| Literature DB >> 17701041 |
Kai Nassenstein1, Kai-Uwe Waltering, Sebastian Kelle, Thomas Schlosser, Frank Breuckmann, Stefan Maderwald, Peter Hunold, Eike Nagel, Jörg Barkhausen.
Abstract
The purpose of the study was to estimate T1 values of blood and myocardium after a single injection of Vasovist and to assess Vasovist for magnetic resonance coronary angiography (MRCA). For all exams 0.05 mmol/kg of Vasovist was injected. T1 values of blood and myocardium were estimated over 30 min after injection. Twelve volunteers were examined on a 1.5-T Siemens system using a SSFP sequence with incrementally increasing inversion times for T1-estimation and a breath-hold 3D IR-FLASH sequence for MRCA. Eleven examinations were performed on 1.5-T Philips system using the Look-Locker approach for T1 estimation and a whole-heart inversion-prepared, 3D SSFP sequence for MRCA. SNR, CNR and image quality were assessed. T1 values of blood (5 min: 230 ms vs. 30 min: 275 ms) and myocardium (5 min: 99 ms vs. 30 min: 130 ms) increased over time. Whereas the blood SNR (1 min: 23.6 vs. 30 min: 21.2) showed no significant differences, the blood-to-myocardium CNR (1 min: 18.1 vs. 30 min: 13.8) and the image quality (1 min: 2.9 vs. 30 min: 3.8) degraded over time. Due to long plasma half-time the T1-shortening effect of Vasovist remains effective over 30 min, which allows for multiple breath-hold or high-resolution MRCA.Mesh:
Substances:
Year: 2007 PMID: 17701041 DOI: 10.1007/s00330-007-0720-0
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315