Audit on source strength determination for HDR and PDR (192)Ir brachytherapy in Sweden.

Asa Carlsson Tedgren1, Jan-Erik Grindborg.   

Abstract

BACKGROUND AND
PURPOSE: To investigate the status of source strength determination in terms of reference air kerma rate (RAKR) for HDR and PDR (192)Ir brachytherapy in Sweden.
MATERIALS AND METHODS: RAKR was determined in each of the 14 Swedish afterloaders, using calibrated equipment from the Swedish Secondary Standard Dosimetry Laboratory.
RESULTS: Values of RAKR from the external audit, the hospitals and vendors agreed within the uncertainty limits guaranteed by the vendors.
CONCLUSIONS: The accuracy in RAKR determination has increased over the last years as a result of increased availability of interpolation standards for HDR (192)Ir and the increased use of robust well-type ion chambers designed for brachytherapy. It is recommended to establish a ratio between the RAKR value from own measurements at the hospital and that of the vendor since such a ratio embeds constant systematic differences due to e.g. varying traceability and therefore has the potential of being less uncertain than the RAKR alone. Traceability to primary standards for HDR (192)Ir sources will in the future significantly decrease the uncertainty in RAKR of (192)Ir brachytherapy.

Mesh:

Substances:

Year:  2008        PMID: 18187219     DOI: 10.1016/j.radonc.2007.12.008

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  4 in total

Review 1.  Dosimetric audit in brachytherapy.

Authors:  A L Palmer; D A Bradley; A Nisbet
Journal:  Br J Radiol       Date:  2014-05-06       Impact factor: 3.039

2.  Source strength determination in iridium-192 and cobalt-60 brachytherapy: A European survey on the level of agreement between clinical measurements and manufacturer certificates.

Authors:  Javier Vijande; Åsa Carlsson Tedgren; Facundo Ballester; Dimos Baltas; Panagiotis Papagiannis; Mark J Rivard; Frank-André Siebert; Larry De Werd; Jose Perez-Calatayud
Journal:  Phys Imaging Radiat Oncol       Date:  2021-07-30

3.  Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.

Authors:  Catharine H Clark; Edwin G A Aird; Steve Bolton; Elizabeth A Miles; Andrew Nisbet; Julia A D Snaith; Russell A S Thomas; Karen Venables; David I Thwaites
Journal:  Br J Radiol       Date:  2015-09-02       Impact factor: 3.039

4.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.