Comparison of FloTrac cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements.

M Cannesson1, Y Attof, P Rosamel, P Joseph, O Bastien, J-J Lehot.   

Abstract

BACKGROUND: Arterial pulse waveform analysis has been proposed for cardiac output (CO) determination and monitoring without calibration or thermodilution (FloTrac/Vigileo; Edwards Lifesciences, Irvine, CA, USA). The accuracy and clinical applicability of this technology has not been fully evaluated. We designed this prospective study to compare the accuracy of the FloTrac system (CO(FT)) vs. pulmonary artery catheter standard bolus thermodilution (CO(PAC) ) in patients undergoing coronary artery bypass grafting.
METHODS: We studied 11 patients referred for coronary artery bypass grafting. CO(FT) and CO(PAC) were determined at six time points in the operating room including before and 5 min after volume expansion (500 mL 6% hetastarch). Measurements were performed on arrival in the intensive care unit and every 4 h afterwards. Bland-Altman analysis was used to assess the agreement between CO(FT) and CO(PAC).
RESULTS: CO(PAC) ranged from 2.0 to 7.6 L min-1 and CO(FT) ranged from 1.9 to 8.2 L min-1. There was a significant relationship between CO(PAC) and CO(FT) (r = 0.662; P < 0.001). Agreement between CO(PAC) and CO(FT) was -0.26 +/- 0.87 L min-1. Volume expansion induced a significant increase in both CO(PAC) and CO(FT) (from 3.4 +/- 0.8 to 4.4 +/- 1.0 L min-1; P < 0.001 and from 3.9 +/- 1.2 to 5.0 +/- 1.1 L min-1; P < 0.001, respectively) and there was a significant relationship between percent change in CO(PAC) and CO(FT) following volume expansion (r = 0.722; P = 0.01).
CONCLUSION: We found clinically acceptable agreement between CO(FT) and CO(PAC) in this setting. This new device has potential clinical applications.

Entities:  

Mesh:

Year:  2007        PMID: 17666154     DOI: 10.1017/S0265021507001056

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  20 in total

Review 1.  Cardiac output monitoring devices: an analytic review.

Authors:  Jahan Porhomayon; Ali El-Solh; Peter Papadakos; Nader Djalal Nader
Journal:  Intern Emerg Med       Date:  2011-12-07       Impact factor: 3.397

Review 2.  Pulse pressure variation: where are we today?

Authors:  Maxime Cannesson; Mateo Aboy; Christoph K Hofer; Mohamed Rehman
Journal:  J Clin Monit Comput       Date:  2011-02       Impact factor: 2.502

3.  Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.

Authors:  Philip J Peyton; Daniel Thompson; Paul Junor
Journal:  J Clin Monit Comput       Date:  2008-07-12       Impact factor: 2.502

Review 4.  Minimally invasive monitoring of cardiac output in the cardiac surgery intensive care unit.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Giorgio della Rocca; Maxime Cannesson; Christoph K Hofer
Journal:  Curr Heart Fail Rep       Date:  2010-09

5.  Effective evaluation of arterial pulse waveform analysis by two-dimensional stroke volume variation-stroke volume index plots.

Authors:  Teiji Sawa; Mao Kinoshita; Atsushi Kainuma; Koichi Akiyama; Yoshifumi Naito; Hideya Kato; Fumimasa Amaya; Keiji Shigemi
Journal:  J Clin Monit Comput       Date:  2016-08-04       Impact factor: 2.502

6.  Lower intra-abdominal pressure has no cardiopulmonary benefits during laparoscopic colorectal surgery: a double-blind, randomized controlled trial.

Authors:  Youn Joung Cho; Hyesun Paik; Seung-Yong Jeong; Ji Won Park; Woo Young Jo; Yunseok Jeon; Kook Hyun Lee; Jeong-Hwa Seo
Journal:  Surg Endosc       Date:  2018-05-14       Impact factor: 4.584

7.  Validation of continuous cardiac output technologies: consensus still awaited.

Authors:  Maurizio Cecconi; Andrew Rhodes
Journal:  Crit Care       Date:  2009-06-26       Impact factor: 9.097

8.  Clinical evaluation of the flotrac/Vigileo system for continuous cardiac output monitoring in patients undergoing regional anesthesia for elective cesarean section: a pilot study.

Authors:  José Otavio Costa Auler; Marcelo L A Torres; Mônica M Cardoso; Thais C Tebaldi; André P Schmidt; Mario M Kondo; Marcelo Zugaib
Journal:  Clinics (Sao Paulo)       Date:  2010-06       Impact factor: 2.365

9.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

10.  Arterial pressure changes during the Valsalva maneuver to predict fluid responsiveness in spontaneously breathing patients.

Authors:  Manuel Ignacio Monge García; Anselmo Gil Cano; Juan Carlos Díaz Monrové
Journal:  Intensive Care Med       Date:  2008-10-02       Impact factor: 17.440