In vivo evaluation of the HeartWare centrifugal ventricular assist device.

Egemen Tuzun1, Katy Roberts, William E Cohn, Murat Sargin, Courtney J Gemmato, Branislav Radovancevic, O H Frazier.   

Abstract

In this study, long-term (90-day) hemocompatibility and end-organ effects of a centrifugal left ventricular assist device (the Heartware HVAD) were evaluated in 6 healthy sheep. The device was implanted into the left ventricular apex on beating hearts. The outflow graft of each device was anastomosed to the descending aorta. None of the sheep received anticoagulation or antiaggregation medication during the study. Hematologic and biochemical tests of liver and kidney function were performed pre-operatively (baseline) and throughout the study. Data associated with pump function were collected continuously until 90 +/- 1 days of support, at which time the sheep were humanely killed, and the end-organs were examined macroscopically and histopathologically. Hematologic and biochemical test results were within normal limits during the study period. There were no significant complications. Postmortem examination of the explanted organs revealed no evidence of ischemia or infarction, except in 2 sheep, in which small foci of infarction were detected in each of their left kidneys. There was no significant device failure. In all sheep, the pump's inflow and outflow conduits were free of thrombus. During the 90-day study, the HeartWare HVAD showed exceptional hemocompatibility and reliability, both of which are crucial to the clinical success of any implantable left ventricular assist device.

Entities:  

Keywords:  Animals; assisted circulation/instrumentation; blood pump, centrifugal; equipment design; heart assist device, left ventricular; hemodynamic processes; heparin; sheep; thrombosis/prevention & control

Mesh:

Substances:

Year:  2007        PMID: 18172519      PMCID: PMC2170489     

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  34 in total

Review 1.  Hemodynamics of chronic nonpulsatile flow: implications for LVAD development.

Authors:  Satoshi Saito; Tomohiro Nishinaka; Stephen Westaby
Journal:  Surg Clin North Am       Date:  2004-02       Impact factor: 2.741

2.  The effect of the impeller-driver magnetic coupling distance on hemolysis in a compact centrifugal pump.

Authors:  T Nakazawa; K Makinouchi; Y Takami; J Glueck; S Takatani; Y Nosé
Journal:  Artif Organs       Date:  1996-03       Impact factor: 3.094

3.  Mechanical circulatory support therapy as a bridge to transplant or recovery (new advances).

Authors:  James K Kirklin; William L Holman
Journal:  Curr Opin Cardiol       Date:  2006-03       Impact factor: 2.161

4.  Development of an implantable small right ventricular assist device.

Authors:  M Yoshikawa; K Nakata; T Takano; T Maeda; J Glueck; S Murabayashi; H Schima; E Wolner; Y Nosé
Journal:  ASAIO J       Date:  2000 May-Jun       Impact factor: 2.872

5.  The effect of intermittent low speed mode upon aortic valve opening in calves supported with a Jarvik 2000 axial flow device.

Authors:  Egemen Tuzun; Igor D Gregoric; Jeff L Conger; Kenny Golden; Robert Jarvik; O H Frazier; Kamuran A Kadipasaoglu
Journal:  ASAIO J       Date:  2005 Mar-Apr       Impact factor: 2.872

6.  Advanced heart failure: feasibility study of long-term continuous axial flow pump support.

Authors:  Michael P Siegenthaler; Stephen Westaby; O H Frazier; Jürgen Martin; Adrian Banning; Desiree Robson; John Pepper; Phillip Poole-Wilson; Friedhelm Beyersdorf
Journal:  Eur Heart J       Date:  2005-03-30       Impact factor: 29.983

7.  Incidence and clinical management of life-threatening left ventricular assist device failure.

Authors:  E J Birks; P D Tansley; M H Yacoub; C T Bowles; M Hipkin; J Hardy; N R Banner; A Khaghani
Journal:  J Heart Lung Transplant       Date:  2004-08       Impact factor: 10.247

8.  Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: physiologic implications and their relationship to patient selection.

Authors:  O H Frazier; Timothy J Myers; Stephen Westaby; Igor D Gregoric
Journal:  Ann Thorac Surg       Date:  2004-01       Impact factor: 4.330

9.  Outcome of patients surviving to heart transplantation after being mechanically bridged for more than 100 days.

Authors:  Christof Schmid; Henryk Welp; Stefan Klotz; Hideo A Baba; Markus J Wilhelm; Hans H Scheld
Journal:  J Heart Lung Transplant       Date:  2003-09       Impact factor: 10.247

10.  Biventricular support with the Jarvik 2000 axial flow pump: a feasibility study.

Authors:  Branislav Radovancevic; Igor D Gregoric; Daniel Tamez; Bojan Vrtovec; Egemen Tuzun; Hyun Keun Chee; Sheila Moore; Robert K Jarvik; O H Frazier
Journal:  ASAIO J       Date:  2003 Sep-Oct       Impact factor: 2.872

View more
  18 in total

1.  Right ventricle-sparing left ventricular resection and replacement with a continuous-flow rotary blood pump: an in vivo experiment.

Authors:  O H Frazier; Egemen Tuzun; Cuneyt Narin; William E Cohn
Journal:  Tex Heart Inst J       Date:  2010

Review 2.  Mechanical circulatory support.

Authors:  Deborah J Kozik; Mark D Plunkett
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

Review 3.  [Noncardiological surgical procedure for patients implanted with a ventricular assist device. Anesthesiological management concepts].

Authors:  F E Blum
Journal:  Med Klin Intensivmed Notfmed       Date:  2014-07-02       Impact factor: 0.840

4.  Preclinical study of a novel hydrodynamically levitated centrifugal pump for long-term cardiopulmonary support : In vivo performance during percutaneous cardiopulmonary support.

Authors:  Tomonori Tsukiya; Toshihide Mizuno; Yoshiaki Takewa; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2015-05-15       Impact factor: 1.731

Review 5.  A ventricular assist device as a bridge to recovery, decision making, or transplantation in patients with advanced cardiac failure.

Authors:  Siyamek Neragi-Miandoab
Journal:  Surg Today       Date:  2012-07-20       Impact factor: 2.549

6.  Differences in clotting parameters between species for preclinical large animal studies of cardiovascular devices.

Authors:  Toshihide Mizuno; Tomonori Tsukiya; Yoshiaki Takewa; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2017-11-09       Impact factor: 1.731

7.  Preclinical animal study of the NIPRO-ventricular assist device for use in pediatric patients.

Authors:  Noritsugu Naito; Yoshiaki Takewa; Satoru Kishimoto; Kei Iizuka; Toshihide Mizuno; Tomonori Tsukiya; Minoru Ono; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2017-11-24       Impact factor: 1.731

8.  Fontan Liver Disease: Review of an Emerging Epidemic and Management Options.

Authors:  Elisa Bradley; Benjamin Hendrickson; Curt Daniels
Journal:  Curr Treat Options Cardiovasc Med       Date:  2015-11

Review 9.  Emerging ventricular assist devices for long-term cardiac support.

Authors:  Rajan Krishnamani; David DeNofrio; Marvin A Konstam
Journal:  Nat Rev Cardiol       Date:  2010-01-12       Impact factor: 32.419

10.  HeartWare and HeartMate II left ventricular assist devices as bridge to transplantation: a comparative analysis.

Authors:  Veli K Topkara; James K O'Neill; Adam Carlisle; Eric Novak; Scott C Silvestry; Gregory A Ewald
Journal:  Ann Thorac Surg       Date:  2013-10-17       Impact factor: 4.330

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.