Javier Vijande1,2,3, Åsa Carlsson Tedgren4,5,6, Facundo Ballester1,2, Dimos Baltas7,8, Panagiotis Papagiannis9, Mark J Rivard10, Frank-André Siebert11, Larry De Werd12, Jose Perez-Calatayud2,13. 1. Departamento de Física Atómica, Molecular y Nuclear, Universitat de Valencia (UV), Burjassot, Spain. 2. Unidad Mixta de Investigación en Radiofísica e Instrumentación Nuclear en Medicina (IRIMED), Instituto de Investigación Sanitaria La Fe (IIS-La Fe)-Universitat de Valencia (UV), Valencia, Spain. 3. Instituto de Física Corpuscular, IFIC (UV-CSIC), Burjassot, Spain. 4. Radiation Physics, Department of Medicine, Health and Caring Sciences (HMV), Linköping University, Linköping, Sweden. 5. Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Stockholm, Sweden. 6. Department of Oncology Pathology, the Karolinska Institute, Stockholm, Sweden. 7. Division of Medical Physics, Department of Radiation Oncology, Medical Center, Faculty of Medicine, University of Freiburg, Germany. 8. German Cancer Consortium (DKTK) Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany. 9. Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, 75 Mikras Asias, Athens, Greece. 10. Department of Radiation Oncology, Alpert Medical School of Brown University, Providence, USA. 11. Clinic of Radiotherapy, UniversityHospital of Schleswig-Holstein, Campus Kiel, Kiel, Germany. 12. Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, United States of America. 13. Radiotherapy Department, La Fe Hospital, Valencia, and Clinica Benidorm, Alicante, Spain.
Abstract
BACKGROUND AND PURPOSE: Brachytherapy treatment outcomes depend on the accuracy of the delivered dose distribution, which is proportional to the reference air-kerma rate (RAKR). Current societal recommendations require the medical physicist to compare the measured RAKR values to the manufacturer source calibration certificate. The purpose of this work was to report agreement observed in current clinical practice in the European Union. MATERIALS AND METHODS: A European survey was performed for high- and pulsed-dose-rate (HDR and PDR) high-energy sources (192Ir and 60Co), to quantify observed RAKR differences. Medical physicists at eighteen hospitals from eight European countries were contacted, providing 1,032 data points from 2001 to 2020. RESULTS: Over the survey period, 77% of the 192Ir measurements used a well chamber instead of the older Krieger phantom method. Mean differences with the manufacturer calibration certificate were 0.01% ± 1.15% for 192Ir and -0.1% ± 1.3% for 60Co. Over 95% of RAKR measurements in the clinic were within 3% of the manufacturer calibration certificate. CONCLUSIONS: This study showed that the agreement level was generally better than that reflected in prior societal recommendations positing 5%. Future recommendations on high-energy HDR and PDR source calibrations in the clinic may consider tightened agreements levels.
BACKGROUND AND PURPOSE: Brachytherapy treatment outcomes depend on the accuracy of the delivered dose distribution, which is proportional to the reference air-kerma rate (RAKR). Current societal recommendations require the medical physicist to compare the measured RAKR values to the manufacturer source calibration certificate. The purpose of this work was to report agreement observed in current clinical practice in the European Union. MATERIALS AND METHODS: A European survey was performed for high- and pulsed-dose-rate (HDR and PDR) high-energy sources (192Ir and 60Co), to quantify observed RAKR differences. Medical physicists at eighteen hospitals from eight European countries were contacted, providing 1,032 data points from 2001 to 2020. RESULTS: Over the survey period, 77% of the 192Ir measurements used a well chamber instead of the older Krieger phantom method. Mean differences with the manufacturer calibration certificate were 0.01% ± 1.15% for 192Ir and -0.1% ± 1.3% for 60Co. Over 95% of RAKR measurements in the clinic were within 3% of the manufacturer calibration certificate. CONCLUSIONS: This study showed that the agreement level was generally better than that reflected in prior societal recommendations positing 5%. Future recommendations on high-energy HDR and PDR source calibrations in the clinic may consider tightened agreements levels.
