Literature DB >> 18158596

Frequency doubling technology threshold visual field results vary with software version.

Nagisa Kojima1, Shiroaki Shirato, Yuko Ohno.   

Abstract

PURPOSE: To determine the comparability of test results obtained using two versions of commercially available frequency doubling technology (FDT) software for visual field testing.
METHODS: One hundred glaucoma patients underwent duplicate N-30 threshold FDT visual field examinations of one eye; one test was performed with the recently released software version 3.0 and the other test with the earlier version 2.6. Results were compared in terms of test duration, mean deviation (MD), pointwise sensitivity, pointwise deviations from the age-normal threshold, and average sensitivity.
RESULTS: Results obtained with version 3.0 differed significantly from those obtained with version 2.6. Version 3.0 tests took on average 14.2% less time (P < 0.0001), and the average MD was 0.53 dB more negative (P < 0.01). Version 3.0 threshold estimates derived from a mixed-model analysis were significantly lower at 15 of 19 test points (P = 0.0074). At 14 of the 19 test point locations, the mixed-model analysis also showed that threshold measurements were more frequently outside the normal range with version 3.0 (P = 0.022).
CONCLUSIONS: We found statistically significant differences between results obtained using FDT versions 3.0 and 2.6, which might affect clinical management. New age-corrected normal values for version 3.0 should be obtained and made available to users, and new baseline field data should be established whenever a patient is switched from version 2.6 to version 3.0. (c) Japanese Ophthalmological Society 2007

Entities:  

Mesh:

Year:  2007        PMID: 18158596     DOI: 10.1007/s10384-007-0469-7

Source DB:  PubMed          Journal:  Jpn J Ophthalmol        ISSN: 0021-5155            Impact factor:   2.447


  7 in total

1.  Visual field indexes for version 3.0 of frequency doubling perimetry in patients with glaucoma.

Authors:  Da Wei Zhang; Katsuyuki Minowa; Tadao Hanawa; Naoya Fujimoto; Kazuhide Tsukamoto; Emiko Adachi-Usami
Journal:  Am J Ophthalmol       Date:  2003-01       Impact factor: 5.258

2.  Screening for glaucomatous visual field loss with frequency-doubling perimetry.

Authors:  C A Johnson; S J Samuels
Journal:  Invest Ophthalmol Vis Sci       Date:  1997-02       Impact factor: 4.799

3.  A new generation of algorithms for computerized threshold perimetry, SITA.

Authors:  B Bengtsson; J Olsson; A Heijl; H Rootzén
Journal:  Acta Ophthalmol Scand       Date:  1997-08

4.  Glaucoma follow-up when converting from long to short perimetric threshold tests.

Authors:  A Heijl; B Bengtsson; V M Patella
Journal:  Arch Ophthalmol       Date:  2000-04

5.  Frequency-doubling technology perimetry and optical defocus.

Authors:  Andrew John Anderson; Chris A Johnson
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-09       Impact factor: 4.799

6.  Clinical evaluation of SITA: a new family of perimetric testing strategies.

Authors:  S Shirato; R Inoue; K Fukushima; Y Suzuki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1999-01       Impact factor: 3.117

7.  Clinical classification of glaucomatous visual field loss by frequency doubling perimetry.

Authors:  W E Sponsel; S Arango; Y Trigo; J Mensah
Journal:  Am J Ophthalmol       Date:  1998-06       Impact factor: 5.258

  7 in total