Literature DB >> 18004915

Mechanical characteristics of various orthodontic mini-screws in relation to artificial cortical bone thickness.

Young-Youn Song1, Jung-Yul Cha, Chung-Ju Hwang.   

Abstract

OBJECTIVE: To evaluate the effect of cortical bone thickness on the maximum insertion and removal torque of different types of self-drilling mini-screws and to determine if torque depends on the screw design.
MATERIALS AND METHODS: Three different types of self-drilling mini-screws (cylindrical type [Cl], taper type [Ta], taper type [Tb]) were inserted with the use of a driving torque tester at a constant speed of 3 rotations per minute. Experimental bone blocks with different cortical bone thicknesses were used as specimens.
RESULTS: Differences in the cortical bone thickness had little effect on the maximum insertion and removal torque in Cl. However, with Ta and Tb, the maximum insertion torque increased as the cortical bone thickness increased. The maximum insertion torque of Tb was highest in all situations, followed by Ta and Tb, in that order. Cl showed less torque loss in all cortical bone thicknesses and a longer removal time compared to Ta or Tb. There were significant relationships between cortical bone thickness, maximum insertion and removal torque, and implantation time in each type of self-drilling mini-screw.
CONCLUSION: Since different screw designs showed different insertion torques with increases in cortical bone thickness, the suitable screw design should be selected according to the cortical thickness at the implant site.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18004915     DOI: 10.2319/090606-363.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  13 in total

1.  Surface characteristics and mechanical behavior of retrieved orthodontic microimplants.

Authors:  Lu Lu; Hyo-Sang Park
Journal:  J Zhejiang Univ Sci B       Date:  2018-05       Impact factor: 3.066

2.  Evaluation of Interdental Spaces of the Mandibular Posterior Area for Orthodontic Mini-Implants with Cone-Beam Computed Tomography.

Authors:  Seyed Hossein Moslemzadeh; Aydin Sohrabi; Ali Rafighi; Yusef Kananizadeh; Amin Nourizadeh
Journal:  J Clin Diagn Res       Date:  2017-04-01

3.  Stress distributions of a bracket type orthodontic miniscrew and the surrounding bone under moment loadings: Three-dimensional finite element analysis.

Authors:  Shabnam Ajami; Ahmad Mina; Seyed Amin Nabavizadeh
Journal:  J Orthod Sci       Date:  2016 Apr-Jun

4.  Effects of predrilling on the osseointegration potential of mini-implants.

Authors:  Yun-Seob Shin; Hyo-Won Ahn; Young-Guk Park; Seong-Hun Kim; Kyu-Rhim Chung; Il-Sik Cho; Gerald Nelson
Journal:  Angle Orthod       Date:  2012-04-12       Impact factor: 2.079

5.  Factors affecting stresses in cortical bone around miniscrew implants: a three-dimensional finite element study.

Authors:  Ramzi Duaibis; Budi Kusnoto; Raghu Natarajan; Linping Zhao; Carla Evans
Journal:  Angle Orthod       Date:  2012-03-05       Impact factor: 2.079

6.  Mechanical properties of self-drilling orthodontic micro-implants with different diameters.

Authors:  Yan Chen; Hee Moon Kyung; Limin Gao; Won-Jae Yu; Eun-Ju Bae; Sang-Min Kim
Journal:  Angle Orthod       Date:  2010-09       Impact factor: 2.079

7.  Biomechanical characteristics and reinsertion guidelines for retrieved orthodontic miniscrews.

Authors:  Chooryung J Chung; Kil-Young Jung; Yoon Jeong Choi; Kyung-Ho Kim
Journal:  Angle Orthod       Date:  2014-03-03       Impact factor: 2.079

8.  Effects of placement angle and direction of orthopedic force application on the stability of orthodontic miniscrews.

Authors:  Jihye Lee; Ji Young Kim; Yoon Jeong Choi; Kyung-Ho Kim; Chooryung J Chung
Journal:  Angle Orthod       Date:  2012-12-14       Impact factor: 2.079

9.  An in vitro study of factors affecting the primary stability of orthodontic mini-implants.

Authors:  Lindsy Holm; Susan J Cunningham; Aviva Petrie; Richard R J Cousley
Journal:  Angle Orthod       Date:  2012-05-07       Impact factor: 2.079

10.  Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies.

Authors:  André Walter; Heinz Winsauer; Jordi Marcé-Nogué; Sergi Mojal; Andreu Puigdollers
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.