Brachytherapy (BT) using photon emitting sources is mainly performed using either a single high dose-rate (HDR) or pulsed dose-rate (PDR) source, or multiple low dose-rate (LDR) sources. HDR sources are, with few exceptions, of high energy (>0.05 MeV) and LDR ones of low energy (<0.05 MeV). PDR sources are used for pulsed treatments and typically have the same design as HDR ones, but with lower source strength. In LDR treatments, several permanently implanted sources (seeds) are used with every patient treatment while a single HDR or PDR source is normally used repeatedly for months or years on multiple patients. For HDR and PDR BT, 192Ir (half-life 73.8 days, mean photon energy 0.4 MeV) is the most common radionuclide while also 60Co sources (half-life 1925 days, mean photon energy 1.25 MeV) are available. The reference quantity used in Europe for BT source strength is the reference air-kerma rate (RAKR) with units cGy h−1 at 1 m [1], [2], while air-kerma strength, SK in units of U (1 U = 1 µGy m2/h = 1 cGy cm2/h) is used in North America with the numerical value of RAKR being 104 more than the numerical value of SK. RAKR plays a key role in dosimetry at the hospital level, since values of absorbed dose in absolute terms used in clinical BT treatment planning are directly proportional to it through the TG-43 formalism for dose calculation [3]. Standards for this quantity are setup and maintained within the international metrology community and requirements on traceability to such standards apply to BT similar to all other radiotherapy modalities [4]. Specifically, vendors issue BT source certificates including a determination of source strength using measuring equipment with traceability to such standards and experimental verification in the clinic of these certificates is regulated in most countries. This way, measurements of RAKR or SK constitute the BT equivalent of external-beam reference dosimetry.The established standards for dosimetric realization of RAKR differ depending on the dose rate [4], [5]. Primary standards used to realize source strength measurement for LDR sources are based on free-air ionization chambers while for HDR and PDR 192Ir sources such standards are based on a spherical graphite-walled large volume cavity ionization chamber and a lead-housing with a dedicated collimator [6], [7]. Other 192Ir standards are based on indirect methods for realization of RAKR
[8]. The primary standard at NPL determines source strength values and instrument calibration coefficients with lower uncertainty (<1%) than the indirect ones (around 3%) at k = 2 [8], [9], [10]. The PTB provides a calibration for HDR 60Co sources and reported on a quality correction factor, k, aimed to transfer a 192Ir calibration coefficient into one for 60Co as derived from measurements with 35 well-type ionization chambers of two different chamber types [11], [12]. Due in part to the logistics of HDR and PDR high-energy sources not being easily shippable and in part to the fact that high-energy sources are less sensitive to source design and manufacturing processes, a system similar to that setup for LDR sources by the AAPM does not exist [13]. It is nevertheless well recognized by the AAPM Task Group No. 56 Report [14] and by the GEC-ESTRO in the ESTRO Booklet #8 [15] that the manufacturer-issued RAKR certificate of each HDR and PDR source must be measured in the clinic using traceably-calibrated equipment. Updated GEC-ESTRO clinical recommendations for calibration traceability of HDR and PDR sources are currently in preparation, also collecting information on available resources of laboratories offering calibration services with traceability to international standards.Use of air-filled, vented well-type ionization chambers in the clinic and secondary standard laboratories has been recommended because of their robustness, stability, and simplicity in setup [16]. An alternative measurement technique is recommended by the German society for Medical Physics (DGMP), consisting of a PMMA phantom, named the Krieger-phantom, housing a thimble ionization chamber [17], [18]. Current societal recommendations establish that differences between clinic-measured RAKR (or SK) values and the manufacturer certificate should be within 5% [14], [15].Uncertainties of secondary/tertiary standard’s calibration coefficients and vendor issued source certificate are lower for HDR-PDR 192Ir and HDR 60Co than for LDR sources. Therefore, it has been suggested in the literature that the current RAKR relative difference limit of 5% could be reduced given that clinics and manufacturers respect the measurement conditions specified on the instrument calibration certificates and follow good practice protocols [19]. Additionally, differences in calibration coefficients for different types of 192Ir sources are small compared to those for low energy sources [20]. RAKR measurement corrections due to source geometry, derived using Monte Carlo calculated factors, of about 0% to 2%, may be applied to further reduce RAKR measurement differences between various 192Ir sources [21]. Such factors are not yet available for all source types (notably short length PDR sources) and all well-type chambers or the Krieger phantom setup.The current study presents the results of a survey performed by the BRAPHYQS WP21 group of GEC-ESTRO to assess the level of agreement between RAKR values as measured in the clinic to verify values reported on source manufacturer certificates for 192Ir and 60Co HDR and PDR BT sources. The survey included clinics throughout Europe, including BRAPHYQS and GEC-ESTRO committee members where HDR and PDR BT is routinely used.
