The effect of intermittent low speed mode upon aortic valve opening in calves supported with a Jarvik 2000 axial flow device.

Egemen Tuzun1, Igor D Gregoric, Jeff L Conger, Kenny Golden, Robert Jarvik, O H Frazier, Kamuran A Kadipasaoglu.   

Abstract

We assessed the effects of an axial flow left ventricular assist device (LVAD) upon aortic valve opening, pump outflow, and biologic and hematologic parameters when operated in intermittent low speed (ILS) mode. An ILS controller equipped Jarvik 2000 LVAD was implanted in six calves. Pump speed was maintained at 10,000 rpm, and pump outflow was measured throughout the study period (71 +/- 6 days [mean +/- SD]). Hematologic and biochemical parameters were analyzed daily for the first 10 days, weekly for the first month, and biweekly thereafter to monitor for kidney or liver dysfunction, hemolysis, bleeding, or infection. Before study termination, esmolol hydrochloride was infused to induce low cardiac output and totally impair aortic valve opening. Radiopaque cineaortography was performed over 30 second intervals (10 seconds before, 10 seconds during, and 10 seconds immediately after ILS controller activation) to assess the effect of ILS mode upon aortic valve opening. After study termination, major end organs and the major vascular tree were removed and examined macroscopically and histologically for thrombus formation and infarction; the aortic valve was examined for thickening and fusion. All pumps were explanted and examined for thrombus formation. All six calves recovered without surgical or mechanical complications. Hematologic and biochemical parameters did not change significantly between baseline and study termination. The aortic valve successfully opened when ILS mode was activated, even under low cardiac output conditions. No thrombus was detected in the major end organs and vascular tree, except for some small renal infarcts in three calves that did not affect renal function. These results indicate that operating an axial flow LVAD in ILS mode allows aortic valve opening and aortic root washout.

Entities:  

Mesh:

Year:  2005        PMID: 15839437     DOI: 10.1097/01.mat.0000155708.75802.c7

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  8 in total

1.  In vivo evaluation of the HeartWare centrifugal ventricular assist device.

Authors:  Egemen Tuzun; Katy Roberts; William E Cohn; Murat Sargin; Courtney J Gemmato; Branislav Radovancevic; O H Frazier
Journal:  Tex Heart Inst J       Date:  2007

2.  Analysis of aortic valve commissural fusion after support with continuous-flow left ventricular assist device.

Authors:  Jerson R Martina; Marguerite E I Schipper; Nicolaas de Jonge; Faiz Ramjankhan; Roel A de Weger; Jaap R Lahpor; Aryan Vink
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-06-24

3.  Outcomes of midterm circulatory support by left ventricular assist device implantation with descending aortic anastomosis.

Authors:  Kan Nawata; Takashi Nishimura; Shunei Kyo; Motoyuki Hisagi; Osamu Kinoshita; Aya Saito; Noboru Motomura; Shinichi Takamoto; Minoru Ono
Journal:  J Artif Organs       Date:  2010-10-30       Impact factor: 1.731

4.  Physiologic and hematologic concerns of rotary blood pumps: what needs to be improved?

Authors:  Tohid Pirbodaghi; Siavash Asgari; Chris Cotter; Kevin Bourque
Journal:  Heart Fail Rev       Date:  2014-03       Impact factor: 4.214

5.  Central aortic valve closure successfully treated aortic insufficiency of the patient with Jarvik 2000 continuous flow left ventricular assist device: a case report.

Authors:  Yoshifumi Itoda; Kan Nawata; Haruo Yamauchi; Osamu Kinoshita; Mitsutoshi Kimura; Minoru Ono
Journal:  J Artif Organs       Date:  2016-10-15       Impact factor: 1.731

6.  Intermittent Aortic Valve Opening and Risk of Thrombosis in Ventricular Assist Device Patients.

Authors:  Claudius Mahr; Venkat Keshav Chivukula; Patrick McGah; Anthony R Prisco; Jennifer A Beckman; Nahush A Mokadam; Alberto Aliseda
Journal:  ASAIO J       Date:  2017 Jul/Aug       Impact factor: 2.872

7.  Development of a novel drive mode to prevent aortic insufficiency during continuous-flow LVAD support by synchronizing rotational speed with heartbeat.

Authors:  Yuichiro Kishimoto; Yoshiaki Takewa; Mamoru Arakawa; Akihide Umeki; Masahiko Ando; Takashi Nishimura; Yutaka Fujii; Toshihide Mizuno; Motonobu Nishimura; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2013-01-23       Impact factor: 1.731

8.  Factors influencing the functional status of aortic valve in ovine models supported by continuous-flow left ventricular assist device.

Authors:  Xin-Yi Yu; Jian-Wei Shi; Yi-Rui Zang; Jie-Min Zhang; Zhi-Gang Liu
Journal:  Artif Organs       Date:  2022-03-03       Impact factor: 2.663