of suitability and clinical outcomes.

Authors:  Peter Helwig; Lukas Konstantinidis; Anja Hirschmüller; Anke Bernstein; Oliver Hauschild; Norbert P Südkamp; Björn G Ochs
Journal:  Int Orthop       Date:  2012-12-07       Impact factor: 3.075

4.  Ceramic-on-ceramic total hip arthroplasty: minimum of six-year follow-up study.

Authors:  Won-Sik Choy; Kap Jung Kim; Sang Ki Lee; Kyoung Wan Bae; Yoon Sub Hwang; Chang Kyu Park
Journal:  Clin Orthop Surg       Date:  2013-08-20

5.  Reduced wear of cross-linked UHMWPE using magnesia-stabilized zirconia femoral heads in a hip simulator.

Authors:  Marcel E Roy; Leo A Whiteside; Mark E Magill; Brian J Katerberg
Journal:  Clin Orthop Relat Res       Date:  2011-02-11       Impact factor: 4.176

6.  Do alumina matrix composite bearings decrease hip noises and bearing fractures at a minimum of 5 years after THA?

Authors:  Seung-Hoon Baek; Won Keun Kim; Jun Young Kim; Shin-Yoon Kim
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

7.  Low aseptic loosening and revision rate in Zweymüller-Plus total hip arthroplasty with ceramic-on-ceramic bearings.

Authors:  T Bouras; T Repantis; P Fennema; P Korovessis
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-12

8.  Primary ceramic-on-ceramic total hip arthroplasty using a 32-mm ceramic head with a titanium-alloy sleeve.

Authors:  Seung-Jae Lim; Seung-Pil Jang; Dong-Wook Kim; Young-Wan Moon; Youn-Soo Park
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

9.  THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

Authors:  William G Hamilton; James P McAuley; Douglas A Dennis; Jeffrey A Murphy; Thomas J Blumenfeld; Joel Politi
Journal:  Clin Orthop Relat Res       Date:  2009-09-19       Impact factor: 4.176

Review 10.  [The problem of fractures of ceramic heads. What should be done?].

Authors:  H R Dürr
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.