Material and methods
Eighteen clinics from eight European countries were contacted to achieve enough statistics and provide basic sample stratification to avoid potential bias due to the use of a particular methodology, clinical practice, or national regulations. Data on HDR 192Ir, PDR 192Ir, and HDR 60Co sources were reported, together with general information about the clinical practice followed for each set of measurements. Data collection included changes during the period reported in instrumentation, calibration certificate, or procedure. Participating clinics were requested to submit their measured values (RAKRCLINIC) together with corresponding values on manufacturer certificates (RAKRMANU). Percentage differences between these were reported as:The number of data points thus obtained were 970 for 192Ir (294 for PDR and 676 for HDR) and 62 for 60Co over the period 2001–2020. In the case of 192Ir, the number of values obtained was large enough to recover the expected normal distribution, hence a Gaussian fit was performed. Not all participants provided the same level of detail, two clinics did not provide detailed lists of measurements, instead providing their mean, standard, and maximum deviations. Those values were combined with the corresponding ones obtained in the fit by a weighted average (mean) and weighted sum in quadrature (standard deviation). A histogram of the RAKR differences reported was produced for both radionuclides. The RAKR interval where more than 95% (k = 2 for a normal distribution) of data points resided was considered a conservative estimate of differences expected between RAKR from clinical user measurements and vendor certificates.
Results
Defining RAKR differences according to Eq. (1), mean differences for 192Ir sources of 0.01% with a standard deviation of 1.15% were found. This was for 750 and 220 clinic measurements using well-type ion chambers and Krieger phantoms, respectively. Hence, values outside 3% corresponded to less than 5% of the reported values. Although this behavior was independent of the measurement technique (well chamber or Krieger phantom datasets), Gaussian-fits performed on each dataset independently yielded standard deviations of 1.0% for the well chamber and 1.5% using the Krieger phantom (Fig. 1).
Fig. 1
Frequency distribution of percentage differences between RAKRCLINIC and RAKRMANU for 192Ir (mean = 0.01%, standard deviation = 1.15%).
Frequency distribution of percentage differences between RAKRCLINIC and RAKRMANU for 192Ir (mean = 0.01%, standard deviation = 1.15%).For 60Co BT sources, where normality of the distribution of RAKR differences could not be assumed due to the limited number of data points, all RAKR difference values were within a ± 3% interval with a mean value of –0.1% and standard deviation of 1.3% (Fig. 2).
Fig. 2
Frequency distribution of percentage differences between RAKRCLINIC and RAKRMANU for 60Co (mean = –0.1%, standard deviation = 1.3%).
Frequency distribution of percentage differences between RAKRCLINIC and RAKRMANU for 60Co (mean = –0.1%, standard deviation = 1.3%).
Discussion
Current recommendations establish that the RAKR value measured by a medical physicist during clinical practice must agree within 5% to that reported in the source calibration certificate provided by the manufacturer [9], [10]. This survey found that such a value underestimates the quality of RAKR measurements at European clinics. It is clear that in the period 2001–2020, more than 95% of the HDR/PDR source strength measurements performed in the sampled European centers agreed within 3% with the BT source manufacturer calibration certificate.The largest differences reported were 9.1% and 3.0% for 192Ir and 60Co, respectively. A reduced number of outliers were found in the data sample. Of those, only three measurements present differences larger than four standard deviations (>4.6%). One center presented larger systematic differences in older results (before 2010) compared to more recent results. If those values were removed from the analysis, the number of measurements within 3% would increase to 99%. A different methodology to approach this problem would analyze institutional results individually. Unfortunately, we were faced with long temporal series (about 20 years in some cases) where the uncertainties changed with time whenever the protocols were actualized. Therefore, it is clear that the data from every single clinic does not always correspond to randomly distributed results around a central value, and hence trying to extract a single mean value and a standard deviation for every clinic might lead to a misleading statement. An example of such pattern for one clinic is shown as Supplementary Material (Suppl. Fig. S1).There are many sources of uncertainty and errors that may have contributed to the RAKR differences observed in this study. Briefly, experimental uncertainties may be divided into two categories: those that are well known and clearly specified, the most important being calibration uncertainties stemming from measurement setup, and systematic errors in the measurement methods or an undetected equipment malfunction. These sources of uncertainty and errors are included within the RAKR differences from the previous section.There are some sources of clinic-related uncertainties that might or might not be folded into the results of the survey. The most relevant ones are differences in source type used at instrument calibration and measurements [21], [20], and sub-optimal practice or non-compliance to the conditions stated in the instrument calibration certificate. An example would be placing the well chamber close to the floor or a wall where typical enhancement of about 3% has been reported, instead of being positioned in the center of the room on a on low-scattering device with more than 30 cm from the floor or wall [19].Although a complete analysis on the protocol followed at each of the institutions participating in this survey is beyond the scope of the present manuscript, it is possible to make general comments. Well-type ion chambers are known for their long-term stability [22], but are more susceptible to room-scattering conditions than the Krieger phantom as the latter is surrounded by a significant amount of PMMA. Furthermore as both instruments contain large amounts of material (air or PMMA), it is important they have reached thermal equilibrium with the other instrumentation, i.e., thermometers and pressure gauges. It is also important to ensure the correct source position inside the well chamber or Krieger phantom. Ideally, the clinical user should maintain a historical record of previous source strength measurements to identify possible systematic error and subsequently correct said measurements.RAKR is determined with an ion chamber as , where Icorr is the measured current corrected for influence quantities and NRAKR is the ion chamber calibration coefficient. The NRAKR bears the largest contribution to the total uncertainty of RAKR measurement as it stems from the realization of the quantity at a standard laboratory, while the Icorr, measured in the clinic or by the manufacturer, contributes less. Clinics are recommended to follow the RAKR difference obtained with the manufacturer over time as such ratio can be expected to vary within the combined uncertainty of the two current determinations around a number set by possible differences in calibration coefficient determination [19], and other potential systematic uncertainties. Logically, every uncertainty budget is affected by the protocol implemented in the corresponding calibration laboratory. Such uncertainty can differ significantly across institutions. Typically, calibrations at the NPL using a primary standard are associated with reduced uncertainty (0.8% at k = 2) relative to calibrations based on indirect interpolation techniques such as 2.6% at k = 2 at the University of Wisconsin ADCL, or 3.0% k = 2 at VSL [19].A protocol used in some of the clinics participating in this survey and enforced by some particular national regulations is to measure all sources twice, once when received and a second after some time, typically a few weeks or when removed from the institution by the vendor (in both cases corrected by the corresponding radioactive decay). Such a procedure allows the user to immediately determine any possible measurement error or equipment malfunction that might have arisen in between measurements and therefore guaranties reproducibility.Summarizing, high dosimetrical accuracy is fundamental to radiation therapy. For HDR and PDR 192Ir and HDR 60Co sources, the mean difference between RAKR values measured at the hospital level and those reported in the source certificates were less than 0.1%, being more than 95% of values reported within 3%. These results will be included in the upcoming GEC-ESTRO recommendations on high energy, HDR and PDR source calibrations in the clinic.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Authors: Mark J Rivard; Bert M Coursey; Larry A DeWerd; William F Hanson; M Saiful Huq; Geoffrey S Ibbott; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson Journal: Med Phys Date: 2004-03 Impact factor: 4.071
Authors: Larry A DeWerd; M Saiful Huq; Indra J Das; Geoffrey S Ibbott; William F Hanson; Thomas W Slowey; Jeffrey F Williamson; Bert M Coursey Journal: Med Phys Date: 2004-03 Impact factor: 4.071